CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 50 MPa

Graph of equivalence

Red diamonds: participants in CCM.P-K13
Blue triangles: participants in APMP.M.P-K13

Green squares: participants in EURAMET.M.P-K13
Orange circles: participants in EURAMET.M.P-K7

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 100 MPa

Graph of equivalence

Red diamonds: participants in CCM.P-K13
Blue triangles: participants in APMP.M.P-K13

Green squares: participants in EURAMET.M.P-K13
Orange circles: participants in EURAMET.M.P-K7

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 150 MPa

Graph of equivalence

Red diamonds: participants in CCM.P-K13
Blue triangles: participants in APMP.M.P-K13

Green squares: participants in EURAMET.M.P-K13
Orange circles: participants in EURAMET.M.P-K7

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 200 MPa

Graph of equivalence

Red diamonds: participants in CCM.P-K13
Blue triangles: participants in APMP.M.P-K13

Green squares: participants in EURAMET.M.P-K13
Orange circles: participants in EURAMET.M.P-K7

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 250 MPa

Graph of equivalence

Red diamonds: participants in CCM.P-K13
Blue triangles: participants in APMP.M.P-K13

Green squares: participants in EURAMET.M.P-K13
Orange circles: participants in EURAMET.M.P-K7

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 300 MPa

Graph of equivalence

Red diamonds: participants in CCM.P-K13
Blue triangles: participants in APMP.M.P-K13

Green squares: participants in EURAMET.M.P-K13
Orange circles: participants in EURAMET.M.P-K7

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 350 MPa

Graph of equivalence

Red diamonds: participants in CCM.P-K13
Blue triangles: participants in APMP.M.P-K13

Green squares: participants in EURAMET.M.P-K13
Orange circles: participants in EURAMET.M.P-K7

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 400 MPa

Graph of equivalence

Red diamonds: participants in CCM.P-K13
Blue triangles: participants in APMP.M.P-K13

Green squares: participants in EURAMET.M.P-K13
Orange circles: participants in EURAMET.M.P-K7

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 450 MPa

Graph of equivalence

Red diamonds: participants in CCM.P-K13
Blue triangles: participants in APMP.M.P-K13

Green squares: participants in EURAMET.M.P-K13
Orange circles: participants in EURAMET.M.P-K7

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 500 MPa

Graph of equivalence

Red diamonds: participants in CCM.P-K13
Blue triangles: participants in APMP.M.P-K13

Green squares: participants in EURAMET.M.P-K13
Orange circles: participants in EURAMET.M.P-K7

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 50 MPa

Degrees of equivalence relative to the CCM.P-K3 key comparison reference value: Di and Ui (k = 2), both expressed in 10-6

Red: participants in CCM.P-K13
Blue: participants in APMP.M.P-K13

Green: participants in EURAMET.M.P-K13
Orange: participants in EURAMET.M.P-K7

 

Results are presented under A4 printable format in Summary Results (PDF file).

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 100 MPa

Degrees of equivalence relative to the CCM.P-K3 key comparison reference value: Di and Ui (k = 2), both expressed in 10-6

Red: participants in CCM.P-K13
Blue: participants in APMP.M.P-K13

Green: participants in EURAMET.M.P-K13
Orange: participants in EURAMET.M.P-K7

 

Results are presented under A4 printable format in Summary Results (PDF file).

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 150 MPa

Degrees of equivalence relative to the CCM.P-K3 key comparison reference value: Di and Ui (k = 2), both expressed in 10-6

Red: participants in CCM.P-K13
Blue: participants in APMP.M.P-K13

Green: participants in EURAMET.M.P-K13
Orange: participants in EURAMET.M.P-K7

 

Results are presented under A4 printable format in Summary Results (PDF file).

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 200 MPa

Degrees of equivalence relative to the CCM.P-K3 key comparison reference value: Di and Ui (k = 2), both expressed in 10-6

Red: participants in CCM.P-K13
Blue: participants in APMP.M.P-K13

Green: participants in EURAMET.M.P-K13
Orange: participants in EURAMET.M.P-K7

 

Results are presented under A4 printable format in Summary Results (PDF file).

