Comparison
Comparison close
Results

BIPM.RI(II)-K1.Am-241, 2003 CCRI(II)-K2.Am-241,
and 2006 COOMET.RI(II)-K2.Am-241

MEASURAND : SIR equivalent activity of 241Am
Degrees of equivalence Di and expanded uncertainty Ui (k = 2) expressed in MBq

Red squares: participants in BIPM.RI(II)-K1.Am-241
Blue circles: participants in CCRI(II)-K2.Am-241 (2003)
Green triangles: participants in COOMET.RI(II)-K2.Am-241 (2006)

Note: the right-hand axis shows approximate relative values only

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Am-241, 2003 CCRI(II)-K2.Am-241,
and 2006 COOMET.RI(II)-K2.Am-241

MEASURAND : SIR equivalent activity of 241Am

 

 

participants in

BIPM.RI(II)-K2.Am-241
participants in CCRI(II)-K2.Am-241
participants in COOMET.RI(II)-K2.Am-241

Results are presented under A4 printable format in Summary Results (.PDF file).

COOMET.RI(II)-K2.Am-241
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity concentration of the same Am-241 solution
Time of measurements 2006
Status Approved for equivalence
Final Reports of the comparisons
Measurand Activity concentration of the same solution
Nominal value: 500 kBq/g

Transfer device Radioactive solution
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by COOMET (Euro-Asian Cooperation of National Metrological Institutions)
RMO Internal Identifier COOMET 359/RU/06
Comments

Results published on 01 November 2007

COOMET.RI(II)-K2.Am-241 results are linked to those of BIPM.RI(II)-K1.Am-241

SIR equivalent activity of the radionuclide 241Am:

Pilot institute VNIIM
D.I. Mendeleyev Institute for Metrology, Rosstandart
Russian Federation
Contact person I.A. Kharitonov

+7 812 323 96 11
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
VNIIM

D.I. Mendeleyev Institute for Metrology, Rosstandart, Russian Federation, COOMET

BelGIM

Belarussian State Institute for Metrology, Belarus, COOMET

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Am-241, CCRI(II)-K2.Am-241,
and COOMET.RI(II)-K2.Am-241

MEASURAND : SIR equivalent activity of 241Am

  • Key comparison BIPM.RI(II)-K1.Am-241
Key comparison reference value:
The SIR reference value for this radionuclide is xR = 2055.8 MBq, with a standard uncertainty uR = 2.8 MBq, see section 4.1 of the
Final Report dated November 2007.

 

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq. With n the number of laboratories, when each laboratory has contributed to the computation of xR.

 

  • Linking CCRI(II)-K2.Am-241 (2003) to BIPM.RI(II)-K1.Am-241

The value xi is the equivalent activity for laboratory i participant in CCRI(II)-K2.Am-241 having been normalized using the SIR measurement of the NPL undiluted solution of the CCRI(II)-K2.Am-241 comparison.

 

The degree of equivalence of laboratory i participant in CCRI(II)-K2.Am-241 with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq.
The approximation Ui = 2(ui2 + uR2)1/2 is used in the Matrix of equivalence as none of these laboratories contributed to the computation of xR.

 

  • Linking COOMET.RI(II)-K2.Am-241 to BIPM.RI(II)-K1.Am-241

The value xi is the equivalent activity for laboratory i participant in COOMET.RI(II)-K2.Am-241 having been normalized to the SIR measurement of the VNIIM solution of the COOMET.RI(II)-K2.Am-241 comparison.

 

The degree of equivalence of laboratory i participant in COOMET.RI(II)-K2.Am-241 with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in MBq.
The approximation Ui = 2(ui2 + uR2)1/2 is used in the Matrix of equivalence as none of these laboratories contributed to the computation of xR.

 

These statements make it possible to extend the BIPM.RI(II)-K1.Am-241 Matrix of equivalence to all participants in the CCRI(II)-K2.Am-241 and the COOMET.RI(II)-K2.Am-241 comparisons.

Comparison
Comparison close
Results

BIPM.RI(II)-K1.Am-241, 2003 CCRI(II)-K2.Am-241,
and 2006 COOMET.RI(II)-K2.Am-241

MEASURAND : SIR equivalent activity of 241Am

  • Key comparison BIPM.RI(II)-K1.Am-241

xi : result of measurement carried out in the SIR for the sample submitted by laboratory i
ui
: combined standard uncertainty of xi

 

  • Key comparison CCRI(II)-K2.Am-241

xi : result of measurement carried out at laboratory i linked to the SIR through the NPL mother solution
ui : combined standard uncertainty of xi

  • Key comparison COOMET.RI(II)-K2.Am-241

xi : result of measurement carried out at laboratory i converted to the equivalent activity through the VNIIM
ui : combined standard uncertainty of xi