BIPM.RI(I)-K1.Co-56

MEASURAND      Activity concentration of Co-56

There is no graphical representation of equivalence, please see the table of degrees of equivalence for CMI-IIR represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, in kBq.

 

BIPM.RI(I)-K1.Co-56

MEASURAND      Activity concentration of Co-56

There is no graphical representation of equivalence, please see the table of degrees of equivalence for CMI-IIR represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, in kBq/MBq.

 

Comparison
Comparison close
Results

BIPM.RI(I)-K1.Co-56

MEASURAND      Activity concentration of Co-56

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq.

 

 

Results given in absolute values.

BIPM.RI(I)-K1.Co-56

MEASURAND      Activity concentration of Co-56

 

Degrees of equivalence represented by Di = (xI - xR)/xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

 

 

 

 

Results given in relative values.

Comparison
Comparison close
BIPM comparison
Linked comparison
BIPM.RI(II)-K1.Co-56
Metrology area, Sub-field Ionizing Radiation, Section II (Measurement of Radionuclides)
Description Activity of radionuclide Co-56
Time of measurements 1991 -
Status Continuous, approved for equivalence
Final Reports of the comparisons
Measurand SIR equivalent activity
Transfer device Radioactive solution
Comparison type Key Comparison
Consultative Committee CCRI (Consultative Committee for Ionizing Radiation)
Conducted by BIPM (Bureau International des Poids et Mesures)
Comments SIR: Système International de Référence, International Reference System
Results published on 14 November 2002, then updated on 30 March 2010

Results older than 20 years are no longer published in the KCDB (see pages 37 and 38 of the CCRI Report of its 18th meeting in 2003)
Pilot institute BIPM
Bureau International des Poids et Mesures
BIPM - International Organization
Contact person Carine Michotte

+33 (0) 1 45 07 70 85
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BIPM

Bureau International des Poids et Mesures, BIPM - International Organization, N/A

BNM-LNHB

Bureau National de Métrologie - Laboratoire National Henri Becquerel, France, EURAMET

CMI-IIR

Czech Metrology Institute - Inspectorate for Ionizing Radiation, Czechia, EURAMET

NPL

National Physical Laboratory, United Kingdom, EURAMET

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

Comparison
Comparison close
Results

BIPM.RI(I)-K1.Co-56

MEASURAND      Activity concentration of Co-56

 

Key comparison reference value:
 

The SIR reference value for this radionuclide is xR = 5066 kBq, with a standard uncertainty uR = 14 kBq.
 

(see section 4.1 of the Final Report dated 23 March 2010)

 

The degree of equivalence of each laboratory i with respect to the reference value is given by a pair of terms:

Di = (x- xR) and Uits expanded uncertainty (k = 2), both expressed in kBq, and

Ui = 2((1 - 2wi)ui2 + u2R)1/2, where wi is the weight of laboratory i contributing to the calculation of xR.

The degrees of equivalence are also given in relative terms by Di = (xI - xR) /xR and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in kBq/MBq.

Comparison
Comparison close
Results

BIPM.RI(I)-K1.Co-56

MEASURAND      Activity concentration of Co-56

 

xi   result of measurement carried out in the SIR for the sample submitted by laboratory i, Labi

ui   combined standard uncertainty of xi

  Labi xi ui Date of measurement
    kBq kBq  
  CMI-IIR 5029 28 2006-05-04