CCQM-K158
MEASURAND Copper in polished rice flour
NOMINAL VALUE 0.5 mg/kg - 5 mg/kg
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
CCQM-K158
MEASURAND Mercury in polished rice flour
NOMINAL VALUE 0.1 mg/kg - 5 mg/kg
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
Enlarged graph
CCQM-K158
MEASURAND Potassium in polished rice flour
NOMINAL VALUE 100 mg/kg - 1000 mg/kg
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
CCQM-K158
MEASURAND Sodium in polished rice flour
NOMINAL VALUE 2 mg/kg - 10 mg/kg
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
CCQM-K158
MEASURAND Lead in polished rice flour
NOMINAL VALUE 0.1 mg/kg - 1 mg/kg
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
CCQM-K158
MEASURAND Antimony in polished rice flour
NOMINAL VALUE 0.5 mg/kg - 5 mg/kg
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
CCQM-K158
MEASURAND Total arsenic in polished rice flour
NOMINAL VALUE > 0.02 mg/kg
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
CCQM-K158
MEASURAND Inorganic arsenic in polished rice flour
NOMINAL VALUE > 0.02 mg/kg
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
CCQM-K158
MEASURAND Copper in polished rice flour
NOMINAL VALUE 0.5 mg/kg - 5 mg/kg
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
mg/kg |
mg/kg |
mg/kg |
||
NIS* | -0.406 | 0.213 | ||
INTI* | -0.291 | 0.229 | ||
NIMT | -0.023 | 0.072 | ||
HSA | 0.007 | 0.070 | ||
KRISS | 0.0188 | 0.0625 | ||
NMIJ | 0.029 | 0.070 | ||
LATU | 0.036 | 0.072 | ||
NMISA | 0.040 | 0.089 | ||
MIRS/IJS/F-2,O-2 | 0.145 | 0.156 |
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
CCQM-K158
MEASURAND Mercury in polished rice flour
NOMINAL VALUE 0.1 mg/kg - 5 mg/kg
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
mg/kg |
mg/kg |
mg/kg |
||
INTI | -0.040 | 0.080 | ||
MIRS/IJS/F-2,O-2 | -0.038 | 0.040 | ||
NMISA* | -0.023 | 0.040 | ||
GL | -0.005 | 0.028 | ||
NMIA | -0.001 | 0.024 | ||
NIM | 0.0042 | 0.0172 | ||
HSA | 0.0050 | 0.0137 | ||
LATU | 0.0122 | 0.0205 | ||
NIMT* | 0.0182 | 0.0398 | ||
KRISS* | 0.0228 | 0.0396 | ||
NIS* | 0.6372 | 0.0831 |
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
CCQM-K158
MEASURAND Potassium in polished rice flour
NOMINAL VALUE 100 mg/kg - 1000 mg/kg
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
mg/kg |
mg/kg |
mg/kg |
||
NIMT* | -72.6 | 43.5 | ||
NIS | -70.7 | 77.9 | ||
NMISA* | -28.6 | 25.2 | ||
MIRS/IJS/F-2,O-2 | -20.6 | 34.1 | ||
PTB | -2.3 | 6.7 | ||
KRISS | 2.0 | 4.2 | ||
INM (CO) | 5.2 | 23.6 | ||
NMIJ | 5.6 | 15.2 | ||
LATU | 10.4 | 14.6 |
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
CCQM-K158
MEASURAND Sodium in polished rice flour
NOMINAL VALUE 2 mg/kg - 10 mg/kg
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
