Results for: Norway "JV"
Filters
CMC Quick search Filters close;
?
Added keywords act as an additional filter. Keywords separated by a space will be treated by the logical function AND automatically. You may also use OR. Keyword searches also find close matches. Placing quotation marks (") before and after the search word, e.g. "word", will allow for exact matches.

In case of multiple search, please use ADVANCED SEARCH.
105 results
Reset all
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Norway, JV (Justervesenet)
Mass , Mass : 50 kg to 100 kg
Mass standard
Absolute expanded uncertainty : 2.2E1 mg to 1.7E2 mg
Comparison in air
Uncertainty scales with measurand level The volume of the mass standards is known.
Approved on 02 December 2003
CMC ID : EURAMET-M-NO-00000503-1
 
AC voltage (up to the MHz range) , AC voltage: AC/DC transfer difference at medium voltage : 0.5 V to 5 V
Thermal converter, AC-DC transfer standard
Relative expanded uncertainty : 5 µV/V to 4.0E1 µV/V
{"row_1":{"col_1":"","col_2":"10 Hz to 30 Hz","col_3":"31 Hz to 300 Hz","col_4":"301 Hz to 10 kHz","col_5":"11 kHz to 50 kHz","col_6":"51 kHz to 100 kHz","col_7":"101 kHz to 200 kHz","col_8":"201 kHz to 500 kHz","col_9":"501 kHz to 700 kHz","col_10":"701 kHz to 1 MHz"},"row_2":{"col_1":"1 mV to 3 mV","col_2":"160","col_3":"150","col_4":"150","col_5":"150","col_6":"150","col_7":"150","col_8":"200","col_9":"250","col_10":"350"},"row_3":{"col_1":"3 mV to 5 mV","col_2":"120","col_3":"100","col_4":"100","col_5":"80","col_6":"80","col_7":"90","col_8":"150","col_9":"220","col_10":"300"},"row_4":{"col_1":"5 mV to 10 mV","col_2":"100","col_3":"90","col_4":"70","col_5":"40","col_6":"50","col_7":"70","col_8":"130","col_9":"190","col_10":"250"},"row_5":{"col_1":"10 mV to 25 mV","col_2":"50","col_3":"40","col_4":"30","col_5":"35","col_6":"40","col_7":"50","col_8":"90","col_9":"140","col_10":"200"},"row_6":{"col_1":"25 mV to 50 mV","col_2":"40","col_3":"30","col_4":"20","col_5":"25","col_6":"30","col_7":"35","col_8":"70","col_9":"100","col_10":"150"},"row_7":{"col_1":"50 mV to 100 mV","col_2":"25","col_3":"20","col_4":"15","col_5":"20","col_6":"25","col_7":"30","col_8":"60","col_9":"70","col_10":"100"},"row_8":{"col_1":"100 mV to 200 mV","col_2":"20","col_3":"17","col_4":"13","col_5":"15","col_6":"18","col_7":"25","col_8":"50","col_9":"65","col_10":"75"},"row_9":{"col_1":"200 mV to 500 mV","col_2":"20","col_3":"15","col_4":"9","col_5":"10","col_6":"12","col_7":"17","col_8":"25","col_9":"30","col_10":"44"},"row_10":{"col_1":"500 mV to 1 V","col_2":"20","col_3":"15","col_4":"7","col_5":"10","col_6":"12","col_7":"17","col_8":"25","col_9":"30","col_10":"40"},"row_11":{"col_1":"1.1 V to 3 V","col_2":"15","col_3":"10","col_4":"5","col_5":"7","col_6":"10","col_7":"15","col_8":"20","col_9":"25","col_10":"35"},"row_12":{"col_1":"3.1 V to 5 V","col_2":"20","col_3":"12","col_4":"7","col_5":"10","col_6":"12","col_7":"17","col_8":"25","col_9":"30","col_10":"40"},"row_13":{"col_1":"5.1 V to 10 V","col_2":"25","col_3":"20","col_4":"10","col_5":"12","col_6":"15","col_7":"20","col_8":"30","col_9":"35","col_10":"45"},"row_14":{"col_1":"10.1 V to 20 V","col_2":"27","col_3":"22","col_4":"15","col_5":"17","col_6":"22","col_7":"25","col_8":"30","col_9":"35","col_10":"45"},"row_15":{"col_1":"20.1 V to 30 V","col_2":"30","col_3":"25","col_4":"20","col_5":"25","col_6":"27","col_7":"30","col_8":"35","col_9":"40","col_10":"50"},"row_16":{"col_1":"30.1 V to 60 V","col_2":"35","col_3":"27","col_4":"25","col_5":"30","col_6":"30","col_7":"30","col_8":"35","col_9":"40","col_10":"50"},"row_17":{"col_1":"60.1 V to 100 V","col_2":"37","col_3":"30","col_4":"30","col_5":"35","col_6":"40","col_7":"-","col_8":"-","col_9":"-","col_10":"-"},"row_18":{"col_1":"100.1 V to 200 V","col_2":"40","col_3":"35","col_4":"33","col_5":"37","col_6":"45","col_7":"-","col_8":"-","col_9":"-","col_10":"-"},"row_19":{"col_1":"200.1 V to 300 V","col_2":"45","col_3":"37","col_4":"35","col_5":"40","col_6":"50","col_7":"-","col_8":"-","col_9":"-","col_10":"-"},"row_20":{"col_1":"300.1 V to 600 V","col_2":"50","col_3":"40","col_4":"37","col_5":"50","col_6":"60","col_7":"-","col_8":"-","col_9":"-","col_10":"-"},"row_21":{"col_1":"600.1 V to 1000 V","col_2":"55","col_3":"45","col_4":"40","col_5":"60","col_6":"80","col_7":"-","col_8":"-","col_9":"-","col_10":"-"}}
The expanded uncertainties given in this table are expressed in µV/V
Comparison
Frequency : 10 Hz to 1 MHz
Approved on 06 August 2013
Institute service identifier : JV/228
CMC ID : EURAMET-EM-NO-000006UM-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 40 °C to 200 °C
Liquid-in-glass thermometer
Absolute expanded uncertainty : 2.0E-2 °C
Comparison in liquid calibration bath
Stirred liquid bath with LIG adapter : liquid is silicon oil
Approved on 03 November 2009
Institute service identifier : JV/38
CMC ID : EURAMET-T-NO-0000075W-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : -40 °C to 40 °C
Liquid-in-glass thermometer
Absolute expanded uncertainty : 2.0E-2 °C
Comparison in liquid calibration bath
Stirred liquid bath with LIG adapter : liquid is alcohol
Approved on 03 November 2009
Institute service identifier : JV/37
CMC ID : EURAMET-T-NO-0000075V-1
 
