CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 300 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 320 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
This is the possibility to place a comment.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 340 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 360 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 380 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 400 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 450 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
This is the possibility to place a comment.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 500 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 550 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 600 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 650 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 700 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 750 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 800 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 850 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 900 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 950 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 1000 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
This is the possibility to place a comment.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 300 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0049 | 0.0080 | ||
| NMC, A*STAR | 0.0107 | 0.0197 | ||
| NIST | -0.0017 | 0.0080 | ||
| IO-CSIC | 0.0564 | 0.0159 | ||
| VNIIOFI | -0.0056 | 0.0107 | ||
| NMIJ | 0.0158 | 0.0198 | ||
| LNE-LCM/Cnam | 0.0006 | 0.0090 | ||
| PTB | -0.0040 | 0.0090 | ||
| NRC | 0.0001 | 0.0093 | ||
| MSL | 0.0144 | 0.0135 | ||
| UME | -0.0284 | 0.0149 | ||
| NPL | 0.0022 | 0.0110 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 320 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0051 | 0.0078 | ||
| NMC, A*STAR | 0.0086 | 0.0187 | ||
| NIST | -0.0050 | 0.0076 | ||
| IO-CSIC | 0.0522 | 0.0153 | ||
| VNIIOFI | 0.0068 | 0.0097 | ||
| NMIJ | 0.0191 | 0.0196 | ||
| LNE-LCM/Cnam | -0.0002 | 0.0089 | ||
| PTB | -0.0022 | 0.0085 | ||
| NRC | -0.0001 | 0.0087 | ||
| MSL | 0.0105 | 0.0133 | ||
| UME | -0.0129 | 0.0139 | ||
| NPL | 0.0041 | 0.0107 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 340 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0103 | 0.0063 | ||
| NMC, A*STAR | 0.0144 | 0.0187 | ||
| NIST | -0.0101 | 0.0059 | ||
| IO-CSIC | 0.0535 | 0.0144 | ||
| VNIIOFI | 0.0127 | 0.0076 | ||
| NMIJ | 0.0174 | 0.0187 | ||
| LNE-LCM/Cnam | -0.0028 | 0.0073 | ||
| PTB | -0.0037 | 0.0070 | ||
| NRC | 0.0020 | 0.0072 | ||
| MSL | 0.0074 | 0.0125 | ||
| UME | 0.0001 | 0.0126 | ||
| NPL | 0.0021 | 0.0094 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 360 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0137 | 0.0148 | ||
| NMC, A*STAR | 0.0064 | 0.0231 | ||
| NIST | -0.0098 | 0.0142 | ||
| IO-CSIC | 0.0479 | 0.0201 | ||
| VNIIOFI | 0.0242 | 0.0150 | ||
| NMIJ | 0.0135 | 0.0231 | ||
| LNE-LCM/Cnam | -0.0074 | 0.0151 | ||
| PTB | -0.0054 | 0.0148 | ||
| NRC | 0.0029 | 0.0151 | ||
| MSL | 0.0054 | 0.0180 | ||