mg/kg |
mg/kg |
mg/kg |
||
NIS* | -1.74 | 0.27 | ||
MIRS/IJS/F-2,O-2 | -0.15 | 0.29 | ||
KRISS | -0.02 | 0.33 | ||
NMIJ | 0.031 | 0.162 | ||
NMISA | 0.05 | 0.24 | ||
NIMT* | 3.35 | 0.87 |
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
CCQM-K158
MEASURAND Lead in polished rice flour
NOMINAL VALUE 0.1 mg/kg - 1 mg/kg
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
mg/kg |
mg/kg |
mg/kg |
||
NMISA* | -0.0596 | 0.0372 | ||
INM (CO)* | -0.0336 | 0.0460 | ||
METAS* | -0.0186 | 0.0382 | ||
MIRS/IJS/F-2,O-2 | -0.0126 | 0.0154 | ||
GL | -0.0016 | 0.0099 | ||
NMIJ | -0.0016 | 0.0085 | ||
NIMT | -0.0016 | 0.0085 | ||
HSA | 0.0004 | 0.0105 | ||
LATU | 0.0014 | 0.0085 | ||
KRISS | 0.0024 | 0.0109 | ||
PTB | 0.0024 | 0.0073 | ||
NIS | 0.0034 | 0.0379 | ||
NIM | 0.0044 | 0.0072 | ||
NMIA* | 0.0134 | 0.0383 | ||
INTI* | 0.0254 | 0.0412 | ||
INMETRO* | 0.0284 | 0.0386 |
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
CCQM-K158
MEASURAND Antimony in polished rice flour
NOMINAL VALUE 0.5 mg/kg - 5 mg/kg
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
mg/kg |
mg/kg |
mg/kg |
||
NIMT* | -0.0743 | 0.0425 | ||
MIRS/IJS/F-2,O-2* | -0.0673 | 0.0638 | ||
INTI | -0.0143 | 0.0725 | ||
GL | -0.0063 | 0.0358 | ||
NMIA | -0.0023 | 0.0340 | ||
KRISS | -0.0017 | 0.0081 | ||
HSA | 0.0037 | 0.0332 | ||
NMIJ | 0.0084 | 0.0114 |
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
CCQM-K158
MEASURAND Total arsenic in polished rice flour
NOMINAL VALUE > 0.02 mg/kg
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
mg/kg |
mg/kg |
mg/kg |
||
NIMT* | -0.0134 | 0.0041 | ||
METAS* | -0.0094 | 0.0060 | ||
MIRS/IJS/F-2,O-2 | -0.0027 | 0.0072 | ||
LGC | -0.0010 | 0.0027 | ||
NIM | -0.0004 | 0.0042 | ||
NMIJ | -0.0001 | 0.0041 | ||
NMIA | 0.0003 | 0.0058 | ||
KRISS | 0.0006 | 0.0139 | ||
LATU | 0.0006 | 0.0041 | ||
HSA | 0.0008 | 0.0040 | ||
NRC | 0.0011 | 0.0040 | ||
GL | 0.0016 | 0.0055 | ||
INTI | 0.0030 | 0.0114 |
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
CCQM-K158
MEASURAND Inorganic arsenic in polished rice flour
NOMINAL VALUE > 0.02 mg/kg
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in mg/kg.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
mg/kg |
mg/kg |
mg/kg |
||
NRC* | -0.0290 | 0.0206 | ||
NIMT* | -0.0093 | 0.0224 | ||
GL | -0.0014 | 0.0076 | ||
HSA | -0.0007 | 0.0058 | ||
LGC | -0.0003 | 0.0035 | ||
NMIJ | 0.0004 | 0.0023 | ||
NIM | 0.0036 | 0.0032 |
*Ui for this participant was estimated after summing in quadrature the participant's reported uncertainty value and the dark uncertainty (tau) associated with the key comparison reference value (KCRV) because estimation using the reported uncertainty value alone results in nonequivalence with the KCRV.
