CCQM-K173

MEASURAND   Amount content of bases expressed as sodium carbonate (Na2CO3)

NOMINAL VALUE 10 mol/kg

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence (k = 2), both expressed in mol/kg.

Amount content of Na2CO3
Di / (mol/kg)
Participating institutes
Comparison
Comparison close
Results

CCQM-K173

MEASURAND   Amount content of bases expressed as sodium carbonate (Na2CO3)

NOMINAL VALUE 10 mol/kg

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence (k = 2), both expressed in mol/kg.

LABi Di Uneg,i Upos,i
mol/kg
mol/kg
mol/kg
INMETRO -0.00681 0.02590
SE "Ukrmetrteststandard" -0.00507 0.01722
NMIJ -0.00159 0.00265
SMU -0.00141 0.00242
NIM -0.00119 0.00250
UNIIM 0.00119 0.00284
CENAM 0.00185 0.00257
BAM 0.00189 0.00409
CMI 0.00595 0.00679
Participating institutes
CCQM-K173
Comparison
Comparison close
CC comparison
Linked comparison
CCQM-K173
Metrology area, Sub-field Chemistry and Biology, Electrochemistry
Description Assay of sodium carbonate
Time of measurements 2021
Status Approved for equivalence
Final Reports of the comparisons
References

CCQM-K173 Technical protocol

Measurand Amount content
Transfer device Solid sodium carbonate samples
Comparison type Key Comparison
Consultative Committee CCQM (Consultative Committee for Amount of Substance)
Conducted by CCQM (Consultative Committee for Amount of Substance)
Comments Results published on 25 January 2023
Pilot institute NIM
National Institute of Metrology
China
Contact person Wu Bing

86-010-64524751
Additional
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NIM

National Institute of Metrology, China, APMP

BAM

Bundesanstalt für Materialforschung und -prüfung, Germany, EURAMET

CENAM

Centro Nacional de Metrologia, Mexico, SIM

CMI

Czech Metrology Institute, Czechia, EURAMET

INMETRO

Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM

INTI

Instituto Nacional de Tecnologia Industrial, Argentina, SIM

NIST

National Institute of Standards and Technology, United States, SIM

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

SE "Ukrmetrteststandard"

State Enterprise "Ukrmetrteststandard", Ukraine, EURAMET

SMU

Slovensky Metrologicky Ustav, Slovakia, EURAMET

VNIIM

D.I. Mendeleyev Institute for Metrology, Rosstandart, Russian Federation, COOMET

Comparison
Comparison close
Results

CCQM-K173

MEASURAND   Amount content of bases expressed as sodium carbonate (Na2CO3)

NOMINAL VALUE 10 mol/kg

The key comparison reference value xR and its associated standard uncertainty uR are computed from the median of the results of the participants, excluding CMI, INMETRO and SE "Ukrmetrteststandard" (See further detailed information in section 8.1 of the Final Report)

         xR = 9.4320 mol/kg     uR = 0.00114 mol/kg

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in mol/kg:

 

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence where Ui = 2 (ui 2 + uR2)1/2

Comparison
Comparison close
Results

CCQM-K173

MEASURAND   Amount content of bases expressed as sodium carbonate (Na2CO3)

NOMINAL VALUE 10 mol/kg

 

xi   measurement value reported by laboratory iLabi

ui   standard uncertainty of xi

  Labi xi ui
    mol kg-1 mol kg-1
  INMETRO 9.4252 0.0129
  SE "Ukrmetrteststandard" 9.4269 0.0085
  NMIJ 9.43042 0.00068
  SMU 9.4306 0.00041
  NIM 9.43082 0.00051
  VNIIM 9.4332 0.00085
  CENAM 9.43386 0.00060
  BAM 9.4339 0.0017
  CMI 9.4380 0.0032