CCAUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       10 Hz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

10 Hz
Di
Participating institutes

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       40 Hz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

40 Hz
Di
Participating institutes

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       80 Hz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

80 Hz
Di
Participating institutes

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       160 Hz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

160 Hz
Di
Participating institutes

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       800 Hz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

800 Hz
Di
Participating institutes

CCAUV.V-K2 and APMP.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       2 kHz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

2 kHz
Di
Participating institutes

CCAUV.V-K2 and APMP.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       5 kHz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

5 kHz
Di
Participating institutes

CCAUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       10 kHz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

10 kHz
Di
Participating institutes

DEVICE : Single-Ended Accelerometer
MEASURAND : Phase
FREQUENCIES : 10 Hz, 40 Hz, 80 Hz, 160 Hz, 800 Hz, 2 kHz, 5 kHz and 10 kHz

Graphs of equivalence for the eight frequency values

DEVICE : Back-to-Back Accelerometer
MEASURAND : Phase
FREQUENCIES : 10 Hz, 40 Hz, 80 Hz, 160 Hz, 800 Hz, 2 kHz, 5 kHz and 10 kHz

Graphs of equivalence for the eight frequency values

CCAUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       10 Hz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

LABi Di Uneg,i Upos,i
10-4 pC/(m/s2)
10-4 pC/(m/s2)
10-4 pC/(m/s2)
PTB -0.5 0.9
CMI -1.9 2.4
NMISA 2.4 6.3
DPLA 0.2 5.0
CEM -0.1 5.0
GUM 2.1 7.6
METAS 3.5 2.7
NMIJ 4.7 16.9
KRISS -7.9 5.0
LNE 0.3 3.7
NIM -0.2 5.0
CENAM -0.5 3.7
INMETRO 0.6 2.9
UME 1.9 6.3
VNIIM -3.0 12.7
BIM 8.4 10.1
Participating institutes
CCAUV.V-K2
EURAMET.AUV.V-K2

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       40 Hz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

LABi Di Uneg,i Upos,i
10-4 pC/(m/s2)
10-4 pC/(m/s2)
10-4 pC/(m/s2)
PTB -0.6 1.0
CMI -0.1 2.4
NMISA 2.1 6.3
DPLA -1.2 5.0
CEM 0.0 5.0
GUM 1.7 6.3
METAS 1.8 2.3
NMIJ 0.0 4.9
KRISS -5.6 4.7
LNE -0.3 3.7
NIM 1.4 5.0
CENAM -0.2 3.7
INMETRO 0.6 3.0
UME -1.0 6.3
VNIIM 0.2 6.3
CMS 2.9 3.9
NIMT -4.7 7.8
A*STAR -3.7 5.9
BIM 0.7 9.0
Participating institutes
CCAUV.V-K2
APMP.AUV.V-K2
EURAMET.AUV.V-K2

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       80 Hz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

LABi Di Uneg,i Upos,i
10-4 pC/(m/s2)
10-4 pC/(m/s2)
10-4 pC/(m/s2)
PTB -0.9 1.0
CMI 0.1 2.4
NMISA 2.6 6.3
DPLA -0.7 5.0
CEM 0.1 5.0
GUM 1.5 6.3
METAS 1.5 2.3
NMIJ -0.1 4.9
KRISS -3.5 4.7
LNE 0.2 3.7
NIM 2.1 5.0
CENAM -0.3 3.7
INMETRO 0.8 3.0
UME 0.1 6.3
VNIIM 0.7 6.3
CMS 2.0 3.2
NIMT -2.8 5.5
A*STAR -3.2 6.2
BIM 2.5 9.0
Participating institutes
CCAUV.V-K2
APMP.AUV.V-K2
EURAMET.AUV.V-K2

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       160 Hz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