Metrology area, Sub-field | Chemistry and Biology, Inorganics |
Description | Elements and inorganic arsenic in rice flour |
Time of measurements | 2021 - 2022 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
Measurand | Mass fraction of Cu, Hg, K, Na, Sb or Pb in rice flour Mass fraction of total arsenic and inorganic arsenic in rice flour |
Transfer device | Sample A: 10 g sample in an amber glass bottle (two bottles) Sample B: 10 g sample in an amber glass bottle (one bottle) |
Comparison type | Key Comparison |
Consultative Committee | CCQM (Consultative Committee for Amount of Substance) |
Conducted by | CCQM (Consultative Committee for Amount of Substance) |
Comments | This study has two parts: Part A which covers the trace elements measurement analysis without arsenic. Part B which covers the measurement analysis of total arsenic and inorganic arsenic. |
Pilot institute |
NMIJ AIST
National Metrology Institute of Japan Japan |
Contact person | Kazumi Inagaki |
Additional |
Pilot laboratory | |
---|---|
NMIJ AIST |
National Metrology Institute of Japan, Japan, APMP |
GL |
Government Laboratory, Hong Kong, China, APMP |
HSA |
Health Science Authority, Singapore, APMP |
INM (CO) |
Instituto Nacional de Metrología de Colombia, Colombia, SIM |
INMETRO |
Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM |
INTI |
Instituto Nacional de Tecnologia Industrial, Argentina, SIM |
KRISS |
Korea Research Institute of Standards and Science, Korea, Republic of, APMP |
LATU |
Laboratorio Tecnologico del Uruguay, Uruguay, SIM |
LGC |
Laboratory of the Government Chemist, United Kingdom, EURAMET |
METAS |
Federal Institute of Metrology, Switzerland, EURAMET |
MIRS/IJS/F-2,O-2 |
MIRS/Jozef Stefan Institute/Low and Medium Energy Physics F2, Environmental Sciences O2, Slovenia, EURAMET |
NIM |
National Institute of Metrology, China, APMP |
NIMT |
National Institute of Metrology (Thailand), Thailand, APMP |
NIS |
National Institute for Standards, Egypt, AFRIMETS |
NMIA |
National Measurement Institute, Australia, Australia, APMP |
NMISA |
National Metrology Institute of South Africa, South Africa, AFRIMETS |
NRC |
National Research Council, Canada, SIM |
PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
CCQM-K158
MEASURAND Copper in polished rice flour
NOMINAL VALUE 0.5 mg/kg - 5 mg/kg
The key comparison reference value xR and its associated standard uncertainty uR were calculated using the Adaptive Weighted Average statistical model following the NIST Decision Tree, as explained on page 23 of the CCQM-K158 final report. The reported result for NIS was not included in the calculation of xR.
xR = 1.345 mg/kg uR = 0.0030 mg/kg
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in mg/kg:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence is calculated as explained on page 27 of the CCQM-K158 final report.
CCQM-K158
MEASURAND Mercury in polished rice flour
NOMINAL VALUE 0.1 mg/kg - 5 mg/kg
The key comparison reference value xR and its associated standard uncertainty uR were calculated using the Adaptive Weighted Average statistical model following the NIST Decision Tree, as explained on page 23 of the CCQM-K158 final report. The reported result for NIS was not included in the calculation of xR.
xR = 0.4788 mg/kg uR = 0.0059 mg/kg
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in mg/kg:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence is calculated as explained on page 27 of the CCQM-K158 final report.
CCQM-K158
MEASURAND Potassium in polished rice flour
NOMINAL VALUE 100 mg/kg - 1000 mg/kg
The key comparison reference value xR and its associated standard uncertainty uR were calculated using the Adaptive Weighted Average statistical model following the NIST Decision Tree, as explained on page 23 of the CCQM-K158 final report. The reported result for NIMT and NIS were not included in the calculation of xR.
xR = 616.6 mg/kg uR = 3.3 mg/kg
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in mg/kg:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence is calculated as explained on page 27 of the CCQM-K158 final report.
CCQM-K158
MEASURAND Sodium in polished rice flour
NOMINAL VALUE 2 mg/kg - 10 mg/kg
The key comparison reference value xR and its associated standard uncertainty uR were calculated using the Adaptive Weighted Average statistical model following the NIST Decision Tree, as explained on page 23 of the CCQM-K158 final report. The reported result for NIS and NIMT were not included in the calculation of xR.
xR = 5.399 mg/kg uR = 0.064 mg/kg
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in mg/kg:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence is calculated as explained on page 27 of the CCQM-K158 final report.
CCQM-K158
MEASURAND Lead in polished rice flour
NOMINAL VALUE 0.1 mg/kg - 1 mg/kg
The key comparison reference value xR and its associated standard uncertainty uR were calculated using the Adaptive Weighted Average statistical model following the NIST Decision Tree, as explained on page 23 of the CCQM-K158 final report.
xR = 0.2166 mg/kg uR = 0.0030 mg/kg
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in mg/kg:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence is calculated as explained on page 27 of the CCQM-K158 final report.
CCQM-K158
MEASURAND Antimony in polished rice flour
NOMINAL VALUE 0.5 mg/kg - 5 mg/kg
The key comparison reference value xR and its associated standard uncertainty uR were calculated using the Adaptive Weighted Average statistical model following the NIST Decision Tree, as explained on page 23 of the CCQM-K158 final report. The reported result for NIMT was not included in the calculation of xR.
xR = 1.0130 mg/kg uR = 0.0050 mg/kg
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in mg/kg:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence is calculated as explained on page 27 of the CCQM-K158 final report.