| UME | -0.0025 | 0.0183 | ||
| NPL | -0.0011 | 0.0162 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 380 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0123 | 0.0062 | ||
| NMC, A*STAR | 0.0079 | 0.0175 | ||
| NIST | -0.0067 | 0.0048 | ||
| IO-CSIC | 0.0459 | 0.0130 | ||
| VNIIOFI | -0.0146 | 0.0070 | ||
| NMIJ | 0.0168 | 0.0132 | ||
| LNE-LCM/Cnam | -0.0001 | 0.0065 | ||
| PTB | -0.0021 | 0.0062 | ||
| NRC | -0.0021 | 0.0064 | ||
| MSL | 0.0067 | 0.0117 | ||
| UME | 0.0076 | 0.0074 | ||
| NPL | 0.0054 | 0.0068 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 400 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0082 | 0.0062 | ||
| NMC, A*STAR | 0.0126 | 0.0175 | ||
| NIST | -0.0070 | 0.0048 | ||
| IO-CSIC | 0.0301 | 0.0130 | ||
| VNIIOFI | -0.0073 | 0.0070 | ||
| NMIJ | 0.0150 | 0.0132 | ||
| LNE-LCM/Cnam | -0.0025 | 0.0065 | ||
| PTB | -0.0042 | 0.0062 | ||
| NRC | 0.0004 | 0.0064 | ||
| MSL | 0.0069 | 0.0117 | ||
| UME | 0.0073 | 0.0074 | ||
| NPL | 0.0002 | 0.0068 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 450 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0013 | 0.0053 | ||
| NMC, A*STAR | 0.0025 | 0.0067 | ||
| NIST | -0.0027 | 0.0055 | ||
| IO-CSIC | 0.0039 | 0.0089 | ||
| VNIIOFI | -0.0011 | 0.0064 | ||
| NMIJ | 0.0052 | 0.0126 | ||
| LNE-LCM/Cnam | -0.0010 | 0.0059 | ||
| PTB | -0.0028 | 0.0067 | ||
| NRC | 0.0015 | 0.0075 | ||
| MSL | 0.0026 | 0.0088 | ||
| UME | 0.0033 | 0.0082 | ||
| NPL | -0.0007 | 0.0064 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 500 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0012 | 0.0043 | ||
| NMC, A*STAR | -0.0006 | 0.0057 | ||
| NIST | -0.0014 | 0.0044 | ||
| IO-CSIC | -0.0001 | 0.0071 | ||
| VNIIOFI | 0.0004 | 0.0056 | ||
| NMIJ | 0.0022 | 0.0113 | ||
| LNE-LCM/Cnam | 0.0008 | 0.0055 | ||
| PTB | -0.0017 | 0.0059 | ||
| NRC | 0.0019 | 0.0063 | ||
| MSL | 0.0019 | 0.0071 | ||
| UME | 0.0027 | 0.0075 | ||
| NPL | -0.0001 | 0.0055 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 550 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0009 | 0.0042 | ||
| NMC, A*STAR | -0.0019 | 0.0054 | ||
| NIST | -0.0009 | 0.0042 | ||
| IO-CSIC | -0.0161 | 0.0076 | ||
| VNIIOFI | 0.0007 | 0.0055 | ||
| NMIJ | 0.0009 | 0.0103 | ||
| LNE-LCM/Cnam | 0.0017 | 0.0053 | ||
| PTB | -0.0016 | 0.0058 | ||
| NRC | 0.0017 | 0.0062 | ||
| MSL | 0.0010 | 0.0067 | ||
| UME | 0.0025 | 0.0074 | ||
| NPL | -0.0001 | 0.0053 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 600 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0010 | 0.0040 | ||
| NMC, A*STAR | -0.0007 | 0.0055 | ||
| NIST | 0.0005 | 0.0042 | ||
| IO-CSIC | -0.0143 | 0.0071 | ||
| VNIIOFI | 0.0021 | 0.0050 | ||
| NMIJ | 0.0017 | 0.0096 | ||
| LNE-LCM/Cnam | 0.0018 | 0.0059 | ||
| PTB | -0.0003 | 0.0058 | ||
| NRC | 0.0024 | 0.0062 | ||
| MSL | 0.0020 | 0.0068 | ||
| UME | 0.0031 | 0.0074 | ||
| NPL | 0.0011 | 0.0053 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 650 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | 0.0001 | 0.0044 | ||
| NMC, A*STAR | -0.0006 | 0.0055 | ||
| NIST | 0.0003 | 0.0043 | ||
| IO-CSIC | -0.0145 | 0.0071 | ||
| VNIIOFI | 0.0018 | 0.0051 | ||
