CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   300 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

300 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   320 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

320 nm
Di 
Participating institutes

This is the possibility to place a comment.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   340 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

340 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   360 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

360 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   380 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

380 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   400 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

400 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   450 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

450 nm
Di 
Participating institutes

This is the possibility to place a comment.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   500 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

500 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   550 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

550 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   600 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

600 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   650 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

650 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   700 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

700 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   750 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

750 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   800 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

800 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   850 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

850 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   900 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

900 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   950 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

950 nm
Di 
Participating institutes

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   1000 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

1000 nm
Di 
Participating institutes

This is the possibility to place a comment.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   300 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0049 0.0080
NMC, A*STAR 0.0107 0.0197
NIST -0.0017 0.0080
IO-CSIC 0.0564 0.0159
VNIIOFI -0.0056 0.0107
NMIJ 0.0158 0.0198
LNE-LCM/Cnam 0.0006 0.0090
PTB -0.0040 0.0090
NRC 0.0001 0.0093
MSL 0.0144 0.0135
UME -0.0284 0.0149
NPL 0.0022 0.0110
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   320 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0051 0.0078
NMC, A*STAR 0.0086 0.0187
NIST -0.0050 0.0076
IO-CSIC 0.0522 0.0153
VNIIOFI 0.0068 0.0097
NMIJ 0.0191 0.0196
LNE-LCM/Cnam -0.0002 0.0089
PTB -0.0022 0.0085
NRC -0.0001 0.0087
MSL 0.0105 0.0133
UME -0.0129 0.0139
NPL 0.0041 0.0107
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   340 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0103 0.0063
NMC, A*STAR 0.0144 0.0187
NIST -0.0101 0.0059
IO-CSIC 0.0535 0.0144
VNIIOFI 0.0127 0.0076
NMIJ 0.0174 0.0187
LNE-LCM/Cnam -0.0028 0.0073
PTB -0.0037 0.0070
NRC 0.0020 0.0072
MSL 0.0074 0.0125
UME 0.0001 0.0126
NPL 0.0021 0.0094
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   360 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0137 0.0148
NMC, A*STAR 0.0064 0.0231
NIST -0.0098 0.0142
IO-CSIC 0.0479 0.0201
VNIIOFI 0.0242 0.0150
NMIJ 0.0135 0.0231
LNE-LCM/Cnam -0.0074 0.0151
PTB -0.0054 0.0148
NRC 0.0029 0.0151
MSL 0.0054 0.0180
UME -0.0025 0.0183
NPL -0.0011 0.0162
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   380 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0123 0.0062
NMC, A*STAR 0.0079 0.0175
NIST -0.0067 0.0048
IO-CSIC 0.0459 0.0130
VNIIOFI -0.0146 0.0070
NMIJ 0.0168 0.0132
LNE-LCM/Cnam -0.0001 0.0065
PTB -0.0021 0.0062
NRC -0.0021 0.0064
MSL 0.0067 0.0117
UME 0.0076 0.0074
NPL 0.0054 0.0068
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   400 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0082 0.0062
NMC, A*STAR 0.0126 0.0175
