Results for: cesmec
Filters
CMC Quick search Filters close;
CMC Area
To filter, either use the tick boxes or click on the underscored item
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Chile, CESMEC (Centro de Estudios, Medición y Certificación de Calidad)
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
10 °C to 90 °C
Platinum Resistance Thermometer
Absolute expanded uncertainty
:
1.0E1 mK
Comparison, stirred water bath
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
CMC ID :
SIM-T-CL-00000BKL-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
250 °C to 450 °C
Digital system thermometer (PRT sensor)
Absolute expanded uncertainty
:
5.0E1 mK
Comparison, vertical block furnace
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
CMC ID :
SIM-T-CL-00000BKS-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
450 °C to 660 °C
Digital system thermometer (PRT sensor)
Absolute expanded uncertainty
:
7.0E1 mK
Comparison, vertical block furnace
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
CMC ID :
SIM-T-CL-00000BKT-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
250 °C to 450 °C
Platinum Resistance Thermometer
Absolute expanded uncertainty
:
5.0E1 mK
Comparison, vertical block furnace
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
CMC ID :
SIM-T-CL-00000BKN-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-40 °C to 10 °C
Digital system thermometer (PRT sensor)
Absolute expanded uncertainty
:
2.5E1 mK
Comparison, stirred alcohol bath
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
CMC ID :
SIM-T-CL-00000BKP-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-40 °C to 10 °C
Platinum Resistance Thermometer
Absolute expanded uncertainty
:
1.5E1 mK
Comparison, stirred alcohol bath
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
CMC ID :
SIM-T-CL-00000BKK-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
90 °C to 250 °C
Digital system thermometer (PRT sensor)
Absolute expanded uncertainty
:
2.5E1 mK
Comparison, stirred silicon oil bath
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
CMC ID :
SIM-T-CL-00000BKR-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
10 °C to 90 °C
Digital system thermometer (PRT sensor)
Absolute expanded uncertainty
:
1.5E1 mK
Comparison, stirred water bath
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
CMC ID :
SIM-T-CL-00000BKQ-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
90 °C to 250 °C
Platinum Resistance Thermometer
Absolute expanded uncertainty
:
1.5E1 mK
Comparison, stirred silicon oil bath
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
CMC ID :
SIM-T-CL-00000BKM-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
450 °C to 660 °C
Platinum Resistance Thermometer
Absolute expanded uncertainty
:
7.0E1 mK
Comparison, vertical block furnace
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
CMC ID :
SIM-T-CL-00000BKO-1
Chile, INN (Instituto Nacional de Normalización)
Mass ,
Conventional mass
:
10 mg
Standard weights
Absolute expanded uncertainty
:
2.0E-3 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055N-1
Mass ,
Conventional mass
:
20 mg
Standard weights
Absolute expanded uncertainty
:
3.0E-3 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055O-1
Mass ,
Conventional mass
:
1 mg
Standard weights
Absolute expanded uncertainty
:
2.0E-3 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055K-1
Mass ,
Conventional mass
:
20 kg
Standard weights
Absolute expanded uncertainty
:
1.0E1 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055I-1
Mass ,
Conventional mass
:
50 kg
Standard weights
Absolute expanded uncertainty
:
7.5E1 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055J-1
Mass ,
Conventional mass
:
2 kg
Standard weights
Absolute expanded uncertainty
:
1 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055F-1
Mass ,
Conventional mass
:
20 g
Standard weights
Absolute expanded uncertainty
:
2.5E-2 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-00000559-1
Mass ,
Conventional mass
:
1 kg
Standard weights
Absolute expanded uncertainty
:
0.5 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055E-1
Mass ,
Conventional mass
:
100 mg
Standard weights
Absolute expanded uncertainty
:
5.0E-3 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055Q-1
Mass ,
Conventional mass
:
50 mg
Standard weights
Absolute expanded uncertainty
:
4.0E-3 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055P-1