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 250 MPa

Degrees of equivalence relative to the CCM.P-K3 key comparison reference value: Di and Ui (k = 2), both expressed in 10-6

Red: participants in CCM.P-K13
Blue: participants in APMP.M.P-K13

Green: participants in EURAMET.M.P-K13
Orange: participants in EURAMET.M.P-K7

 

Results are presented under A4 printable format in Summary Results (PDF file).

CCM.P-K13, APMP.M.P-K13 and EURAMET.M.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 300 MPa

Degrees of equivalence relative to the CCM.P-K3 key comparison reference value: Di and Ui (k = 2), both expressed in 10-6

Red: participants in CCM.P-K13
Blue: participants in APMP.M.P-K13

Green: participants in EURAMET.M.P-K13
Orange: participants in EURAMET.M.P-K7

 

Results are presented under A4 printable format in Summary Results (PDF file).

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 350 MPa

Degrees of equivalence relative to the CCM.P-K3 key comparison reference value: Di and Ui (k = 2), both expressed in 10-6

Red: participants in CCM.P-K13
Blue: participants in APMP.M.P-K13

Green: participants in EURAMET.M.P-K13
Orange: participants in EURAMET.M.P-K7

 

Results are presented under A4 printable format in Summary Results (PDF file).

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 400 MPa

Degrees of equivalence relative to the CCM.P-K3 key comparison reference value: Di and Ui (k = 2), both expressed in 10-6

Red: participants in CCM.P-K13
Blue: participants in APMP.M.P-K13

Green: participants in EURAMET.M.P-K13
Orange: participants in EURAMET.M.P-K7

 

Results are presented under A4 printable format in Summary Results (PDF file).

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 450 MPa

Degrees of equivalence relative to the CCM.P-K3 key comparison reference value: Di and Ui (k = 2), both expressed in 10-6

Red: participants in CCM.P-K13
Blue: participants in APMP.M.P-K13

Green: participants in EURAMET.M.P-K13
Orange: participants in EURAMET.M.P-K7

 

Results are presented under A4 printable format in Summary Results (PDF file).

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 500 MPa

Degrees of equivalence relative to the CCM.P-K3 key comparison reference value: Di and Ui (k = 2), both expressed in 10-6

Red: participants in CCM.P-K13
Blue: participants in APMP.M.P-K13

Green: participants in EURAMET.M.P-K13
Orange: participants in EURAMET.M.P-K7

 

Results are presented under A4 printable format in Summary Results (PDF file).

Comparison
Comparison close
EURAMET.M.P-K13
Metrology area, Sub-field Mass and related quantities, Pressure
Description Hydraulic gauge pressure
Time of measurements 2009 - 2011
Status Approved for equivalence
Final Reports of the comparisons
Measurand Hydraulic gauge pressure: 50 MPa to 500 MPa
Transfer device Pressure balance
Comparison type Key Comparison
Consultative Committee CCM (Consultative Committee for Mass and Related Quantities)
Conducted by EURAMET (European Association of National Metrology Institutes)
RMO Internal Identifier EURAMET Project No 1091
Comments

Results published on 23 March 2015

APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7 results are linked to those of CCM.P-K13

Hydraulic gauge pressure

Pilot institute UME
TÜBITAK Ulusal Metroloji Enstitüsü
Türkiye
Contact person I. Kocas

+90 262 679 5000 ext 5200
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
UME

TÜBITAK Ulusal Metroloji Enstitüsü, Türkiye, EURAMET

INRIM

Istituto Nazionale di Ricerca Metrologica, Italy, EURAMET

MIRS/IMT/LMT

MIRS/Institute of Metals and Technology/Laboratory of Pressure Metrology, Slovenia, EURAMET

NIS

National Institute for Standards, Egypt, AFRIMETS

NMISA

National Metrology Institute of South Africa, South Africa, AFRIMETS

NPL

National Physical Laboratory, United Kingdom, EURAMET

PTB

Physikalisch-Technische Bundesanstalt, Germany, EURAMET

SMU

Slovensky Metrologicky Ustav, Slovakia, EURAMET

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 50 MPa

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MED |xi - xR| / (n - 1)1/2, where MED means median and n is the number of participants.

xR = 1.961152 mm2
uR / xR = 13.5 10-6

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) / xR, the relative deviation from the reference value, and
Ui = 2(ui2 + uDC2 + uR2)1/2 / xR, its expanded uncertainty (k = 2).

