Comparison
Comparison close

CCPR-K3.a, CCPR-K3.a.1, EURAMET.PR-K3.a, EURAMET.PR-K3.a.1 and COOMET.PR-K3.a

MEASURAND       Luminous intensity

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-2.

Luminous intensity
Di
Participating institutes
Comparison
Comparison close

CCPR-K3.a, CCPR-K3.a.1, EURAMET.PR-K3.a, EURAMET.PR-K3.a.1 and COOMET.PR-K3.a

MEASURAND       Luminous intensity

 

Degrees of equivalence represented by Di = (xI - xR) and its expanded uncertainty Ui at a 95 % level of confidence, both expressed in 10-2.

LABi Di Uneg,i Upos,i
10-2
10-2
10-2
LNE-INM/Cnam 0.89 0.60
NMISA 0.51 0.88
NMIA -0.07 0.60
NMIJ -0.09 0.56
INRIM -0.43 0.90
IO-CSIC -0.48 0.72
NIM -0.16 0.48
NIST 0.12 0.48
NPL 0.04 0.30
NRC 0.19 1.06
METAS -0.50 1.12
MKEH 0.05 0.94
PTB -0.31 0.40
SMU -0.36 1.40
VNIIOFI 0.33 0.92
INTI -0.25 0.74
BIPM 0.30 1.00
NMC, A*STAR 0.03 0.33
BIM 0.04 1.50
BelGIM -0.53 0.72
SP -0.61 0.93
UME 0.43 1.50
BEV -0.26 1.10
VSL -0.54 0.64
CMI -0.10 0.80
IPQ 0.57 1.60
DMDM 0.22 1.20
GUM 0.33 1.10
INM (RO) -0.21 1.40
SMU 0.42 0.72
MIKES -0.32 0.44
INM (RO) -0.72 0.90
BelGIM -0.57 0.94
NSC IM -0.66 1.28
KazInMetr -0.22 0.94
Participating institutes
CCPR-K3.a
CCPR-K3.a.1
EURAMET-K3.a
EURAMET-K3.a.1
COOMET-K3.a
COOMET.PR-K3.a
Metrology area, Sub-field Photometry and Radiometry, Photometry
Description Luminous intensity
Time of measurements 2011 - 2013
Status Approved for equivalence
Final Reports of the comparisons
References

COOMET.PR-K3.a Technical Protocol

Measurand Luminous intensity, 100 cd
100 cd
Transfer device Lamps for luminous intensity of the SIS 40-100 type
Comparison type Key Comparison
Consultative Committee CCPR (Consultative Committee for Photometry and Radiometry)
Conducted by COOMET (Euro-Asian Cooperation of National Metrological Institutions)
RMO Internal Identifier COOMET 368/BY/06
Comments Results published on 22 June 2020.
Pilot institute BelGIM
Belarussian State Institute for Metrology
Belarus
Contact person Olga Tarasova

+375 017 365 50 61
Additional
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
BelGIM

Belarussian State Institute for Metrology, Belarus, COOMET

KazStandard

Kazakhstan Institute of Standardization and Metrology, Kazakhstan, COOMET

NSC IM

National Scientific Centre "Institute of Metrology", Ukraine, EURAMET

SMU

Slovensky Metrologicky Ustav, Slovakia, EURAMET

VNIIOFI

All-Russian Scientific Institute for Optical and Physical Measurements, Rosstandart, Russian Federation, COOMET

Comparison
Comparison close

CCPR-K3.a, CCPR-K3.a.1, EURAMET.PR-K3.a, EURAMET.PR-K3.a.1 and COOMET.PR-K3.a

MEASURAND       Luminous intensity

 

The individual measurements, xi, of the participating laboratories take the form of ratios which depend on the reference used at the pilot laboratory. This is cancelled out by normalization using the key comparison reference value, xR. This procedure leads to normalized individual measurements: xi / xR.

 

CCPR-K3.a

The key comparison reference value, xR, is calculated as the weighted average of the invidual results xi, weighted by the inverse square of the individual standard uncertainties, ui, with the application of a minimum cutoff of 0.25%. The INTI and the BIPM are excluded from the calculation of xR. The standard uncertainty of xR is uR = 0.09 %. It is negligible compared to the ui values.

The degree of equivalence of each laboratory with respect to the reference value is given by a pair of terms: Di = (xi - xR) / xR and Ui, its expanded uncertainty
(k = 2), both expressed in relative units. Ui = 2ui.

 

Linking CCPR-K3.a.1 to CCPR-K3.a

The link to the key comparison CCPR-K3.a is made though CSIRO-NML (now NMIA) that participated in both comparisons, see Section 9 of the CCPR-K3.a.1 Final Report.

 

Linking EURAMET.PR-K3.a to CCPR-K3.a

The link to the key comparison CCPR-K3.a is realized through the common participation of LNE-INM/Cnam, INRiM (named "BNM-INM" and "IEN" at the time of the CCPR key comparison) and PTB that participated in both comparisons, see EURAMET.PR-K3.a Final Report.

 

Linking EURAMET.PR-K3.a.1 to CCPR-K3.a

The link to the key comparison CCPR-K3.a is realized through the common participation of LNE see EURAMET.PR-K3.a.1 Final Report.

 

Linking COOMET.PR-K3.a to CCPR-K3.a

The link to the key comparison CCPR-K3.a is realized through the common participation of SMU and VNIIOFI, see COOMET.PR-K3.a Final Report.

Comparison
Comparison close

CCPR-K3.a, CCPR-K3.a.1, EURAMET.PR-K3.a, EURAMET.PR-K3.a.1 and COOMET.PR-K3.a

MEASURAND       Luminous intensity

 

For individual measurements see Final Reports.