In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Serbia, DMDM (Directorate of Measures and Precious Metals)
Matrix or material Analyte or component Dissemination range of measurement capability Range of certified values in reference materials
Amount-of-substance fraction in nmol/mol Absolute expanded uncertainty in nmol/mol Amount-of-substance fraction in Absolute expanded uncertainty in
purified air ozone 0 to 1000
  • Uncertainty equation CMC : Q[1.1, 0.022 x(O3)]

    The uncertainty is expressed in nmol/mol

  • Mechanism for measurement service delivery : Instrumental comparison against primary UV photometer Calibration
  • The notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a2 + b2]1/2
  • Approved on 28 June 2016
  • Institute service identifier : GMH-RU-H09
Photometry , Luminous intensity : 1 cd to 1.00E4 cd
Tungsten lamp
Relative expanded uncertainty : 2 %
Photometric bench, reference lamps / photometer
Colour temperature : 2353 K, 2800 K and 2856 K
Approved on 30 September 2010
Measurement microphones , Pressure sensitivity level
Measurement microphone type LS1P
Absolute expanded uncertainty : 8.0E-2 dB
IEC 61094-2:1992
Frequency : 63 Hz to 3.15 kHz
Approved on 27 February 2014
Institute service identifier : DMDM/1.1
Measurement microphones , Pressure sensitivity level
Measurement microphone type LS1P
Absolute expanded uncertainty : 9.0E-2 dB
IEC 61094-2:1992
Frequency : 4 kHz
Approved on 27 February 2014
Institute service identifier : DMDM/1.1
Measurement microphones , Pressure sensitivity level
Measurement microphone type LS2P
Absolute expanded uncertainty : 0.15 dB
IEC 61094-2:1992
Frequency : 16 kHz
Approved on 27 February 2014
Institute service identifier : DMDM/1.2
Measurement microphones , Pressure sensitivity level
Measurement microphone type LS1P
Absolute expanded uncertainty : 0.14 dB
IEC 61094-2:1992
Frequency : 8 kHz
Approved on 27 February 2014
Institute service identifier : DMDM/1.1
Frequency , Frequency : 10 MHz
Local frequency standard
Relative expanded uncertainty : 1.0E-13 Hz/Hz
Phase time measurement
Averaging time : 1 day
Approved on 10 March 2008
Institute service identifier : DMDM/TF-2.1.1b
Items for disseminating ITS-90 and PLTS 2000 , Temperature : -80 °C to -40 °C
Liquid-in-glass thermometer resolution 0.1 °C
Absolute expanded uncertainty : 3.5E-2 °C
Calibration by comparison
Halocarbon bath
Approved on 19 March 2012
Institute service identifier : DMDM/T06-11
Items for disseminating ITS-90 and PLTS 2000 , Temperature : -40 °C to 20 °C
Liquid-in-glass thermometer resolution 0.1 °C
Absolute expanded uncertainty : 3.3E-2 °C
Calibration by comparison
Ethanol bath
Approved on 03 November 2009
Institute service identifier : DMDM/T06-4
Items for disseminating ITS-90 and PLTS 2000 , Temperature : -40 °C to 20 °C
Liquid-in-glass thermometer resolution 1 °C
Absolute expanded uncertainty : 0.35 °C
Calibration by comparison
Ethanol bath
Approved on 03 November 2009
Institute service identifier : DMDM/T06-7
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 90 °C to 250 °C
Temperature sensor with display unit
Absolute expanded uncertainty : 2.0E1 mK
Calibration by comparison
Oil bath
Approved on 19 March 2012
Institute service identifier : DMDM/T05-4
Items for disseminating ITS-90 and PLTS 2000 , Temperature : -80 °C to -40 °C
Platinum Resistance Thermometer
Absolute expanded uncertainty : 1.0E1 mK
Calibration by comparison
Halocarbon bath
Approved on 19 March 2012
Institute service identifier : DMDM/T02-4
Items for disseminating ITS-90 and PLTS 2000 , Temperature : -40 °C to 20 °C
Platinum Resistance Thermometer
Absolute expanded uncertainty : 9.0E-3 °C
Calibration by comparison
Ethanol bath
Approved on 03 November 2009
Institute service identifier : DMDM/T02-1
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 300 °C to 600 °C
Base-metal thermocouple
Absolute expanded uncertainty : 0.51 °C to 0.87 °C
Calibration by comparison
Comparison furnace
Approved on 19 March 2012
Institute service identifier : DMDM/T04-5
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 600 °C to 1000 °C
Base-metal thermocouple
Absolute expanded uncertainty : 0.87 °C to 1.5 °C
Calibration by comparison
Comparison furnace
Approved on 19 March 2012
Institute service identifier : DMDM/T04-6
Measurement microphones , Pressure sensitivity level
Measurement microphone type LS2P
Absolute expanded uncertainty : 0.46 dB
IEC 61094-2:1992
Frequency : 25 kHz
Approved on 27 February 2014
Institute service identifier : DMDM/1.2
Fluid flow , Volume of liquid : 1 mL to 1.00E4 mL
Glassware equipment (Pipettes, flasks, burettes, cylinders)
Relative expanded uncertainty : 1.5E-2 %
Gravimetric
Liquid : water
Reference temperature : 20 °C
Approved on 17 November 2017
Institute service identifier : DMDM/RU-Z03
Mass , Conventional mass : 1 mg to 100 mg
Mass standard
Absolute expanded uncertainty : 1.3 µg to 2 µg
Comparison in air
Temperature : (20 to 24) °C with max ± 0.3 °C/h and max ± 0.5 °C/12 h
Relative humidity : 40 % to 60 % with max ± 5 % per 4 hours
Pressure : 995 hPa to 1020 hPa
Approved on 04 May 2010
Mass , Conventional mass : 100 kg to 500 kg
Mass standard
Absolute expanded uncertainty : 0.5 g to 8 g
Comparison in air
Temperature : (20 to 24) °C with max ± 1.5 °C/h and max ± 2 °C/12 h
Relative humidity : 40 % to 60 % with max ± 10 % per 4 hours
Pressure : 995 hPa to 1020 hPa
Approved on 04 May 2010
Fluid flow , Volume of liquid : 5 L to 5000 L
Proving tanks
Relative expanded uncertainty : 3.0E-2 %
Volumetric
Liquid : water
Reference temperature : 20 °C
Approved on 17 November 2017
Institute service identifier : DMDM/RU-Z06

Select Metrology Area

General physics
Acoustics, Ultrasound, Vibration

This metrology area covers the fields of acoustics, ultrasound and vibration.

Electricity and Magnetism

This metrology area covers the fields of DC and AC measurements, impedance, electric and magnetic fields, radiofrequencies and measurements on materials.

Length

This metrology area covers the fields of laser frequencies and dimensional metrology.

Mass and related quantities

This metrology area covers the fields of mass standards, force, pressure, density, hardness, torque, gravity and viscosity.

Chemistry and Biology
Chemistry and Biology

This metrology area covers all fields relevant to chemistry and biology.

Ionizing radiation
Ionizing Radiation

This metrology area covers the fields of dosimetry, radioactivity and neutron measurements.