CCRI(II)-K2.Fe-55.2019
MEASURAND Activity concentration of a solution
NOMINAL VALUE 100 kBq/g
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in (kBq/g).
CCRI(II)-K2.Fe-55.2019
MEASURAND Activity concentration of a solution
NOMINAL VALUE 100 kBq/g
Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in (kBq/g).
LABi | Di | Uneg,i | Upos,i | |
---|---|---|---|---|
kBq/g |
kBq/g |
kBq/g |
||
BIPM | -1.7 | 1.2 | ||
CIEMAT | -0.3 | 1.9 | ||
ENEA-INMRI | -1.4 | 1.4 | ||
KRISS | -3.0 | 1.6 | ||
LNE-LNHB | -0.3 | 0.8 | ||
NIM | -1.2 | 0.9 | ||
NMIJ | -1.1 | 1.2 | ||
NMISA | 0.7 | 1.3 | ||
NRC | 0.6 | 3.2 | ||
POLATOM | -0.2 | 0.8 | ||
PTB | 0.2 | 1.0 | ||
SMU | 0.4 | 0.9 |
Metrology area, Sub-field | Ionizing Radiation, Section II (Measurement of Radionuclides) |
Description | Activity per mass of Fe-55 |
Time of measurements | 2019 |
Status | Approved for equivalence |
Final Reports of the comparisons | |
References | |
Measurand | The activity per mass in the master solution |
Transfer device | Glass ampoules |
Comparison type | Key Comparison |
Consultative Committee | CCRI (Consultative Committee for Ionizing Radiation) |
Conducted by | CCRI (Consultative Committee for Ionizing Radiation) |
Pilot institute |
POLATOM
Radioisotope Centre Polatom Poland |
Contact person | Ryszard Broda +48 22 273 19 53 |
Additional |
Pilot laboratory | |
---|---|
POLATOM |
Radioisotope Centre Polatom, Poland, EURAMET |
BIPM |
Bureau International des Poids et Mesures, BIPM - International Organization, N/A |
CIEMAT |
Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas, Spain, EURAMET |
ENEA-INMRI |
Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, Italy, EURAMET |
KRISS |
Korea Research Institute of Standards and Science, Korea, Republic of, APMP |
LNE-LNHB |
Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET |
NIM |
National Institute of Metrology, China, APMP |
NMIJ AIST |
National Metrology Institute of Japan, Japan, APMP |
NMISA |
National Metrology Institute of South Africa, South Africa, AFRIMETS |
NRC |
National Research Council, Canada, SIM |
PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
SMU |
Slovensky Metrologicky Ustav, Slovakia, EURAMET |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
CCRI(II)-K2.Fe-55.2019
MEASURAND Activity concentration of a solution
NOMINAL VALUE 100 kBq/g
The key comparison reference value xR and its associated standard uncertainty uR are computed from a power moderated mean of the reported results, see CCRI(II)-K2.Fe-55 Final Report where
xR = 101.49 kBq/g uR = 0.23 kBq/g
The degree of equivalence of laboratory i relative to the key comparison reference value xR is given by a pair of terms, both expressed in kBq/g.
CCRI(II)-K2.Fe-55.2019
MEASURAND Activity concentration of a solution
NOMINAL VALUE 100 kBq/g
xi measured activity concentration reported by laboratory i, Labi
ui combined standard uncertainty of xi