When required, the degree of equivalence between two laboratories i and j is given by a pair of terms:
Dij = Di - Dj = (xi - xj) / xR, the relative difference between their results, and
Uij = 2(ui2 + uj2 + uDC2)1/2 / xR, its expanded uncertainty (k = 2).

No pair-wise degrees of equivalence are displayed here, but they can be found in the relevant Final Report.

  • Key comparison APMP.M.P-K13

The results of the participants in APMP.M.P-K13 are linked to those of CCM.P-K13 through the results of the two linking laboratories, NMIJ and NPLI, as explained in Section 7 of the APMP.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K13

The results of the participants in EURAMET.M.P-K13 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 6 of the EURAMET.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K7

The results of the participants in EURAMET.M.P-K7 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 9 of the EURAMET.M.P-K7 Final Report.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 100 MPa

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MED |xi - xR| / (n - 1)1/2, where MED means median and n is the number of participants.

xR = 1.961261 mm2
uR / xR = 4.6 10-6

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) / xR, the relative deviation from the reference value, and
Ui = 2(ui2 + uDC2 + uR2)1/2 / xR, its expanded uncertainty (k = 2).

When required, the degree of equivalence between two laboratories i and j is given by a pair of terms:
Dij = Di - Dj = (xi - xj) / xR, the relative difference between their results, and
Uij = 2(ui2 + uj2 + uDC2)1/2 / xR, its expanded uncertainty (k = 2).

No pair-wise degrees of equivalence are displayed here, but they can be found in the relevant Final Report.

  • Key comparison APMP.M.P-K13

The results of the participants in APMP.M.P-K13 are linked to those of CCM.P-K13 through the results of the two linking laboratories, NMIJ and NPLI, as explained in Section 7 of the APMP.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K13

The results of the participants in EURAMET.M.P-K13 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 6 of the EURAMET.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K7

The results of the participants in EURAMET.M.P-K7 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 9 of the EURAMET.M.P-K7 Final Report.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 150 MPa

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MED |xi - xR| / (n - 1)1/2, where MED means median and n is the number of participants.

xR = 1.961376 mm2
uR / xR = 3.4 10-6

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) / xR, the relative deviation from the reference value, and
Ui = 2(ui2 + uDC2 + uR2)1/2 / xR, its expanded uncertainty (k = 2).

When required, the degree of equivalence between two laboratories i and j is given by a pair of terms:
Dij = Di - Dj = (xi - xj) / xR, the relative difference between their results, and
Uij = 2(ui2 + uj2 + uDC2)1/2 / xR, its expanded uncertainty (k = 2).

No pair-wise degrees of equivalence are displayed here, but they can be found in the relevant Final Report.

  • Key comparison APMP.M.P-K13

The results of the participants in APMP.M.P-K13 are linked to those of CCM.P-K13 through the results of the two linking laboratories, NMIJ and NPLI, as explained in Section 7 of the APMP.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K13

The results of the participants in EURAMET.M.P-K13 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 6 of the EURAMET.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K7

The results of the participants in EURAMET.M.P-K7 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 9 of the EURAMET.M.P-K7 Final Report.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 200 MPa

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MED |xi - xR| / (n - 1)1/2, where MED means median and n is the number of participants.

xR = 1.961486 mm2
uR / xR = 3.4 10-6

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) / xR, the relative deviation from the reference value, and
Ui = 2(ui2 + uDC2 + uR2)1/2 / xR, its expanded uncertainty (k = 2).

