BIPM.RI(II)-K1.Ba-133 and APMP.RI(II)-K2.Ba-133
MEASURAND SIR equivalent activity of Ba-133
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.
Results given in absolute values.
BIPM.RI(II)-K1.Ba-133 and APMP.RI(II)-K2.Ba-133
MEASURAND SIR equivalent activity of Ba-133
Degrees of equivalence represented by Di = (xI - xR) / xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.
Results given in relative values.
BIPM.RI(II)-K1.Ba-133 and APMP.RI(II)-K2.Ba-133
MEASURAND SIR equivalent activity of Ba-133
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in MBq.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
MBq |
MBq |
MBq |
||
IFIN-HH | -0.31 | 0.29 | ||
NMIJ | 0.13 | 0.22 | ||
IRA | 0.02 | 0.23 | ||
LNE-LNHB | -0.02 | 0.38 | ||
BEV | 0.16 | 0.61 | ||
NRC | -0.59 | 0.52 | ||
TENMAK-NÜKEN | 0.0 | 1.4 | ||
NMISA | 0.22 | 0.32 | ||
NIST | -0.02 | 0.52 | ||
ANSTO | -0.01 | 0.38 | ||
BARC | -1.5 | 1.2 | ||
INER | 0.12 | 0.30 | ||
KRISS | 0.20 | 0.32 | ||
NIM | 0.35 | 0.59 | ||
OAP | -0.1 | 1.5 | ||
PTKMR | 1.3 | 1.4 |
Results given in absolute values.
BIPM.RI(II)-K1.Ba-133 and APMP.RI(II)-K2.Ba-133
MEASURAND SIR equivalent activity of Ba-133
Degrees of equivalence represented by Di = (xI - xR) / xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
kBq/MBq |
kBq/MBq |
kBq/MBq |
||
IFIN-HH | -7.0 | 6.5 | ||
NMIJ | 3.0 | 5.0 | ||
IRA | 0.5 | 5.3 | ||
LNE-LNHB | -0.4 | 8.6 | ||
BEV | 4 | 14 | ||
NRC | -13 | 12 | ||
TENMAK-NÜKEN | 0 | 32 | ||
NMISA | 5.0 | 7.4 | ||
NIST | 0 | 12 | ||
ANSTO | -0.2 | 8.6 | ||
BARC | -35 | 27 | ||
INER | 2.8 | 6.9 | ||
KRISS | 4.6 | 7.3 | ||
NIM | 8 | 13 | ||
OAP | -3 | 35 | ||
PTKMR | 30 | 33 |
Results given in relative values.
Metrology area, Sub-field | Ionizing Radiation, Section II (Measurement of Radionuclides) |
Description | Activity of radionuclide Ba-133 |
Time of measurements | 2006 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
Measurand | Activity concentration in kBq/g |
Transfer device | Radioactive solution |
Comparison type | Key Comparison |
Consultative Committee | CCRI (Consultative Committee for Ionizing Radiation) |
Conducted by | APMP (Asia Pacific Metrology Programme) |
Comments | Results published on 27 October 2009 APMP.RI(II)-K2.Ba-133 results are linked to those of BIPM.RI(II)-K1.Ba-133 SIR equivalent activity of the radionuclide 133Ba
|
Pilot institute |
NMIJ AIST
National Metrology Institute of Japan Japan |
Contact person | Y. Hino +81 298 61 5667 |
Pilot laboratory | |
---|---|
NMIJ AIST |
National Metrology Institute of Japan, Japan, APMP |
ANSTO |
Australian Nuclear Science and Technology Organisation, Australia, APMP |
BARC |
Bhabha Atomic Research Centre, India, APMP |
KRISS |
Korea Research Institute of Standards and Science, Korea, Republic of, APMP |
NIM |
National Institute of Metrology, China, APMP |
NMISA |
National Metrology Institute of South Africa, South Africa, AFRIMETS |
NRSL/INER |
Institute of Nuclear Energy Research, Chinese Taipei, APMP |
OAP |
Office of Atoms for Peace, Thailand, APMP |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
BIPM.RI(II)-K1.Ba-133 and APMP.RI(II)-K2.Ba-133
MEASURAND SIR equivalent activity of Ba-133
The key comparison reference value xR for this radionuclide is estimated to be
xR = 43 899 kBq with a standard uncertainty uR = 59 kBq
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence are both expressed in MBq, where xi is the equivalent activity for laboratory i.
The Degrees of equivalence are also given in relative terms by Di = (xI - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.
BIPM.RI(II)-K1.Ba-133 and APMP.RI(II)-K2.Ba-133
MEASURAND SIR equivalent activity of Ba-133
xi measurement value submitted by laboratory i, Labi
ui combined standard uncertainty of xi
BIPM.RI(II)-K1.Ba-133
Labi | xi | u(xi) | Date of measurement | |
kBq | kBq | |||
IFIN-HH | 43590 | 140 | 2005-12-08 | |
NMIJ | 44030 | 100 | 2006-07-07 | |
IRA | 43920 | 100 | 2009-04-15 | |
LNE-LNHB | 43880 | 190 | 2012-03-07 | |
BEV | 44060 | 300 | 2012-06-12 | |
NRC | 43310 | 260 | 2016-01-28 | |
TENMAK-NÜKEN | 43920 | 700 | 2018-01-10 | |
NMISA | 44120 | 160 | 2019-03-12 | |
NIST | 43880 | 260 | 2019-04-15 | |
APMP.RI(II)-K2.Ba-133
Labi | xi | u(xi) | Date of measurement | |
kBq | kBq | |||
ANSTO | 43890 | 180 | 2006 | |
BARC | 42370 | 590 | 2006 | |
INER | 44020 | 140 | 2006 | |
KRISS | 44100 | 150 | 2006 | |
NIM | 44250 | 290 | 2006 | |
OAP | 43750 | 770 | 2006 | |
PTKMR | 45220 | 720 | 2006 | |