Results for: National Metrology Centre, Agency for Science, Technology and Research
Filters
CMC Quick search Filters close;
CMC Area
To filter, either use the tick boxes or click on the underscored item
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Singapore, NMC, A*STAR (National Metrology Centre, Agency for Science, Technology and Research)
Spectral properties of materials ,
Transmittance, regular, spectral
:
1.00E-3 to 1
Spectrally-neutral material
Relative expanded uncertainty
:
1.5 % to 0.5 %
Scanning spectrometer
Wavelength range : 350 nm to 860 nm
Bandwidth : 1 nm to 5 nm
Specific measurement conditions : normal incidence
Approved on 16 January 2011
CMC ID :
APMP-PR-SG-0000099N-1
Fibre optics ,
Wavelength
:
1520 nm to 1620 nm
Optical spectrum analyser
Absolute expanded uncertainty
:
2.0E-2 nm
Tunable laser source and wavelength meter
Optical power : -30 dBm to +3 dBm
Approved on 16 January 2011
CMC ID :
APMP-PR-SG-0000098Y-1
Properties of detectors ,
Responsivity, spectral, irradiance
Broad band detector
Relative expanded uncertainty
:
6 %
Transfer standard silicon photodiode, UV source and filter
Wavelength range : 365 nm
Bandwidth : 10 nm
Irradiance level : 150 W/m2 to 3000 W/m2
Approved on 16 January 2011
CMC ID :
APMP-PR-SG-0000099F-1
Items for defining ITS-90 ,
Temperature
:
29.7646 °C
SPRT
Absolute expanded uncertainty
:
0.65 mK
Gallium melting point cell
Temperature controlled furnace
Approved on 27 February 2018
Institute service identifier : NMC, A*STAR/TM050
CMC ID :
APMP-T-SG-000009SC-1
Items for defining ITS-90 ,
Temperature
:
660.323 °C
SPRT
Absolute expanded uncertainty
:
6 mK
Aluminum freezing point cell
Temperature controlled furnace
Approved on 09 February 2015
Institute service identifier : NMC, A*STAR/TM047
CMC ID :
APMP-T-SG-000009SF-1
Items for defining ITS-90 ,
Temperature
:
-189.3442 °C
SPRT
Absolute expanded uncertainty
:
0.77 mK
Argon triple point cell
Temperature controlled Dewar
Approved on 27 February 2018
Institute service identifier : NMC, A*STAR/TM048
CMC ID :
APMP-T-SG-000009SA-1
Items for defining ITS-90 ,
Temperature
:
419.527 °C
SPRT
Absolute expanded uncertainty
:
2.2 mK
Zinc freezing point cell
Temperature controlled furnace
Approved on 27 February 2018
Institute service identifier : NMC, A*STAR/TM053
CMC ID :
APMP-T-SG-000009SE-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
660.323 °C
Thermocouple (type S, R &B)
Absolute expanded uncertainty
:
0.11 °C
Calibration at aluminum freezing point
Approved on 16 March 2011
Institute service identifier : NMC, A*STAR/TM022
CMC ID :
APMP-T-SG-000009SQ-1
Items for defining ITS-90 ,
Temperature
:
-189.3442 °C to 0.01 °C
SPRT
Absolute expanded uncertainty
:
1 mK
W4 (Ar-Hg-TPW)
Temperature controlled Bath
Approved on 27 February 2018
Institute service identifier : NMC, A*STAR/TM055
CMC ID :
APMP-T-SG-000009SG-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
419.527 °C
Thermocouple (type B)
Absolute expanded uncertainty
:
0.1 °C
Calibration at zinc freezing point
Approved on 16 March 2011
Institute service identifier : NMC, A*STAR/TM021
CMC ID :
APMP-T-SG-000009SP-1
Items for defining ITS-90 ,
Temperature
:
29.7646 °C
Gallium melting point cell
Absolute expanded uncertainty
:
0.4 mK
Comparison with reference cell(s)
Temperature controlled furnace