When required, the degree of equivalence between two laboratories i and j is given by a pair of terms:
Dij = Di - Dj = (xi - xj) / xR, the relative difference between their results, and
Uij = 2(ui2 + uj2 + uDC2)1/2 / xR, its expanded uncertainty (k = 2).

No pair-wise degrees of equivalence are displayed here, but they can be found in the relevant Final Report.

  • Key comparison APMP.M.P-K13

The results of the participants in APMP.M.P-K13 are linked to those of CCM.P-K13 through the results of the two linking laboratories, NMIJ and NPLI, as explained in Section 7 of the APMP.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K13

The results of the participants in EURAMET.M.P-K13 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 6 of the EURAMET.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K7

The results of the participants in EURAMET.M.P-K7 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 9 of the EURAMET.M.P-K7 Final Report.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 250 MPa

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MED |xi - xR| / (n - 1)1/2, where MED means median and n is the number of participants.

xR = 1.961589 mm2
uR / xR = 5.4 10-6

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) / xR, the relative deviation from the reference value, and
Ui = 2(ui2 + uDC2 + uR2)1/2 / xR, its expanded uncertainty (k = 2).

When required, the degree of equivalence between two laboratories i and j is given by a pair of terms:
Dij = Di - Dj = (xi - xj) / xR, the relative difference between their results, and
Uij = 2(ui2 + uj2 + uDC2)1/2 / xR, its expanded uncertainty (k = 2).

No pair-wise degrees of equivalence are displayed here, but they can be found in the relevant Final Report.

  • Key comparison APMP.M.P-K13

The results of the participants in APMP.M.P-K13 are linked to those of CCM.P-K13 through the results of the two linking laboratories, NMIJ and NPLI, as explained in Section 7 of the APMP.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K13

The results of the participants in EURAMET.M.P-K13 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 6 of the EURAMET.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K7

The results of the participants in EURAMET.M.P-K7 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 9 of the EURAMET.M.P-K7 Final Report.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 300 MPa

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MED |xi - xR| / (n - 1)1/2, where MED means median and n is the number of participants.

xR = 1.961688 mm2
uR / xR = 5.9 10-6

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) / xR, the relative deviation from the reference value, and
Ui = 2(ui2 + uDC2 + uR2)1/2 / xR, its expanded uncertainty (k = 2).

When required, the degree of equivalence between two laboratories i and j is given by a pair of terms:
Dij = Di - Dj = (xi - xj) / xR, the relative difference between their results, and
Uij = 2(ui2 + uj2 + uDC2)1/2 / xR, its expanded uncertainty (k = 2).

No pair-wise degrees of equivalence are displayed here, but they can be found in the relevant Final Report.

  • Key comparison APMP.M.P-K13

The results of the participants in APMP.M.P-K13 are linked to those of CCM.P-K13 through the results of the two linking laboratories, NMIJ and NPLI, as explained in Section 7 of the APMP.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K13

The results of the participants in EURAMET.M.P-K13 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 6 of the EURAMET.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K7

The results of the participants in EURAMET.M.P-K7 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 9 of the EURAMET.M.P-K7 Final Report.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 350 MPa

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MED |xi - xR| / (n - 1)1/2, where MED means median and n is the number of participants.

xR = 1.961780 mm2
uR / xR = 7.7 10-6

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) / xR, the relative deviation from the reference value, and
Ui = 2(ui2 + uDC2 + uR2)1/2 / xR, its expanded uncertainty (k = 2).

When required, the degree of equivalence between two laboratories i and j is given by a pair of terms:
Dij = Di - Dj = (xi - xj) / xR, the relative difference between their results, and
Uij = 2(ui2 + uj2 + uDC2)1/2 / xR, its expanded uncertainty (k = 2).

No pair-wise degrees of equivalence are displayed here, but they can be found in the relevant Final Report.

