EURAMET.EM-S43
Metrology area, Sub-field Electricity and Magnetism, High voltage and current
Description Bilateral comparison of high voltage transformer measuring systems
Time of measurements 2016 - 2020
Status Approved
Final Reports of the comparisons
References

EURAMET.EM-S43 Technical Protocol

Measurand Ratio error and phase displacement of instrument voltage transformers
Nominal primary voltages: 3 kV, 6 kV, 10 kV, 20 kV, 30 kV
Nominal secondary voltage(s): 100 V
Burden: 3 VA
Frequency: 50 Hz
Parameters Ratio error in % or ppm
Phase displacement in rad or min
Transfer device Cast-resin inductive multi-ratio voltage transformer
Comparison type Supplementary Comparison
Consultative Committee CCEM (Consultative Committee for Electricity and Magnetism)
Conducted by EURAMET (European Association of National Metrology Institutes)
RMO Internal Identifier EURAMET Project No 1492
Comments TUBITAK UME and PTB performed a bilateral comparison with their high voltage transformer measuring systems. Travelling standards were a cast-resin inductive multi-ratio voltage transformer. Measurements were held at 50 Hz and at the nominal primary voltages of 3 kV, 6 kV, 10 kV, 20 kV and 30 kV. The nominal secondary voltage was 100 V and the burden was 3 VA.
Pilot institute UME
TÜBITAK Ulusal Metroloji Enstitüsü
Türkiye
Contact person Hüseyin Çaycı

+905327384106
Additional
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
UME

TÜBITAK Ulusal Metroloji Enstitüsü, Türkiye, EURAMET

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

Comparison
Comparison close
Results
Comparison
Comparison close
Results