Items for defining ITS-90 , Temperature : -189.3442 °C
Long stem SPRT
Absolute expanded uncertainty : 0.85 mK
Fixed points, Argon
Ambient temperature : (22 ± 2) °C
Approved on 20 May 2010
Institute service identifier : JV/10
CMC ID : EURAMET-T-NO-00000760-1
 
Hygrometers , Humidity : -50 °C to 0 °C
Dew-point hygrometer
Absolute expanded uncertainty : 0.3 °C
Measurement against humidity generator
Ambient temperature : 21 °C
Approved on 03 November 2009
Institute service identifier : JV/40
CMC ID : EURAMET-T-NO-0000075T-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 419.527 °C
Noble metal thermocouple (type S, R)
Absolute expanded uncertainty : 0.2 °C
Fixed point cell
Heat pipe furnace
Approved on 19 March 2012
Institute service identifier : JV/32
CMC ID : EURAMET-T-NO-0000076A-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 660.323 °C
Noble metal thermocouple (type S, R)
Absolute expanded uncertainty : 0.2 °C
Fixed point cell
Heat pipe furnace
Approved on 19 March 2012
Institute service identifier : JV/31
CMC ID : EURAMET-T-NO-00000769-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 961.78 °C
Noble metal thermocouple (type S, R)
Absolute expanded uncertainty : 0.3 °C
Fixed point cell
Heat pipe furnace
Approved on 19 March 2012
Institute service identifier : JV/30
CMC ID : EURAMET-T-NO-00000768-1
 
Properties of detectors , Responsivity, spectral, power
Broadband detector
Relative expanded uncertainty : 2.4 %
Double grating monochromator, cut -in filters
Wavelength range : 930 nm to 1600 nm
Bandwith : 4 nm
Power : 1 µW to 6 µW
Approved on 29 September 2001
CMC ID : EURAMET-PR-NO-000004YD-1
 