NIST -0.0070 0.0048
IO-CSIC 0.0301 0.0130
VNIIOFI -0.0073 0.0070
NMIJ 0.0150 0.0132
LNE-LCM/Cnam -0.0025 0.0065
PTB -0.0042 0.0062
NRC 0.0004 0.0064
MSL 0.0069 0.0117
UME 0.0073 0.0074
NPL 0.0002 0.0068
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   450 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0013 0.0053
NMC, A*STAR 0.0025 0.0067
NIST -0.0027 0.0055
IO-CSIC 0.0039 0.0089
VNIIOFI -0.0011 0.0064
NMIJ 0.0052 0.0126
LNE-LCM/Cnam -0.0010 0.0059
PTB -0.0028 0.0067
NRC 0.0015 0.0075
MSL 0.0026 0.0088
UME 0.0033 0.0082
NPL -0.0007 0.0064
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   500 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0012 0.0043
NMC, A*STAR -0.0006 0.0057
NIST -0.0014 0.0044
IO-CSIC -0.0001 0.0071
VNIIOFI 0.0004 0.0056
NMIJ 0.0022 0.0113
LNE-LCM/Cnam 0.0008 0.0055
PTB -0.0017 0.0059
NRC 0.0019 0.0063
MSL 0.0019 0.0071
UME 0.0027 0.0075
NPL -0.0001 0.0055
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   550 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0009 0.0042
NMC, A*STAR -0.0019 0.0054
NIST -0.0009 0.0042
IO-CSIC -0.0161 0.0076
VNIIOFI 0.0007 0.0055
NMIJ 0.0009 0.0103
LNE-LCM/Cnam 0.0017 0.0053
PTB -0.0016 0.0058
NRC 0.0017 0.0062
MSL 0.0010 0.0067
UME 0.0025 0.0074
NPL -0.0001 0.0053
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   600 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0010 0.0040
NMC, A*STAR -0.0007 0.0055
NIST 0.0005 0.0042
IO-CSIC -0.0143 0.0071
VNIIOFI 0.0021 0.0050
NMIJ 0.0017 0.0096
LNE-LCM/Cnam 0.0018 0.0059
PTB -0.0003 0.0058
NRC 0.0024 0.0062
MSL 0.0020 0.0068
UME 0.0031 0.0074
NPL 0.0011 0.0053
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   650 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS 0.0001 0.0044
NMC, A*STAR -0.0006 0.0055
NIST 0.0003 0.0043
IO-CSIC -0.0145 0.0071
VNIIOFI 0.0018 0.0051
NMIJ 0.0009 0.0089
LNE-LCM/Cnam 0.0027 0.0054
PTB -0.0004 0.0059
NRC 0.0020 0.0063
MSL 0.0011 0.0068
UME 0.0025 0.0075
NPL 0.0005 0.0054
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   700 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0002 0.0041
NMC, A*STAR -0.0003 0.0051
NIST 0.0003 0.0040
IO-CSIC -0.0139 0.0069
VNIIOFI 0.0017 0.0048
NMIJ 0.0003 0.0088
LNE-LCM/Cnam 0.0020 0.0048
PTB -0.0006 0.0056
NRC 0.0022 0.0060
MSL 0.0015 0.0067
UME 0.0020 0.0073
NPL 0.0009 0.0051
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   750 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0015 0.0048
NMC, A*STAR -0.0006 0.0054
NIST -0.0007 0.0044
IO-CSIC 0.0025 0.0070
VNIIOFI 0.0010 0.0051
NMIJ -0.0013 0.0093
LNE-LCM/Cnam 0.0015 0.0052
PTB -0.0017 0.0059
NRC 0.0013 0.0064
MSL 0.0002 0.0069
UME 0.0005 0.0075
NPL -0.0002 0.0054
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   800 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0002 0.0066
NMC, A*STAR -0.0003 0.0055
NIST -0.0012 0.0044
IO-CSIC 0.0044 0.0074
VNIIOFI 0.0005 0.0051
NMIJ -0.0025 0.0094
LNE-LCM/Cnam 0.0002 0.0054
PTB -0.0016 0.0059
NRC 0.0012 0.0066
MSL 0.0000 0.0070
UME 0.0004 0.0075
NPL -0.0001 0.0055
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   850 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0010 0.0059
NMC, A*STAR 0.0005 0.0054
NIST -0.0013 0.0044
IO-CSIC 0.0037 0.0091
VNIIOFI 0.0006 0.0051
NMIJ -0.0034 0.0095
LNE-LCM/Cnam 0.0004 0.0052
PTB -0.0018 0.0059
NRC 0.0024 0.0068
MSL 0.0004 0.0071
UME 0.0006 0.0075
NPL 0.0002 0.0054
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   900 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0038 0.0044
NMC, A*STAR 0.0015 0.0054
NIST -0.0009 0.0044
IO-CSIC 0.0047 0.0090
VNIIOFI 0.0019 0.0051
NMIJ -0.0021 0.0097
LNE-LCM/Cnam 0.0000 0.0050
PTB -0.0009 0.0059
NRC 0.0021 0.0064
MSL 0.0011 0.0072
UME 0.0004 0.0125
NPL 0.0003 0.0054
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   950 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0022 0.0039
NMC, A*STAR 0.0033 0.0156
NIST -0.0009 0.0040
IO-CSIC 0.0072 0.0092
VNIIOFI 0.0005 0.0059
NMIJ -0.0009 0.0115
LNE-LCM/Cnam 0.0012 0.0055
PTB -0.0011 0.0056
NRC 0.0009 0.0060
MSL 0.0005 0.0072
UME 0.0004 0.0123
NPL 0.0007 0.0060
Participating institutes
CCPR-K2.b.2016

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   1000 nm

 

Degrees of equivalence represented by Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both dimensionless.