| NMIJ | 0.0009 | 0.0089 | ||
| LNE-LCM/Cnam | 0.0027 | 0.0054 | ||
| PTB | -0.0004 | 0.0059 | ||
| NRC | 0.0020 | 0.0063 | ||
| MSL | 0.0011 | 0.0068 | ||
| UME | 0.0025 | 0.0075 | ||
| NPL | 0.0005 | 0.0054 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 700 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0002 | 0.0041 | ||
| NMC, A*STAR | -0.0003 | 0.0051 | ||
| NIST | 0.0003 | 0.0040 | ||
| IO-CSIC | -0.0139 | 0.0069 | ||
| VNIIOFI | 0.0017 | 0.0048 | ||
| NMIJ | 0.0003 | 0.0088 | ||
| LNE-LCM/Cnam | 0.0020 | 0.0048 | ||
| PTB | -0.0006 | 0.0056 | ||
| NRC | 0.0022 | 0.0060 | ||
| MSL | 0.0015 | 0.0067 | ||
| UME | 0.0020 | 0.0073 | ||
| NPL | 0.0009 | 0.0051 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 750 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0015 | 0.0048 | ||
| NMC, A*STAR | -0.0006 | 0.0054 | ||
| NIST | -0.0007 | 0.0044 | ||
| IO-CSIC | 0.0025 | 0.0070 | ||
| VNIIOFI | 0.0010 | 0.0051 | ||
| NMIJ | -0.0013 | 0.0093 | ||
| LNE-LCM/Cnam | 0.0015 | 0.0052 | ||
| PTB | -0.0017 | 0.0059 | ||
| NRC | 0.0013 | 0.0064 | ||
| MSL | 0.0002 | 0.0069 | ||
| UME | 0.0005 | 0.0075 | ||
| NPL | -0.0002 | 0.0054 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 800 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0002 | 0.0066 | ||
| NMC, A*STAR | -0.0003 | 0.0055 | ||
| NIST | -0.0012 | 0.0044 | ||
| IO-CSIC | 0.0044 | 0.0074 | ||
| VNIIOFI | 0.0005 | 0.0051 | ||
| NMIJ | -0.0025 | 0.0094 | ||
| LNE-LCM/Cnam | 0.0002 | 0.0054 | ||
| PTB | -0.0016 | 0.0059 | ||
| NRC | 0.0012 | 0.0066 | ||
| MSL | 0.0000 | 0.0070 | ||
| UME | 0.0004 | 0.0075 | ||
| NPL | -0.0001 | 0.0055 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 850 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0010 | 0.0059 | ||
| NMC, A*STAR | 0.0005 | 0.0054 | ||
| NIST | -0.0013 | 0.0044 | ||
| IO-CSIC | 0.0037 | 0.0091 | ||
| VNIIOFI | 0.0006 | 0.0051 | ||
| NMIJ | -0.0034 | 0.0095 | ||
| LNE-LCM/Cnam | 0.0004 | 0.0052 | ||
| PTB | -0.0018 | 0.0059 | ||
| NRC | 0.0024 | 0.0068 | ||
| MSL | 0.0004 | 0.0071 | ||
| UME | 0.0006 | 0.0075 | ||
| NPL | 0.0002 | 0.0054 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 900 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0038 | 0.0044 | ||
| NMC, A*STAR | 0.0015 | 0.0054 | ||
| NIST | -0.0009 | 0.0044 | ||
| IO-CSIC | 0.0047 | 0.0090 | ||
| VNIIOFI | 0.0019 | 0.0051 | ||
| NMIJ | -0.0021 | 0.0097 | ||
| LNE-LCM/Cnam | 0.0000 | 0.0050 | ||
| PTB | -0.0009 | 0.0059 | ||
| NRC | 0.0021 | 0.0064 | ||
| MSL | 0.0011 | 0.0072 | ||
| UME | 0.0004 | 0.0125 | ||
| NPL | 0.0003 | 0.0054 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 950 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0022 | 0.0039 | ||
| NMC, A*STAR | 0.0033 | 0.0156 | ||
| NIST | -0.0009 | 0.0040 | ||
| IO-CSIC | 0.0072 | 0.0092 | ||
| VNIIOFI | 0.0005 | 0.0059 | ||
| NMIJ | -0.0009 | 0.0115 | ||
| LNE-LCM/Cnam | 0.0012 | 0.0055 | ||
| PTB | -0.0011 | 0.0056 | ||
| NRC | 0.0009 | 0.0060 | ||
| MSL | 0.0005 | 0.0072 | ||
| UME | 0.0004 | 0.0123 | ||
| NPL | 0.0007 | 0.0060 |
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 1000 nm
Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.