AC voltage (up to the MHz range) , AC voltage up to 1000 V: sources : 1.00E-3 V to 1000 V
Source
Relative expanded uncertainty : 3.0E1 µV/V to 8.0E3 µV/V
{"row_1":{"col_1":"","col_2":"10 Hz to 30 Hz","col_3":"31 Hz to 300 Hz","col_4":"301 Hz to 10 kHz","col_5":"11 kHz to 50 kHz","col_6":"51 kHz to 100 kHz","col_7":"101 kHz to 200 kHz","col_8":"201 kHz to 500 kHz","col_9":"501 kHz to 700 kHz","col_10":"701 kHz to 1 MHz"},"row_2":{"col_1":"1 mV to 2 mV","col_2":"2100","col_3":"2000","col_4":"2000","col_5":"2200","col_6":"2400","col_7":"3000","col_8":"3600","col_9":"5500","col_10":"8000"},"row_3":{"col_1":"10 mV to 20 mV","col_2":"310","col_3":"300","col_4":"300","col_5":"400","col_6":"500","col_7":"700","col_8":"900","col_9":"1800","col_10":"3100"},"row_4":{"col_1":"100 mV to 200 mV","col_2":"100","col_3":"80","col_4":"55","col_5":"70","col_6":"110","col_7":"150","col_8":"320","col_9":"850","col_10":"1500"},"row_5":{"col_1":"500 mV to 3 V","col_2":"90","col_3":"70","col_4":"40","col_5":"45","col_6":"50","col_7":"70","col_8":"100","col_9":"200","col_10":"700"},"row_6":{"col_1":"1 V / 2 V","col_2":"85","col_3":"40","col_4":"30","col_5":"30","col_6":"35","col_7":"50","col_8":"70","col_9":"140","col_10":"300"},"row_7":{"col_1":"3.1 V to 20 V","col_2":"100","col_3":"80","col_4":"75","col_5":"75","col_6":"80","col_7":"95","col_8":"110","col_9":"250","col_10":"600"},"row_8":{"col_1":"10 V / 20 V","col_2":"70","col_3":"50","col_4":"45","col_5":"45","col_6":"50","col_7":"70","col_8":"90","col_9":"130","col_10":"320"},"row_9":{"col_1":"20.1 to 200 V","col_2":"100","col_3":"90","col_4":"85","col_5":"90","col_6":"100","col_7":"-","col_8":"-","col_9":"-","col_10":"-"},"row_10":{"col_1":"100 V / 200 V","col_2":"75","col_3":"60","col_4":"55","col_5":"65","col_6":"70","col_7":"-","col_8":"-","col_9":"-","col_10":"-"},"row_11":{"col_1":"200.1 V to 1000 V","col_2":"120","col_3":"105","col_4":"90","col_5":"200","col_6":"300","col_7":"-","col_8":"-","col_9":"-","col_10":"-"},"row_12":{"col_1":"700 V","col_2":"85","col_3":"70","col_4":"65","col_5":"110","col_6":"150","col_7":"-","col_8":"-","col_9":"-","col_10":"-"},"row_13":{"col_1":"1000 V","col_2":"100","col_3":"90","col_4":"85","col_5":"150","col_6":"250","col_7":"-","col_8":"-","col_9":"-","col_10":"-"}}
The expanded uncertainties given in this table are expressed in µV/V
AC/DC transfer standard, ACV-meter
Frequency : 10 Hz to 1 MHz
Approved on 06 August 2013
Institute service identifier : JV/230
CMC ID : EURAMET-EM-NO-000006UO-1
 
DC resistance , DC resistance standards and sources: multiple ranges : 1 Ω to 1.00E8 Ω
Multifunction calibrator
Relative expanded uncertainty : 4 µΩ/Ω to 2.4E1 µΩ/Ω
{"row_1":{"col_1":"","col_2":"Expanded uncertainty"},"row_2":{"col_1":"1 Ω to 1.9 Ω","col_2":"9"},"row_3":{"col_1":"10 Ω to 19E+03 Ω","col_2":"4"},"row_4":{"col_1":"1E+05 Ω to 1.9E+05 Ω","col_2":"7"},"row_5":{"col_1":"1E+06 Ω to 1.9E+06 Ω","col_2":"11"},"row_6":{"col_1":"1E+07 Ω","col_2":"12"},"row_7":{"col_1":"1.9E+07 Ω","col_2":"15"},"row_8":{"col_1":"1E+08 Ω","col_2":"24"}}
The expanded uncertainties given in this table are expressed in µΩ/Ω
Automatic DCC bridge / automatic high resistance ratio bridge
Power : 1 mW to 10 mW
Approved on 06 August 2013
Institute service identifier : JV/194a
CMC ID : EURAMET-EM-NO-000006V3-1
 