  • Key comparison APMP.M.P-K13

The results of the participants in APMP.M.P-K13 are linked to those of CCM.P-K13 through the results of the two linking laboratories, NMIJ and NPLI, as explained in Section 7 of the APMP.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K13

The results of the participants in EURAMET.M.P-K13 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 6 of the EURAMET.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K7

The results of the participants in EURAMET.M.P-K7 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 9 of the EURAMET.M.P-K7 Final Report.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 400 MPa

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MED |xi - xR| / (n - 1)1/2, where MED means median and n is the number of participants.

xR = 1.961868 mm2
uR / xR = 6.7 10-6

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) / xR, the relative deviation from the reference value, and
Ui = 2(ui2 + uDC2 + uR2)1/2 / xR, its expanded uncertainty (k = 2).

When required, the degree of equivalence between two laboratories i and j is given by a pair of terms:
Dij = Di - Dj = (xi - xj) / xR, the relative difference between their results, and
Uij = 2(ui2 + uj2 + uDC2)1/2 / xR, its expanded uncertainty (k = 2).

No pair-wise degrees of equivalence are displayed here, but they can be found in the relevant Final Report.

  • Key comparison APMP.M.P-K13

The results of the participants in APMP.M.P-K13 are linked to those of CCM.P-K13 through the results of the two linking laboratories, NMIJ and NPLI, as explained in Section 7 of the APMP.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K13

The results of the participants in EURAMET.M.P-K13 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 6 of the EURAMET.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K7

The results of the participants in EURAMET.M.P-K7 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 9 of the EURAMET.M.P-K7 Final Report.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 450 MPa

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MED |xi - xR| / (n - 1)1/2, where MED means median and n is the number of participants.

xR = 1.961955 mm2
uR / xR = 8.8 10-6

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) / xR, the relative deviation from the reference value, and
Ui = 2(ui2 + uDC2 + uR2)1/2 / xR, its expanded uncertainty (k = 2).

When required, the degree of equivalence between two laboratories i and j is given by a pair of terms:
Dij = Di - Dj = (xi - xj) / xR, the relative difference between their results, and
Uij = 2(ui2 + uj2 + uDC2)1/2 / xR, its expanded uncertainty (k = 2).

No pair-wise degrees of equivalence are displayed here, but they can be found in the relevant Final Report.

  • Key comparison APMP.M.P-K13

The results of the participants in APMP.M.P-K13 are linked to those of CCM.P-K13 through the results of the two linking laboratories, NMIJ and NPLI, as explained in Section 7 of the APMP.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K13

The results of the participants in EURAMET.M.P-K13 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 6 of the EURAMET.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K7

The results of the participants in EURAMET.M.P-K7 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 9 of the EURAMET.M.P-K7 Final Report.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 500 MPa

The key comparison reference value, xR, is calculated as the median of the participants' values. The standard uncertainty of the key comparison reference value, uR, is calculated by the formula uR = 1.858 MED |xi - xR| / (n - 1)1/2, where MED means median and n is the number of participants.

xR = 1.962041 mm2
uR / xR = 11.1 10-6

The degree of equivalence of laboratory i with respect to the key comparison reference value is given by a pair of terms:
Di = (xi - xR) / xR, the relative deviation from the reference value, and
Ui = 2(ui2 + uDC2 + uR2)1/2 / xR, its expanded uncertainty (k = 2).

When required, the degree of equivalence between two laboratories i and j is given by a pair of terms:
Dij = Di - Dj = (xi - xj) / xR, the relative difference between their results, and
Uij = 2(ui2 + uj2 + uDC2)1/2 / xR, its expanded uncertainty (k = 2).

No pair-wise degrees of equivalence are displayed here, but they can be found in the relevant Final Report.

  • Key comparison APMP.M.P-K13

The results of the participants in APMP.M.P-K13 are linked to those of CCM.P-K13 through the results of the two linking laboratories, NMIJ and NPLI, as explained in Section 7 of the APMP.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K13

The results of the participants in EURAMET.M.P-K13 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 6 of the EURAMET.M.P-K13 Final Report.