LABi Di Uneg,i Upos,i
10-4 pC/(m/s2)
10-4 pC/(m/s2)
10-4 pC/(m/s2)
PTB -0.4 1.0
CMI 0.1 2.4
NMISA 0.6 6.3
DPLA -1.0 5.0
CEM 0.1 5.0
GUM 1.6 6.3
METAS 1.0 2.3
NMIJ -0.4 4.9
KRISS -3.2 6.7
LNE 0.3 3.7
NIM 1.9 5.0
CENAM -0.2 3.7
INMETRO 0.6 3.0
UME -1.8 6.3
VNIIM 1.4 6.3
CMS 1.8 3.3
NIMT -2.4 4.9
A*STAR -3.9 8.3
BIM 5.1 9.1
Participating institutes
CCAUV.V-K2
APMP.AUV.V-K2
EURAMET.AUV.V-K2

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       800 Hz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

LABi Di Uneg,i Upos,i
10-4 pC/(m/s2)
10-4 pC/(m/s2)
10-4 pC/(m/s2)
PTB -0.6 1.0
CMI 1.5 3.7
NMISA -0.7 6.3
DPLA -0.8 5.0
CEM 0.5 5.0
GUM 0.7 6.3
METAS 1.2 2.4
NMIJ 0.3 4.9
KRISS -1.7 5.2
LNE -0.2 3.7
NIM 1.5 5.0
CENAM -0.7 3.7
INMETRO 0.5 2.9
UME 0.6 6.3
VNIIM 1.5 6.3
CMS 0.8 4.3
NIMT -2.7 5.2
BIM 1.7 9.0
Participating institutes
CCAUV.V-K2
APMP.AUV.V-K2
EURAMET.AUV.V-K2

CCAUV.V-K2 and APMP.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       2 kHz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

LABi Di Uneg,i Upos,i
10-4 pC/(m/s2)
10-4 pC/(m/s2)
10-4 pC/(m/s2)
PTB -0.2 1.0
CMI 3.4 5.1
NMISA -2.2 10.2
DPLA -0.3 5.1
CEM -2.6 5.1
GUM 0.6 6.4
METAS 0.4 2.8
NMIJ 3.3 5.1
KRISS -1.5 5.2
LNE -0.4 3.8
NIM 1.0 5.1
CENAM 0.5 6.4
INMETRO -0.3 3.0
UME -0.1 12.8
VNIIM 1.2 6.4
CMS -0.8 4.5
NIMT -0.5 9.9
Participating institutes
CCAUV.V-K2
APMP.AUV.V-K2

CCAUV.V-K2 and APMP.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       5 kHz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

LABi Di Uneg,i Upos,i
10-4 pC/(m/s2)
10-4 pC/(m/s2)
10-4 pC/(m/s2)
PTB -0.6 2.5
CMI 6.2 5.7
NMISA -4.9 15.8
DPLA 0.4 8.2
CEM -1.8 5.6
GUM 1.4 8.2
METAS -0.7 3.7
NMIJ 4.0 5.5
LNE -1.6 8.1
NIM 2.0 5.7
CENAM 3.7 13.3
INMETRO 0.2 8.2
UME 1.7 13.3
VNIIM -14.0 13.2
CMS -2.8 12.1
NIMT 0.2 10.4
Participating institutes
CCAUV.V-K2
APMP.AUV.V-K2

CCAUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       10 kHz

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-4 pC/(m/s2).

LABi Di Uneg,i Upos,i
10-4 pC/(m/s2)
10-4 pC/(m/s2)
10-4 pC/(m/s2)
PTB -0.7 11.2
CMI -2.0 11.8
NMISA -12.5 19.9
DPLA -4.1 17.6
CEM -6.0 15.4
GUM 12.6 21.4
METAS -10.9 14.1
NMIJ 11.2 12.4
LNE -3.8 17.6
NIM 4.4 17.7
CENAM 15.2 17.8
INMETRO -4.5 15.4
UME 11.2 26.6
VNIIM -63.2 29.4
Participating institutes
CCAUV.V-K2

DEVICE : Single-Ended Accelerometer
MEASURAND : Phase
FREQUENCIES : 10 Hz, 40 Hz, 80 Hz, 160 Hz, 800 Hz, 2 kHz, 5 kHz and 10 kHz

Degrees of equivalence: Di and Ui expressed in degrees

Results are presented under A4 printable format in Summary Results (PDF file).

Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO 80000-1:2009 Annex B Rule B.

DEVICE : Back-to-Back Accelerometer
MEASURAND : Phase
FREQUENCIES : 10 Hz, 40 Hz, 80 Hz, 160 Hz, 800 Hz, 2 kHz, 5 kHz and 10 kHz

Degrees of equivalence: Di and Ui expressed in degrees

Results are presented under A4 printable format in Summary Results (PDF file).

Unless otherwise stated, in the final numbers presented here, rounding has been applied according to ISO 80000-1:2009 Annex B Rule B.

Comparison
Comparison close
CC comparison
Linked comparison
APMP.AUV.V-K2
Metrology area, Sub-field Acoustics, Ultrasound, Vibration, Vibration
Description Vibration acceleration
Time of measurements 2010
Status Approved for equivalence
Final Reports of the comparisons
References

APMP.AUV.V-K1.1 Technical Protocol

Measurand Amplified charge sensitivity
Parameters Frequency: 40 Hz to 5 kHz
Transfer device Standard accelerometer
Comparison type Key Comparison
Consultative Committee CCAUV (Consultative Committee for Acoustics, Ultrasound and Vibration)
Conducted by APMP (Asia Pacific Metrology Programme)
Comments Results published on 17 July 2017
A limited set of frequencies have been selected to be displayed in the KCDB web. A larger set of frequencies are presented in the Final Report.
This comparison was initially registered as APMP.AUV.V-K1.1.
Pilot institute NMIJ AIST
National Metrology Institute of Japan
Japan
Contact person Takashi USUDA

+81 29 861 4356
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NMIJ AIST

National Metrology Institute of Japan, Japan, APMP

CMS

ITRI Center for Measurement Standards, Chinese Taipei, APMP

NIMT

National Institute of Metrology (Thailand), Thailand, APMP

NMC, A*STAR

National Metrology Centre, Agency for Science, Technology and Research, Singapore, APMP

CCAUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       10 Hz

 

CCAUV.V-K2

The determination of the key comparison reference value, xR, and of its expanded uncertainty (k = 2), UR, is explained in Section 8 of the Final Report.

 

Linking EURAMET.AUV.V-K2 to CCAUV.V-K2

The GUM provides the link between the key comparison EURAMET.AUV.V-K2 and CCAUV.V-K2, having participated in both comparisons. The linking process is described in Section 6 of the EURAMET.AUV.V-K2 Final Report.

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       40 Hz

 

CCAUV.V-K2

The determination of the key comparison reference value, xR, and of its expanded uncertainty (k = 2), UR, is explained in Section 8 of the CCAUV.V-K2 Final Report.

 

Linking APMP.AUV.V-K2 to CCAUV.V-K2

The NMIJ-AIST provides the link between the key comparison APMP.AUV.V-K2 and CCAUV.V-K2, having participated in both comparisons. The linking process is described in Section 8.3 of the APMP.AUV.V-K2 Final Report.

 

Linking EURAMET.AUV.V-K2 to CCAUV.V-K2

The GUM provides the link between the key comparison EURAMET.AUV.V-K2 and CCAUV.V-K2, having participated in both comparisons. The linking process is described in Section 6 of the EURAMET.AUV.V-K2 Final Report.

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       80 Hz

 

CCAUV.V-K2

The determination of the key comparison reference value, xR, and of its expanded uncertainty (k = 2), UR, is explained in Section 8 of the CCAUV.V-K2 Final Report.

 

Linking APMP.AUV.V-K2 to CCAUV.V-K2

The NMIJ-AIST provides the link between the key comparison APMP.AUV.V-K2 and CCAUV.V-K2, having participated in both comparisons. The linking process is described in Section 8.3 of the APMP.AUV.V-K2 Final Report.

 

Linking EURAMET.AUV.V-K2 to CCAUV.V-K2

The GUM provides the link between the key comparison EURAMET.AUV.V-K2 and CCAUV.V-K2, having participated in both comparisons. The linking process is described in Section 6 of the EURAMET.AUV.V-K2 Final Report.