CCQM-K158
MEASURAND Total arsenic in polished rice flour
NOMINAL VALUE > 0.02 mg/kg
The key comparison reference value xR and its associated standard uncertainty uR were calculated using the Adaptive Weighted Average statistical model following the NIST Decision Tree, as explained on pages 50 and 51 of the CCQM-K158 final report. The reported result for NIMT and METAS were not included in the calculation of xR.
xR = 0.1064 mg/kg uR = 0.0006 mg/kg
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in mg/kg:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence is calculated as explained on page 54 of the CCQM-K158 final report.
CCQM-K158
MEASURAND Inorganic arsenic in polished rice flour
NOMINAL VALUE > 0.02 mg/kg
The key comparison reference value xR and its associated standard uncertainty uR were calculated using the Adaptive Weighted Average statistical model following the NIST Decision Tree, as explained on pages 50 and 51 of the CCQM-K158 final report. The reported result for NRC and NIMT were not included in the calculation of xR.
xR = 0.0903 mg/kg uR = 0.0006 mg/kg
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in mg/kg:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence is calculated as explained on page 54 of the CCQM-K158 final report.
CCQM-K158
MEASURAND Copper in polished rice flour
NOMINAL VALUE 0.5 mg/kg - 5 mg/kg
xi measured mass fraction reported by laboratory i, Labi
ui combined standard uncertainty of xi
Labi | xi | ui | |
mg/kg | mg/kg | ||
NIS* | 0.939 | 0.088# | |
INTI | 1.054 | 0.097# | |
NIMT | 1.322 | 0.020 | |
HSA | 1.352 | 0.015 | |
KRISS | 1.3640 | 0.0062 | |
NMIJ | 1.3740 | 0.018 | |
LATU | 1.381 | 0.020 | |
NMISA | 1.385 | 0.033 | |
MIRS/IJS/F-2,O-2 | 1.490 | 0.060 |
* Reported value is not included for the calculation of the key comparison reference value
# Participant's value includes the uncertainty and the dark uncertainty, τ, summed in quadrature
CCQM-K158
MEASURAND Mercury in polished rice flour
NOMINAL VALUE 0.1 mg/kg - 5 mg/kg
xi measured mass fraction reported by laboratory i, Labi
ui combined standard uncertainty of xi
Labi | xi | ui | |
mg/kg | mg/kg | ||
INTI | 0.439 | 0.040 | |
MIRS/IJS/F-2,O-2 | 0.441 | 0.015 | |
NMISA | 0.456 | 0.017# | |
GL | 0.474 | 0.013 | |
NMIA | 0.478 | 0.010 | |
NIM | 0.4830 | 0.0064 | |
HSA | 0.4838 | 0.0032 | |
LATU | 0.4910 | 0.008 | |
NIMT | 0.4970 | 0.0171# | |
KRISS | 0.5016 | 0.0169# | |
NIS* | 1.116 | 0.0409# |
* Reported value is not included for the calculation of the key comparison reference value
# Participant's value includes the uncertainty and the dark uncertainty, τ, summed in quadrature
CCQM-K158
MEASURAND Potassium in polished rice flour
NOMINAL VALUE 100 mg/kg - 1000 mg/kg
xi measured mass fraction reported by laboratory i, Labi
ui combined standard uncertainty of xi
Labi | xi | ui | |
mg/kg | mg/kg | ||
NIMT* | 539 | 22# | |
NIS* | 541 | 40 | |
NMISA | 583 | 13# | |
MIRS/IJS/F-2,O-2 | 591 | 18 | |
PTB | 609.3 | 3.9 | |
KRISS | 613.6 | 1.9 | |
INM (CO) | 617 | 12 | |
NMIJ | 617.2 | 8.2 | |
LATU | 622.0 | 7.8 |
* Reported value is not included for the calculation of the key comparison reference value
# Participant's value includes the uncertainty and the dark uncertainty, τ, summed in quadrature
CCQM-K158