LABi Di Uneg,i Upos,i
dimensionless
dimensionless
dimensionless
KRISS -0.0034 0.0075
NMC, A*STAR 0.0036 0.0168
NIST -0.0060 0.0080
IO-CSIC 0.0071 0.0107
VNIIOFI 0.0142 0.0091
NMIJ -0.0011 0.0183
LNE-LCM/Cnam 0.0002 0.0096
PTB -0.0023 0.0085
NRC -0.0039 0.0091
MSL -0.0012 0.0100
UME 0.0015 0.0139
NPL -0.0016 0.0085
Participating institutes
CCPR-K2.b.2016
Comparison
Comparison close
CC comparison
Linked comparison
CCPR-K2.b.2016
Metrology area, Sub-field Photometry and Radiometry, Properties of detectors and sources
Description Spectral responsitivity
Time of measurements 2016 - 2017
Status Approved for equivalence
Final Reports of the comparisons
References

CCPR-K2.b.2016 Technical Protocol

CCPR-K2.b.2016 Registration and progress form

Measurand Spectral responsivity in A/W
Transfer device Silicon trap detectors and Silicon photodiodes (2 types)
Comparison type Key Comparison
Consultative Committee CCPR (Consultative Committee for Photometry and Radiometry)
Conducted by CCPR (Consultative Committee for Photometry and Radiometry)
Comments This is a repeat of the CCPR-K2.b carried out 2000 - 2001

Results published 06 November 2026
 
Pilot institute KRISS
Korea Research Institute of Standards and Science
Korea, Republic of
Contact person Dong-Hoon Lee

+82 42 868 5706
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
KRISS

Korea Research Institute of Standards and Science, Korea, Republic of, APMP

IO-CSIC

Instituto de Optica Daza de Valdés, Spain, EURAMET

LNE-LCM/Cnam

Conservatoire National des Arts et Métiers/Laboratoire Commun de Métrologie, France, EURAMET

MSL

Measurement Standards Laboratory, New Zealand, APMP

NIST

National Institute of Standards and Technology, United States, SIM

NMC, A*STAR

National Metrology Centre, Agency for Science, Technology and Research, Singapore, APMP

NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

NPL

National Physical Laboratory, United Kingdom, EURAMET

NRC

National Research Council, Canada, SIM

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

UME

TÜBITAK Ulusal Metroloji Enstitüsü, Türkiye, EURAMET

VNIIOFI

All-Russian Scientific Institute for Optical and Physical Measurements, Rosstandart, Russian Federation, COOMET

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   300 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0049     uR = 0.0017

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

 

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   320 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0051     uR = 0.0016

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   340 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0103     uR = 0.0014

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   360 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0137     uR = 0.0026

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   380 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0123     uR = 0.0012

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   400 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0082     uR = 0.0015

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   450 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0013     uR = 0.0010

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   500 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0012     uR = 0.0009

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   550 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0009     uR = 0.0009

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   600 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0010     uR = 0.0009

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   650 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0001     uR = 0.0009

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   700 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0002     uR = 0.0008

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   750 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0015     uR = 0.0009

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   800 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0002     uR = 0.0009

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   850 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0010     uR = 0.0009

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   900 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0038     uR = 0.0009

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   950 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0022     uR = 0.0009

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

WAVELENGTH   1000 nm

 

The key comparison reference value xR and its associated standard uncertainty uR are computed from the weighted mean of the relative differences defined from  participants' measurement results with exclusions, as described in Section 6 of the CCPR-K2.b.2016 Final report.

 

        xR = 0.0034     uR = 0.0015

 

The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both dimensionless:

Di = (xi - xR) and its expanded uncertainty Ui at a 95 % level of confidence, calculated as described in Section 6 of the CCPR-K2.b.2016 Final report .

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.

CCPR-K2.b.2016

 

MEASURAND   Spectral responsivity

TRANSFER STANDARD Single-element detector

 

The measurement data of participants that was used to process the comparison results is listed in Section 4 of the CCPR-K2.b.2016 Final report.