  • Key comparison EURAMET.M.P-K7

The results of the participants in EURAMET.M.P-K7 are linked to those of CCM.P-K13 through the results of the linking laboratory, PTB, as explained in Section 9 of the EURAMET.M.P-K7 Final Report.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 50 MPa

xi : result of measurement of A' (p', 20 °C) taken as the arithmetic average of 10 values obtained by laboratory i and corrected for drift
ui : combined standard uncertainty of xi as reported by laboratory i (without including uncertainty of the drift correction)
uDC : standard uncertainty of the drift correction of xi
uDC / xi = 0.5 x 10-6

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NIST

1.961187

17

Jan 2009

 CENAM 1.961171 85 Mar 2009
 NPLI 1.961104 44 May 2009
 NIM 1.961098 17 Jun 2009
 NMIJ 1.961152 17 Oct 2009
 LNE 1.961191 11 Feb 2010
 PTB 1.961148 15 Mar 2010

  • Key comparison APMP.M.P-K13

The results of the participating laboratories in key comparison APMP.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2010 and 2013.

  • Key comparison EURAMET.M.P-K13

The results of the participating laboratories in key comparison EURAMET.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2009 and 2011.

  • Key comparison EURAMET.M.P-K7

The results of the participating laboratories in key comparison EURAMET.M.P-K7 are given in Section 7 of the Final Report.
Measurements were carried out between 2005 and 2007.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 100 MPa

xi : result of measurement of A' (p', 20 °C) taken as the arithmetic average of 10 values obtained by laboratory i and corrected for drift
ui : combined standard uncertainty of xi as reported by laboratory i (without including uncertainty of the drift correction)
uDC : standard uncertainty of the drift correction of xi
uDC / xi = 0.5 x 10-6

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NIST

1.961285

17

Jan 2009

 CENAM 1.961258 85 Mar 2009
 NPLI 1.961244 44 May 2009
 NIM 1.961175 17 Jun 2009
 NMIJ 1.961269 17 Oct 2009
 LNE 1.961273 13 Feb 2010
 PTB 1.961261 15 Mar 2010

  • Key comparison APMP.M.P-K13

The results of the participating laboratories in key comparison APMP.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2010 and 2013.

  • Key comparison EURAMET.M.P-K13

The results of the participating laboratories in key comparison EURAMET.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2009 and 2011.

  • Key comparison EURAMET.M.P-K7

The results of the participating laboratories in key comparison EURAMET.M.P-K7 are given in Section 7 of the Final Report.
Measurements were carried out between 2005 and 2007.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 150 MPa

xi : result of measurement of A' (p', 20 °C) taken as the arithmetic average of 10 values obtained by laboratory i and corrected for drift
ui : combined standard uncertainty of xi as reported by laboratory i (without including uncertainty of the drift correction)
uDC : standard uncertainty of the drift correction of xi
uDC / xi = 0.6 x 10-6

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NIST

1.961409

17

Jan 2009

 CENAM 1.961367 85 Mar 2009
 NPLI 1.961376 44 May 2009
 NIM 1.961280 17 Jun 2009
 NMIJ 1.961386 19 Oct 2009
 LNE 1.961373 16 Feb 2010
 PTB 1.961378 16 Mar 2010

  • Key comparison APMP.M.P-K13

The results of the participating laboratories in key comparison APMP.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2010 and 2013.

  • Key comparison EURAMET.M.P-K13

The results of the participating laboratories in key comparison EURAMET.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2009 and 2011.

  • Key comparison EURAMET.M.P-K7

The results of the participating laboratories in key comparison EURAMET.M.P-K7 are given in Section 7 of the Final Report.
Measurements were carried out between 2005 and 2007.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 200 MPa

xi : result of measurement of A' (p', 20 °C) taken as the arithmetic average of 10 values obtained by laboratory i and corrected for drift
ui : combined standard uncertainty of xi as reported by laboratory i (without including uncertainty of the drift correction)
uDC : standard uncertainty of the drift correction of xi
uDC / xi = 0.2 x 10-6

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NIST

1.961519

17

Jan 2009

 CENAM 1.961481 85 Mar 2009
 NPLI 1.961493 44 May 2009
 NIM 1.961381 18 Jun 2009
 NMIJ 1.961495 20 Oct 2009
 LNE 1.961474 18 Feb 2010
 PTB 1.961486 17 Mar 2010

  • Key comparison APMP.M.P-K13

The results of the participating laboratories in key comparison APMP.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2010 and 2013.