DC voltage (up to 1100 V) , DC voltage sources: low values : 0.1 V to 10 V
Multifunction calibrator
Relative expanded uncertainty : 0.8 µV/V to 1.2E1 µV/V
{"row_1":{"col_1":"","col_2":"Expanded uncertainty"},"row_2":{"col_1":"0.1 V","col_2":"9"},"row_3":{"col_1":"1 V","col_2":"1"},"row_4":{"col_1":"10 V","col_2":"0.8"},"row_5":{"col_1":"0.1 V to 10 V","col_2":"2 + 1/<I>U</I>"}}
The voltage U is expressed in volt V
The expanded uncertainties given in this table are expressed in µV/V
Comparison with standard using a voltage divider
Approved on 06 August 2013
Institute service identifier : JV/2a
CMC ID : EURAMET-EM-NO-000006UU-1
 
DC current (up to 100 A) , DC current sources: low values : 1.00E-4 A
Multifunction calibrator
Relative expanded uncertainty : 9 µA/A
Voltage measurement across a shunt
Approved on 06 August 2013
Institute service identifier : JV/210
CMC ID : EURAMET-EM-NO-000006V5-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 200 °C to 550 °C
Digital thermometer: IPRT and SPRT
Absolute expanded uncertainty : 1.8E1 mK to 3.8E1 mK
Comparison in liquid calibration bath
Stirred liquid bath : Liquid: melted salt
Approved on 10 November 2015
Institute service identifier : JV/39
CMC ID : EURAMET-T-NO-0000075J-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 550 °C to 660 °C
Digital thermometer: IPRT and SPRT
Absolute expanded uncertainty : 3.8E1 mK to 4.5E1 mK
Comparison in heat pipe
Isothermal enclosure : Heat pipe furnace
Approved on 10 November 2015
Institute service identifier : JV/42
CMC ID : EURAMET-T-NO-0000075K-1
 
Hygrometers , Humidity : 10 %rh to 95 %rh
Relative humidity sensors
Absolute expanded uncertainty : 0.4 %rh to 1.2 %rh
Calibration by comparison
Temperature : -10 °C to 70 °C
Approved on 25 February 2020
CMC ID : EURAMET-T-NO-00000K8E-1
 
Linear dimensions , Length instruments : 0.3 m to 50 m
EDM, "Electronic (Laser) Distance Measuring" instrument
Absolute expanded uncertainty : 0.62 mm to 0.8 mm
Q[0.62 mm, 10E-06 L]
Comparison to master length interferometer
Reference to the back plane of instrument.
Approved on 17 July 2014
Institute service identifier : JV-NL-LEN-018
CMC ID : EURAMET-L-NO-00000AO6-2
 
Radiations of the mise en pratique , Laser radiations : 633 nm
Frequency stabilized laser: vacuum wavelength
Absolute expanded uncertainty : 4.0E-2 fm
Optical beat frequency, bandwidth 1 GHz
Approved on 21 October 2001
Institute service identifier : JV-NL-LEN-002
CMC ID : EURAMET-L-NO-00000ANZ-2
 
Radiations of the mise en pratique , Laser radiations : 633 nm
Frequency stabilized laser: vacuum wavelength
Relative expanded uncertainty : 5.0E-11
Optical beat frequency, bandwidth 1 GHz
Approved on 21 October 2001
Institute service identifier : JV-NL-LEN-002
CMC ID : EURAMET-L-NO-00000AOA-2
 

Select Metrology Area

General physics
Acoustics, Ultrasound, Vibration

This metrology area covers the fields of acoustics, ultrasound and vibration.

Electricity and Magnetism

This metrology area covers the fields of DC and AC measurements, impedance, electric and magnetic fields, radiofrequencies and measurements on materials.

Length

This metrology area covers the fields of laser frequencies and dimensional metrology.

Mass and related quantities

This metrology area covers the fields of mass standards, force, pressure, density, hardness, torque, gravity and viscosity.

Chemistry and Biology
Chemistry and Biology

This metrology area covers all fields relevant to chemistry and biology.

Ionizing radiation
Ionizing Radiation

This metrology area covers the fields of dosimetry, radioactivity and neutron measurements.