MEASURAND : Equivalent activity of 207Bi
Degrees of equivalence relative to the BIPM.RI(II)-K1.Bi-207 key comparison reference value
.bi-207_g_apr15.gif)
MEASURAND : Equivalent activity of 207Bi
Degrees of equivalence relative to the BIPM.RI(II)-K1.Bi-207 key comparison reference value
.bi-207_doe_apr15.gif)
Results are presented under A4 printable format in Summary Results (PDF file).
| Metrology area, Sub-field | Ionizing Radiation, Section II (Measurement of Radionuclides) |
| Description | Activity of radionuclide Bi-207 |
| Time of measurements | 1982 - |
| Status | Continuous, approved for equivalence |
| Final Reports of the comparisons | |
| Measurand | SIR equivalent activity |
| Transfer device | Radioactive solution |
| Comparison type | Key Comparison |
| Consultative Committee | CCRI (Consultative Committee for Ionizing Radiation) |
| Conducted by | BIPM (Bureau International des Poids et Mesures) |
| Comments | SIR: Système International de Référence, International Reference System Results published on 14 November 2002 SIR equivalent activity of the radionuclide 207Bi: |
| Pilot institute |
BIPM
Bureau International des Poids et Mesures BIPM - International Organization |
| Contact person | Carine Michotte +33 (0) 1 45 07 70 85 |
| Pilot laboratory | |
|---|---|
| BIPM |
Bureau International des Poids et Mesures, BIPM - International Organization, N/A |
| LNE-LNHB |
Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel, France, EURAMET |
| 2010 |
| PTB |
Physikalisch-Technische Bundesanstalt, Germany, EURAMET |
| 1982 |
| VNIIM |
D.I. Mendeleyev Institute for Metrology, Rosstandart, Russian Federation, COOMET |
| 1991 |
This page proposes print-out on A4 paper (portrait) of the comparison details (best printed out using a black and white printer).
Please, select items to be printed out, then click on "OK" :
MEASURAND : Equivalent activity of 207Bi
|
Key comparison reference value: |
| The degree of equivalence of each laboratory with respect to the reference value is given by a pair of terms: Di = (xi - xR) and Ui, its expanded uncertainty (k = 2), both expressed in kBq, and Ui = 2((1 - 2wi)ui2 + uR2)1/2 when each laboratory has contributed to the calculation of xR with a weight wi. |
MEASURAND : Equivalent activity of 207Bi
xi : result of measurement carried out in the SIR for the sample submitted by laboratory i
ui : combined standard uncertainty of xi
|
Lab i |
xi |
ui |
Date of measurement |
| LNE-LNHB | 10 889 | 55 | 2010-03-30 |