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       160 Hz

 

CCAUV.V-K2

The determination of the key comparison reference value, xR, and of its expanded uncertainty (k = 2), UR, is explained in Section 8 of the CCAUV.V-K2 Final Report.

 

Linking APMP.AUV.V-K2 to CCAUV.V-K2

The NMIJ-AIST provides the link between the key comparison APMP.AUV.V-K2 and CCAUV.V-K2, having participated in both comparisons. The linking process is described in Section 8.3 of the APMP.AUV.V-K2 Final Report.

 

Linking EURAMET.AUV.V-K2 to CCAUV.V-K2

The GUM provides the link between the key comparison EURAMET.AUV.V-K2 and CCAUV.V-K2, having participated in both comparisons. The linking process is described in Section 6 of the EURAMET.AUV.V-K2 Final Report.

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       800 Hz

 

CCAUV.V-K2

The determination of the key comparison reference value, xR, and of its expanded uncertainty (k = 2), UR, is explained in Section 8 of the CCAUV.V-K2 Final Report.

 

Linking APMP.AUV.V-K2 to CCAUV.V-K2

The NMIJ-AIST provides the link between the key comparison APMP.AUV.V-K2 and CCAUV.V-K2, having participated in both comparisons. The linking process is described in Section 8.3 of the APMP.AUV.V-K2 Final Report.

 

Linking EURAMET.AUV.V-K2 to CCAUV.V-K2

The GUM provides the link between the key comparison EURAMET.AUV.V-K2 and CCAUV.V-K2, having participated in both comparisons. The linking process is described in Section 6 of the EURAMET.AUV.V-K2 Final Report.

CCAUV.V-K2 and APMP.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       2 kHz

 

CCAUV.V-K2

The determination of the key comparison reference value, xR, and of its expanded uncertainty (k = 2), UR, is explained in Section 8 of the CCAUV.V-K2 Final Report.

 

Linking APMP.AUV.V-K2 to CCAUV.V-K2

The NMIJ-AIST provides the link between the key comparison APMP.AUV.V-K2 and CCAUV.V-K2, having participated in both comparisons. The linking process is described in Section 8.3 of the APMP.AUV.V-K2 Final Report.

CCAUV.V-K2 and APMP.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       5 kHz

 

CCAUV.V-K2

The determination of the key comparison reference value, xR, and of its expanded uncertainty (k = 2), UR, is explained in Section 8 of the CCAUV.V-K2 Final Report.

 

Linking APMP.AUV.V-K2 to CCAUV.V-K2

The NMIJ-AIST provides the link between the key comparison APMP.AUV.V-K2 and CCAUV.V-K2, having participated in both comparisons. The linking process is described in Section 8.3 of the APMP.AUV.V-K2 Final Report.

CCAUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       10 kHz

CCAUV.V-K2

The determination of the key comparison reference value, xR, and of its expanded uncertainty (k = 2), UR, is explained in Section 8 of the CCAUV.V-K2 Final Report.

 

DEVICE : Single-Ended Accelerometer
MEASURAND : Phase
FREQUENCIES : 10 Hz, 40 Hz, 80 Hz, 160 Hz, 800 Hz, 2 kHz, 5 kHz and 10 kHz

The computation of the key comparison reference value, xR, and of its expanded uncertainty (k = 2), UR, is explained in Section 8 of the Final Report.

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed as explained in Section 8 of the Final Report.
The pair-wise degrees of equivalence are available from the Final report starting on page 58.

DEVICE : Back-to-Back Accelerometer
MEASURAND : Phase
FREQUENCIES : 10 Hz, 40 Hz, 80 Hz, 160 Hz, 800 Hz, 2 kHz, 5 kHz and 10 kHz

The computation of the key comparison reference value, xR, and of its expanded uncertainty (k = 2), UR, is explained in Section 8 of the Final Report.