  • Key comparison EURAMET.M.P-K13

The results of the participating laboratories in key comparison EURAMET.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2009 and 2011.

  • Key comparison EURAMET.M.P-K7

The results of the participating laboratories in key comparison EURAMET.M.P-K7 are given in Section 7 of the Final Report.
Measurements were carried out between 2005 and 2007.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 250 MPa

xi : result of measurement of A' (p', 20 °C) taken as the arithmetic average of 10 values obtained by laboratory i and corrected for drift
ui : combined standard uncertainty of xi as reported by laboratory i (without including uncertainty of the drift correction)
uDC : standard uncertainty of the drift correction of xi
uDC / xi = 0.4 x 10-6

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NIST

1.961607

17

Jan 2009

 CENAM 1.961589 85 Mar 2009
 NPLI 1.961603 44 May 2009
 NIM 1.961477 19 Jun 2009
 NMIJ 1.961597 22 Oct 2009
 LNE 1.961574 20 Feb 2010
 PTB 1.961589 18 Mar 2010

  • Key comparison APMP.M.P-K13

The results of the participating laboratories in key comparison APMP.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2010 and 2013.

  • Key comparison EURAMET.M.P-K13

The results of the participating laboratories in key comparison EURAMET.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2009 and 2011.

  • Key comparison EURAMET.M.P-K7

The results of the participating laboratories in key comparison EURAMET.M.P-K7 are given in Section 7 of the Final Report.
Measurements were carried out between 2005 and 2007.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 300 MPa

xi : result of measurement of A' (p', 20 °C) taken as the arithmetic average of 10 values obtained by laboratory i and corrected for drift
ui : combined standard uncertainty of xi as reported by laboratory i (without including uncertainty of the drift correction)
uDC : standard uncertainty of the drift correction of xi
uDC / xi = 1.2 x 10-6

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NIST

-

-

-

 CENAM 1.961699 85 Mar 2009
 NPLI 1.961705 44 May 2009
 NIM 1.961564 20 Jun 2009
 NMIJ 1.961692 24 Oct 2009
 LNE 1.961671 23 Feb 2010
 PTB 1.961685 20 Mar 2010

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

The results of the participating laboratories in key comparison APMP.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2010 and 2013.

  • Key comparison EURAMET.M.P-K13

The results of the participating laboratories in key comparison EURAMET.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2009 and 2011.

  • Key comparison EURAMET.M.P-K7

The results of the participating laboratories in key comparison EURAMET.M.P-K7 are given in Section 7 of the Final Report.
Measurements were carried out between 2005 and 2007.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 350 MPa

xi : result of measurement of A' (p', 20 °C) taken as the arithmetic average of 10 values obtained by laboratory i and corrected for drift
ui : combined standard uncertainty of xi as reported by laboratory i (without including uncertainty of the drift correction)
uDC : standard uncertainty of the drift correction of xi
uDC / xi = 1.4 x 10-6

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NIST

-

-

-

 CENAM 1.961802 85 Mar 2009
 NPLI 1.961802 44 May 2009
 NIM 1.961646 21 Jun 2009
 NMIJ 1.961782 27 Oct 2009
 LNE 1.961766 25 Feb 2010
 PTB 1.961777 22 Mar 2010

  • Key comparison APMP.M.P-K13

The results of the participating laboratories in key comparison APMP.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2010 and 2013.

  • Key comparison EURAMET.M.P-K13

The results of the participating laboratories in key comparison EURAMET.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2009 and 2011.