Approved on 09 February 2015
CMC ID :
APMP-T-SG-000009S5-1
Items for defining ITS-90 ,
Temperature
:
419.527 °C
Zinc freezing point cell
Absolute expanded uncertainty
:
1 mK
Comparison with reference cell(s)
Temperature controlled furnace
Approved on 09 February 2015
CMC ID :
APMP-T-SG-000009S7-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
960 °C to 1400 °C
Disappearing filament pyrometer
Absolute expanded uncertainty
:
4 °C
Calibration by reference strip lamp
Approved on 28 May 2012
Institute service identifier : NMC, A*STAR/TM045
CMC ID :
APMP-T-SG-000009UK-1
Items for defining ITS-90 ,
Temperature
:
800 °C to 1600 °C
Radiation thermometer
Absolute expanded uncertainty
:
0.24 °C to 0.84 °C
Uncertainty table
{"row_1":{"col_1":"","col_2":"0.9 µm"},"row_2":{"col_1":"800 °C","col_2":"0.24"},"row_3":{"col_1":"900 °C","col_2":"0.32"},"row_4":{"col_1":"1000 °C","col_2":"0.40"},"row_5":{"col_1":"1100 °C","col_2":"0.46"},"row_6":{"col_1":"1200 °C","col_2":"0.52"},"row_7":{"col_1":"1300 °C","col_2":"0.59"},"row_8":{"col_1":"1400 °C","col_2":"0.66"},"row_9":{"col_1":"1500 °C","col_2":"0.75"},"row_10":{"col_1":"1600 °C","col_2":"0.84"}}
Direct scale realization Measurement of relative spectral responsivity
Wavelength : 0.9 µm
Approved on 28 May 2012
Institute service identifier : NMC, A*STAR/TM032
CMC ID :
APMP-T-SG-000009U9-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-196 °C
IPRT
Absolute expanded uncertainty
:
1.6E1 mK
Calibration by comparison at temperature points
Temperature controlled bath : Liquid nitrogen comparator
CMC uncertainty does not include DUT hysteresis
Approved on 27 February 2018
Institute service identifier : NMC, A*STAR/TM059
CMC ID :
APMP-T-SG-000009SX-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-90 °C to 80 °C
Liquid-in-glass thermometer
Absolute expanded uncertainty
:
0.5 °C
Partial immersion, 2 °C graduation. Comparison method
Temperature controlled bath : Ultra low temperature bath, alcohol, water bath
Approved on 27 February 2018
Institute service identifier : NMC, A*STAR/TM015
CMC ID :
APMP-T-SG-000009TZ-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-75 °C to 80 °C
Liquid-in-glass thermometer
Absolute expanded uncertainty
:
1.0E-2 °C
Total immersion, 0.01 °C graduation. Comparison method
Temperature controlled bath : alcohol, water bath
Approved on 16 March 2011
Institute service identifier : NMC, A*STAR/TM015
CMC ID :
APMP-T-SG-000009T2-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
80 °C to 250 °C
Liquid-in-glass thermometer
Absolute expanded uncertainty
:
0.15 °C
Total immersion, 0.5 °C graduation. Comparison method
Temperature controlled bath : oil bath
Approved on 16 March 2011
Institute service identifier : NMC, A*STAR/TM016
CMC ID :
APMP-T-SG-000009TD-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
80 °C to 250 °C
Liquid-in-glass thermometer
Absolute expanded uncertainty
:
0.3 °C
Total immersion, 1 °C graduation. Comparison method
Temperature controlled bath : oil bath
Approved on 16 March 2011
Institute service identifier : NMC, A*STAR/TM016
CMC ID :
APMP-T-SG-000009TF-1
Hygrometers ,
Dew-point temperature
:
-60 °C to -50 °C
Dew-point hygrometer
Absolute expanded uncertainty
:
0.13 °C to 0.1 °C
Comparison
Pressure : ambient
Approved on 28 June 2007
CMC ID :
APMP-T-SG-000009U2-1