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed as explained in Section 8 of the Final Report.
The pair-wise degrees of equivalence are available from the Final report starting on page 58.

CCAUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       10 Hz

 

CCAUV.V-K2

The laboratories' individual measurement results are given in Section 7.1.1 of the CCAUV.V-K2 Final Report (starting page 12).

 

EURAMET.AUV.V-K2

The laboratories' individual measurement results are given in Section 5 the EURAMET.AUV.V-K2 Final Report.

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       40 Hz

 

CCAUV.V-K2

The laboratories' individual measurement results are given in Section 7.1.1 of the CCAUV.V-K2 Final Report (starting page 12).

 

APMP.AUV.V-K2

The laboratories' individual measurement results are given in Section 7 the APMP.AUV.V-K2 Final Report.

 

EURAMET.AUV.V-K2

The laboratories' individual measurement results are given in Section 5 the EURAMET.AUV.V-K2 Final Report.

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       80 Hz

 

CCAUV.V-K2

The laboratories' individual measurement results are given in Section 7.1.1 of the CCAUV.V-K2 Final Report (starting page 12).

 

APMP.AUV.V-K2

The laboratories' individual measurement results are given in Section 7 the APMP.AUV.V-K2 Final Report.

 

EURAMET.AUV.V-K2

The laboratories' individual measurement results are given in Section 5 the EURAMET.AUV.V-K2 Final Report.

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       160 Hz

 

CCAUV.V-K2

The laboratories' individual measurement results are given in Section 7.1.1 of the CCAUV.V-K2 Final Report (starting page 12).

 

APMP.AUV.V-K2

The laboratories' individual measurement results are given in Section 7 the APMP.AUV.V-K2 Final Report.

 

EURAMET.AUV.V-K2

The laboratories' individual measurement results are given in Section 5 the EURAMET.AUV.V-K2 Final Report.

CCAUV.V-K2, APMP.AUV.V-K2 and EURAMET.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       800 Hz

 

CCAUV.V-K2

The laboratories' individual measurement results are given in Section 7.1.1 of the CCAUV.V-K2 Final Report (starting page 12).

 

APMP.AUV.V-K2

The laboratories' individual measurement results are given in Section 7 the APMP.AUV.V-K2 Final Report.

 

EURAMET.AUV.V-K2

The laboratories' individual measurement results are given in Section 5 the EURAMET.AUV.V-K2 Final Report.

CCAUV.V-K2 and APMP.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       2 kHz

 

CCAUV.V-K2

The laboratories' individual measurement results are given in Section 7.1.1 of the CCAUV.V-K2 Final Report (starting page 12).

 

APMP.AUV.V-K2

The laboratories' individual measurement results are given in Section 7 the APMP.AUV.V-K2 Final Report.

CCAUV.V-K2 and APMP.AUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       5 kHz

 

CCAUV.V-K2

The laboratories' individual measurement results are given in Section 7.1.1 of the CCAUV.V-K2 Final Report (starting page 12).

 

APMP.AUV.V-K2

The laboratories' individual measurement results are given in Section 7 the APMP.AUV.V-K2 Final Report.

CCAUV.V-K2

DEVICE                 Single-Ended Accelerometer

MEASURAND      Magnitude

FREQUENCY       10 kHz

 

CCAUV.V-K2

The laboratories' individual measurement results are given in Section 7.1.1 of the CCAUV.V-K2 Final Report (starting page 12).

DEVICE : Single-Ended Accelerometer
MEASURAND : Phase
FREQUENCIES : 10 Hz, 40 Hz, 80 Hz, 160 Hz, 800 Hz, 2 kHz, 5 kHz and 10 kHz

The laboratories' individual measurement may be found in Section 7.2.1 of the Final Report (starting page 18).

DEVICE : Back-to-Back Accelerometer
MEASURAND : Phase
FREQUENCIES : 10 Hz, 40 Hz, 80 Hz, 160 Hz, 800 Hz, 2 kHz, 5 kHz and 10 kHz

The laboratories' individual measurement may be found in Section 7.2.2 of the Final Report (starting page 21).