Comparison
Comparison close
Results

BIPM.RI(II)-K1.I-123 and EUROMET.RI(II)-K2.I-123

MEASURAND : Equivalent activity of 123I

Degrees of equivalence Di and expanded uncertainty Ui (k = 2) expressed in MBq

Left part of the graph : participants in BIPM.RI(II)-K1.I-123
Right part of the graph : participant in EUROMET.RI(II)-K2.I-123 (year 1983)
Black squares : participants' results prior to 1985

Note: the right-hand axis shows approximate relative values only.

Comparison
Comparison close
Results

BIPM.RI(II)-K1.I-123 and EUROMET.RI(II)-K2.I-123

MEASURAND : Equivalent activity of 123I

   
Lab iDiUi
 / MBq/ MBq
IRMM-0.11.3
LNE-LNHB-0.21.1
IRA-0.81.2
PTB1.22.0
   
NPL0.21.2

In blue: participant in EUROMET.RI(II)-K2.I-123 only.

Results are presented under A4 printable format in Summary Results (.PDF file).

Comparison
Comparison close
BIPM comparison
Linked comparison
BIPM.RI(II)-K4.I-123
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity of radionuclide I-123
Time of measurements 2022 -
Status Continuous, approved for equivalence
References

BIPM.RI(II)-K4 protocol

Measurand SIR-TI equivalent activity in kBq
Transfer device SIR Transfer Instrument
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by BIPM (Bureau International des Poids et Mesures)
Comments SIR: Système International de Référence, International Reference System
SIR-TI: SIR Transfer Instrument

The status for not yet completed comparisons is indicated in "Participants".
Pilot institute BIPM
Bureau International des Poids et Mesures
BIPM - International Organization
Contact person Carine Michotte

+33 (0) 1 45 07 70 85
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BIPM

Bureau International des Poids et Mesures, BIPM - International Organization, N/A

ANSTO

Australian Nuclear Science and Technology Organisation, Australia, APMP

report in progress draft A

KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

Measurements complete

LNMRI/IRD

Laboratorio Nacional de Metrologia das Radiacoes Ionizantes, IRD, Brazil, SIM

Planned

Comparison
Comparison close
Results

BIPM.RI(II)-K1.I-123 and EUROMET.RI(II)-K2.I-123

MEASURAND : Equivalent activity of 123I

• Key comparison BIPM.RI(II)-K1.I-123

Key comparison reference value:
The SIR reference value for this radionuclide is xR = 120.25 MBq, with a standard uncertainty, uR = 0.41 MBq, see section 4.1 of the Final Report.

The degree of equivalence of each laboratory with respect to the reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. With n the number of laboratories, when each laboratory has contributed to the computation of xR, see Metrologia, 42, 140-144.

The degree of equivalence between two laboratories is given by a pair of terms:
Dij = Di - Dj = (xi - xj) and Uij, its expanded uncertainty (k = 2), both expressed in MBq. The approximation Uij ~ 2(ui2 + uj2)1/2 is used for the computation of the pair-wise degrees of equivalence.

• Linking EUROMET.RI(II)-K2.I-123 to BIPM.RI(II)-K1.I-123

The value xi is the equivalent activity for laboratory i participant in EUROMET.RI(II)-K2.I-123 having been normalized to the value of the LNE-LNHB and IRMM as linking laboratories (see Final Report).

The degree of equivalence of laboratory i participant in EUROMET.RI(II)-K2.I-123 with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq.
The approximation Ui = 2(ui2 + uR2)1/2 is used in the Matrix of equivalence.

The degree of equivalence between two laboratories i and j, one participant in BIPM.RI(II)-K1.I-123 and one in EUROMET.RI(II)-K2.I-123, or both participants in EUROMET.RI(II)-K2.I-123, is given by a pair of terms expressed in MBq:
Dij = Di - Dj and Uij, its expanded uncertainty (k = 2), approximated by
Uij = 2(ui2 + uj2 - 2ful2)1/2 with l being the linking laboratory when each laboratory is from the EUROMET and f is the correlation coefficient.


These statements make it possible to extend the BIPM.RI(II)-K1.I-123 Matrix of equivalence to all participants in EUROMET.RI(II)-K2.I-123.
Comparison
Comparison close
Results

BIPM.RI(II)-K1.I-123 and EUROMET.RI(II)-K2.I-123

MEASURAND : Equivalent activity of 123I

Key comparison BIPM.RI(II)-K1.I-123

xi : result of measurement carried out in the SIR for the sample submitted by laboratory i
ui
: combined standard uncertainty of xi

 Lab i xi
/ kBq
ui
/ kBq
Date of measurement
 IRMM 120120 730 83-05-31
 LNE-LNHB 120020 470 83-05-31
 IRA 119450 570 85-06-27
 PTB 121400 1300 85-10-25

• Key comparison EUROMET.RI(II)-K2.I-123

xi : result of measurement carried out at laboratory i converted to the equivalent activity through the LNE-LNHB and the IRMM
ui : combined standard uncertainty of xi

 Lab i xi
/ kBq
ui
/ kBq
Measurement report date
 NPL 120480 440 1983