Results for: Tanzania
Filters
CMC Quick search Filters close;
CMC Area
To filter, either use the tick boxes or click on the underscored item
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Tanzania, TBS (Tanzania Bureau of Standards)
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-30 °C to 50 °C
Base-metal thermocouples
Absolute expanded uncertainty
:
0.3 °C to 0.5 °C
Comparison at Liqud Bath
Liquid Bath : Ethanol
Approved on 15 May 2025
CMC ID :
AFRIMETS-T-TZ-00000PVC-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-30 °C to 50 °C
Liquid in glass thermometer: scale division 0.1 oC, Total Immersion
Absolute expanded uncertainty
:
5.0E-2 °C to 0.1 °C
Comparison at stirred liquid bath
Liquid Bath : Ethanol
Approved on 15 May 2025
CMC ID :
AFRIMETS-T-TZ-00000PVB-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-30 °C to 50 °C
Industrial platinum resistance thermometers (IPRTs)
Absolute expanded uncertainty
:
5.0E-2 °C to 6.0E-2 °C
Comparison at liquid bath
Comparison at Liquid Bath : Ethanol
Approved on 15 May 2025
CMC ID :
AFRIMETS-T-TZ-00000PZX-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-30 °C to 50 °C
Noble-metal thermocouples
Absolute expanded uncertainty
:
0.3 °C to 0.5 °C
Comparison at a strried Liquid Bath
Liquid Bath : Ethanol
Approved on 15 May 2025
CMC ID :
AFRIMETS-T-TZ-00000Q03-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-30 °C to 50 °C
Temperature sensor with display unit (RTD and Thermistor based)
Absolute expanded uncertainty
:
5.0E-2 °C to 0.1 °C
Comparison at liquid bath
Liquid Bath : Ethanol
Approved on 15 May 2025
CMC ID :
AFRIMETS-T-TZ-00000PP1-1