MEASURAND Sodium in polished rice flour
NOMINAL VALUE 2 mg/kg - 10 mg/kg
xi measured mass fraction reported by laboratory i, Labi
ui combined standard uncertainty of xi
Labi | xi | ui | |
mg/kg | mg/kg | ||
NIS* | 3.66 | 0.12# | |
MIRS/IJS/F-2,O-2 | 5.25 | 0.16 | |
KRISS | 5.38 | 0.18 | |
NMIJ | 5.431 | 0.090 | |
NMISA | 5.45 | 0.14 | |
NIMT* | 8.75 | 0.44# |
* Reported value is not included for the calculation of the key comparison reference value
# Participant's value includes the uncertainty and the dark uncertainty, τ, summed in quadrature
CCQM-K158
MEASURAND Lead in polished rice flour
NOMINAL VALUE 0.1 mg/kg - 1 mg/kg
xi measured mass fraction reported by laboratory i, Labi
ui combined standard uncertainty of xi
Labi | xi | ui | |
mg/kg | mg/kg | ||
NMISA | 0.1570 | 0.0170# | |
INM (CO) | 0.183 | 0.022# | |
METAS | 0.198 | 0.018# | |
MIRS/IJS/F-2,O-2 | 0.204 | 0.006 | |
GL | 0.2146 | 0.0045 | |
NMIJ | 0.2146 | 0.0028 | |
NIMT | 0.215 | 0.003 | |
HSA | 0.2168 | 0.0043 | |
LATU | 0.2181 | 0.0033 | |
KRISS | 0.2186 | 0.0033 | |
PTB | 0.2189 | 0.0022 | |
NIS | 0.220 | 0.019 | |
NIM | 0.221 | 0.002 | |
NMIA | 0.2305 | 0.0173# | |
INTI | 0.2415 | 0.019# | |
INMETRO | 0.2448 | 0.0175# |
# Participant's value includes the uncertainty and the dark uncertainty, τ, summed in quadrature
CCQM-K158
MEASURAND Antimony in polished rice flour
NOMINAL VALUE 0.5 mg/kg - 5 mg/kg
xi measured mass fraction reported by laboratory i, Labi
ui combined standard uncertainty of xi
Labi | xi | ui | |
mg/kg | mg/kg | ||
NIMT* | 0.939 | 0.021# | |
MIRS/IJS/F-2,O-2 | 0.946 | 0.033# | |
INTI | 0.999 | 0.038 | |
GL | 1.007 | 0.019 | |
NMIA | 1.011 | 0.018 | |
KRISS | 1.0115 | 0.0038 | |
HSA | 1.017 | 0.018 | |
NMIJ | 1.0217 | 0.0066 |
* Reported value is not included for the calculation of the key comparison reference value
# Participant's value includes the uncertainty and the dark uncertainty, τ, summed in quadrature
CCQM-K158
MEASURAND Total arsenic in polished rice flour
NOMINAL VALUE > 0.02 mg/kg
xi measured mass fraction reported by laboratory i, Labi
ui combined standard uncertainty of xi
Labi | xi | ui | |
mg/kg | mg/kg | ||
NIMT* | 0.093 | 0.004# | |
METAS* | 0.0970 | 0.0050# | |
MIRS/IJS/F-2,O-2 | 0.1037 | 0.0037 | |
LGC | 0.1053 | 0.0012 | |
NIM | 0.1060 | 0.0020 | |
NMIJ | 0.1063 | 0.0020 | |
NMIA | 0.1067 | 0.0029 | |
KRISS | 0.107 | 0.007 | |
LATU | 0.1070 | 0.0019 | |
HSA | 0.1072 | 0.0019 | |
NRC | 0.1075 | 0.0020 | |
GL | 0.1080 | 0.0027 | |
INTI | 0.1094 | 0.0057 |
* Reported value is not included for the calculation of the key comparison reference value
# Participant's value includes the uncertainty and the dark uncertainty, τ, summed in quadrature
CCQM-K158
MEASURAND Inorganic arsenic in polished rice flour
NOMINAL VALUE > 0.02 mg/kg
xi measured mass fraction reported by laboratory i, Labi
ui combined standard uncertainty of xi
Labi | xi | ui | |
mg/kg | mg/kg | ||
NRC* | 0.0613 | 0.0103# | |
NIMT* | 0.0810 | 0112# | |
GL | 0.0889 | 0.0039 | |
HSA | 0.0896 | 0.0029 | |
LGC | 0.0901 | 0.0016 | |
NMIJ | 0.09070 | 0.00096 | |
NIM | 0.0939 | 0.0015 |
* Reported value is not included for the calculation of the key comparison reference value
# Participant's value includes the uncertainty and the dark uncertainty, τ, summed in quadrature