| LABi | Di | Uneg,i | Upos,i | |
|---|---|---|---|---|
dimensionless |
dimensionless |
dimensionless |
||
| KRISS | -0.0034 | 0.0075 | ||
| NMC, A*STAR | 0.0036 | 0.0168 | ||
| NIST | -0.0060 | 0.0080 | ||
| IO-CSIC | 0.0071 | 0.0107 | ||
| VNIIOFI | 0.0142 | 0.0091 | ||
| NMIJ | -0.0011 | 0.0183 | ||
| LNE-LCM/Cnam | 0.0002 | 0.0096 | ||
| PTB | -0.0023 | 0.0085 | ||
| NRC | -0.0039 | 0.0091 | ||
| MSL | -0.0012 | 0.0100 | ||
| UME | 0.0015 | 0.0139 | ||
| NPL | -0.0016 | 0.0085 |
| Metrology area, Sub-field | Photometry and Radiometry, Properties of detectors and sources |
| Description | Spectral responsitivity |
| Time of measurements | 2016 - 2017 |
| Status | Approved for equivalence |
| Final Reports of the comparisons | |
| References | |
| Measurand | Spectral responsivity in A/W |
| Transfer device | Silicon trap detectors and Silicon photodiodes (2 types) |
| Comparison type | Key Comparison |
| Consultative Committee | CCPR (Consultative Committee for Photometry and Radiometry) |
| Conducted by | CCPR (Consultative Committee for Photometry and Radiometry) |
| Comments | This is a repeat of the CCPR-K2.b carried out 2000 - 2001 Results published 06 November 2026 |
| Pilot institute |
KRISS
Korea Research Institute of Standards and Science Korea, Republic of |
| Contact person | Dong-Hoon Lee +82 42 868 5706 |
| Pilot laboratory | |
|---|---|
| KRISS |
Korea Research Institute of Standards and Science, Korea, Republic of, APMP |
| IO-CSIC |
Instituto de Optica Daza de Valdés, Spain, EURAMET |
| LNE-LCM/Cnam |
Conservatoire National des Arts et Métiers/Laboratoire Commun de Métrologie, France, EURAMET |
| MSL |
Measurement Standards Laboratory, New Zealand, APMP |
| NIST |
National Institute of Standards and Technology, United States, SIM |
| NMC, A*STAR |
National Metrology Centre, Agency for Science, Technology and Research, Singapore, APMP |
| NMIJ AIST |
National Metrology Institute of Japan, Japan, APMP |
| NPL |
National Physical Laboratory, United Kingdom, EURAMET |
| NRC |
National Research Council, Canada, SIM |
| PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
| UME |
TÜBITAK Ulusal Metroloji Enstitüsü, Türkiye, EURAMET |
| VNIIOFI |
All-Russian Scientific Institute for Optical and Physical Measurements, Rosstandart, Russian Federation, COOMET |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 300 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0049 uR = 0.0017
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 320 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0051 uR = 0.0016
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 340 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0103 uR = 0.0014
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 360 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0137 uR = 0.0026
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 380 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0123 uR = 0.0012
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 400 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0082 uR = 0.0015
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 450 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0013 uR = 0.0010
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 500 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0012 uR = 0.0009
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 550 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0009 uR = 0.0009
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 600 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0010 uR = 0.0009
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 650 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0001 uR = 0.0009
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 700 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0002 uR = 0.0008
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 750 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0015 uR = 0.0009
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 800 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0002 uR = 0.0009
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 850 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0010 uR = 0.0009
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 900 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0038 uR = 0.0009
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 950 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0022 uR = 0.0009
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
WAVELENGTH 1000 nm
The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.
xR = 0.0034 uR = 0.0015
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:
Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.
CCPR-K2.b.2016
MEASURAND Spectral responsivity
TRANSFER STANDARD Single-element detector
The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.