Results for: Sri Lanka
				Filters
				
					CMC Quick search Filters close;
				 
				
					
					
						
							CMC Area
							
							To filter, either use the tick boxes or click on the underscored item
						 
						
						
						
					 
			 
			
				
				
				
					In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
					Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
				
					Sri Lanka, MUSSD (Department of Measurement Units, Standards and Services)
					
						
    
        
            
                Pressure ,
				 Gauge pressure: gas medium
                :
                500 kPa to 7000 kPa
            
            
                Instruments for pneumatic pressure measurement
            
            
                Absolute expanded uncertainty
                :
                5.6E1 Pa to 3.4E2 Pa
            
            
            
                Q[50 , 0.000048] = [(50 Pa)2 + (0.000048*P )2 ]1/2     P = Gauge Pressure in Pa
The Uncertainty is expressed in Pa
            
            
                Pressure balance
            
            
                DKD-R 6-1 “Calibration of pressure gauges”, Edition 03/2014
            
            
            
                MUSSD, National Metrology Institute of Sri Lanka
            
            
                Approved on 01 July 2022
            
            
            
                CMC ID :
                APMP-M-LK-00000NDI-1 
             
         
        
        
     
 
					 
					
						
    
        
            
                Pressure ,
				 Gauge pressure: liquid medium
                :
                0.5 MPa to 200 MPa
            
            
                Instruments for hydraulic pressure measurement
            
            
                Absolute expanded uncertainty
                :
                0.12 kPa to 1.2E1 kPa
            
            
            
                Q [0.12 , 0.000062] = [ (0.12 kPa)2 + (0.000062*P)2] 1/2     P= Hydraulic gauge Pressure in kPa
The uncertainty is expressed in kPa
            
            
                Pressure balance
            
            
                DKD-R 6-1 “Calibration of pressure gauges”, Edition 03/2014
            
            
            
                MUSSD, National Metrology Institute of Sri Lanka
            
            
                Approved on 01 July 2022
            
            
            
                CMC ID :
                APMP-M-LK-00000NDJ-1