  • Key comparison EURAMET.M.P-K7

The results of the participating laboratories in key comparison EURAMET.M.P-K7 are given in Section 7 of the Final Report.
Measurements were carried out between 2005 and 2007.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 400 MPa

xi : result of measurement of A' (p', 20 °C) taken as the arithmetic average of 10 values obtained by laboratory i and corrected for drift
ui : combined standard uncertainty of xi as reported by laboratory i (without including uncertainty of the drift correction)
uDC : standard uncertainty of the drift correction of xi
uDC / xi = 1.6 x 10-6

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NIST

-

-

-

 CENAM 1.961905 85 Mar 2009
 NPLI 1.961887 44 May 2009
 NIM 1.961729 22 Jun 2009
 NMIJ 1.961871 30 Oct 2009
 LNE 1.961856 28 Feb 2010
 PTB 1.961866 24 Mar 2010

  • Key comparison APMP.M.P-K13

The results of the participating laboratories in key comparison APMP.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2010 and 2013.

  • Key comparison EURAMET.M.P-K13

The results of the participating laboratories in key comparison EURAMET.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2009 and 2011.

  • Key comparison EURAMET.M.P-K7

The results of the participating laboratories in key comparison EURAMET.M.P-K7 are given in Section 7 of the Final Report.
Measurements were carried out between 2005 and 2007.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 450 MPa

xi : result of measurement of A' (p', 20 °C) taken as the arithmetic average of 10 values obtained by laboratory i and corrected for drift
ui : combined standard uncertainty of xi as reported by laboratory i (without including uncertainty of the drift correction)
uDC : standard uncertainty of the drift correction of xi
uDC / xi = 2.1 x 10-6

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NIST

-

-

-

 CENAM 1.962013 85 Mar 2009
 NPLI 1.961987 44 May 2009
 NIM 1.961774 23 Jun 2009
 NMIJ 1.961956 33 Oct 2009
 LNE 1.961945 30 Feb 2010
 PTB 1.961953 26 Mar 2010

  • Key comparison APMP.M.P-K13

The results of the participating laboratories in key comparison APMP.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2010 and 2013.

  • Key comparison EURAMET.M.P-K13

The results of the participating laboratories in key comparison EURAMET.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2009 and 2011.

  • Key comparison EURAMET.M.P-K7

The results of the participating laboratories in key comparison EURAMET.M.P-K7 are given in Section 7 of the Final Report.
Measurements were carried out between 2005 and 2007.

CCM.P-K13, APMP.M.P-K13, EURAMET.M.P-K13 and EURAMET.M.P-K7

  • Key comparison CCM.P-K13

MEASURAND : Effective area A' (p', 20 °C) of piston-cylinder unit DH-Budenberg 4603 mounted in a DH-Budenberg pressure balance model 5306 determined from comparisons against participants primary pressure standards
NOMINAL VALUE : 2 mm2
NOMINAL PRESSURE : p' = 500 MPa

xi : result of measurement of A' (p', 20 °C) taken as the arithmetic average of 10 values obtained by laboratory i and corrected for drift
ui : combined standard uncertainty of xi as reported by laboratory i (without including uncertainty of the drift correction)
uDC : standard uncertainty of the drift correction of xi
uDC / xi = 2.2 x 10-6

 Lab i xi
/ mm2
ui / xi
/ 10-6
Date of measurement
 NIST

-

-

-

 CENAM 1.962120 85 Mar 2009
 NPLI 1.962093 44 May 2009
 NIM 1.961853 25 Jun 2009
 NMIJ 1.962041 36 Oct 2009
 LNE 1.962040 33 Feb 2010
 PTB 1.962040 28 Mar 2010

  • Key comparison APMP.M.P-K13

The results of the participating laboratories in key comparison APMP.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2010 and 2013.

  • Key comparison EURAMET.M.P-K13

The results of the participating laboratories in key comparison EURAMET.M.P-K13 are given in Section 6 of the Final Report.
Measurements were carried out between 2009 and 2011.

  • Key comparison EURAMET.M.P-K7

The results of the participating laboratories in key comparison EURAMET.M.P-K7 are given in Section 7 of the Final Report.
Measurements were carried out between 2005 and 2007.