Results for: Denmark "HBK-DPLA"
Filters
CMC Quick search Filters close;
CMC Area
To filter, either use the tick boxes or click on the underscored item
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Denmark, HBK-DPLA (Hottinger Brüel & Kjaer A/S - Danish Primary Laboratory for Acoustics Section)
Measurement microphones ,
Pressure sensitivity level
Measurement microphone type LS2
Absolute expanded uncertainty
:
8.0E-2 dB
IEC 61094-2:1992
Frequency : 16.1 kHz to 20.0 kHz
Approved on 27 December 2007
KCDB ID :
EURAMET-AUV-DK-000009PM-1
Measurement microphones ,
Pressure sensitivity level
Measurement microphone type LS1
Absolute expanded uncertainty
:
8.0E-2 dB
IEC 61094-2:1992
Frequency : 8.1 kHz to 10 kHz
Approved on 27 December 2007
KCDB ID :
EURAMET-AUV-DK-000009PF-1
Measurement microphones ,
Pressure sensitivity level
Measurement microphone type LS2
Absolute expanded uncertainty
:
8.0E-2 dB
IEC 61094-2:1992
Frequency : 25 Hz to 49 Hz
Approved on 27 December 2007
KCDB ID :
EURAMET-AUV-DK-000009PH-1
Linear vibration ,
Voltage sensitivity (modulus)
Acceleration measuring chain
Relative expanded uncertainty
:
0.5 %
ISO 16063-11
Frequency : 0.1 Hz to <0.2 Hz
Approved on 17 January 2018
KCDB ID :
EURAMET-AUV-DK-000009Q0-1
Linear vibration ,
Voltage sensitivity (modulus)
Acceleration measuring chain
Relative expanded uncertainty
:
0.4 %
ISO 16063-11
Frequency : 0.2 Hz to 5 kHz
Approved on 17 January 2018
KCDB ID :
EURAMET-AUV-DK-000009Q1-1
Linear vibration ,
Charge sensitivity (modulus)
Accelerometer
Relative expanded uncertainty
:
0.5 %
ISO 16063-11
Frequency : 0.1 Hz to <0.2 Hz
Approved on 17 January 2018
KCDB ID :
EURAMET-AUV-DK-000009Q3-1
Measurement microphones ,
Pressure sensitivity level
Measurement microphone type LS2
Absolute expanded uncertainty
:
0.14 dB
IEC 61094-2:1992
Frequency : 20.1 kHz to 25.2 kHz
Approved on 27 December 2007
KCDB ID :
EURAMET-AUV-DK-000009PN-1
Measurement microphones ,
Pressure sensitivity level
Measurement microphone type LS1
Absolute expanded uncertainty
:
5.0E-2 dB
IEC 61094-2:1992
Frequency : 25 Hz to 49 Hz
Approved on 27 December 2007
KCDB ID :
EURAMET-AUV-DK-000009PA-1
Linear vibration ,
Charge sensitivity (phase shift)
:
-180 ° to 180 °
Accelerometer
Absolute expanded uncertainty
:
0.4 °
ISO 16063-11
Frequency : 0.1 Hz to <0.2 Hz
Approved on 17 January 2018
KCDB ID :
EURAMET-AUV-DK-000009P6-1
Linear vibration ,
Charge sensitivity (phase shift)
:
-180 ° to 180 °
Accelerometer
Absolute expanded uncertainty
:
1 °
ISO 16063-11
Frequency : >5 kHz to 10 kHz
Approved on 17 January 2018
KCDB ID :
EURAMET-AUV-DK-000009P8-1
Measurement microphones ,
Pressure sensitivity level
Measurement microphone type LS2
Absolute expanded uncertainty
:
5.0E-2 dB
IEC 61094-2:1992
Frequency : 12.7 kHz to 16.0 kHz
Approved on 27 December 2007
KCDB ID :
EURAMET-AUV-DK-000009PL-1
Measurement microphones ,
Pressure sensitivity level
Measurement microphone type LS1
Absolute expanded uncertainty
:
6.0E-2 dB
IEC 61094-2:1992
Frequency : 19 Hz to 24 Hz
Approved on 27 December 2007
KCDB ID :
EURAMET-AUV-DK-000009P9-1
Measurement microphones ,
Pressure sensitivity level
Measurement microphone type LS2
Absolute expanded uncertainty
:
4.0E-2 dB
IEC 61094-2:1992
Frequency : 50 Hz to 99 Hz
Approved on 27 December 2007
KCDB ID :
EURAMET-AUV-DK-000009PI-1
Measurement microphones ,
Pressure sensitivity level
Measurement microphone type LS1
Absolute expanded uncertainty
:
3.0E-2 dB
IEC 61094-2:1992
Frequency : 50 Hz to 99 Hz
Approved on 27 December 2007
KCDB ID :
EURAMET-AUV-DK-000009PB-1
Linear vibration ,
Voltage sensitivity (modulus)
Acceleration measuring chain
Relative expanded uncertainty
:
0.6 %
ISO 16063-11
Frequency : >5 kHz to 10 kHz
Approved on 17 January 2018
KCDB ID :
EURAMET-AUV-DK-000009Q2-1
Linear vibration ,
Charge sensitivity (modulus)
Accelerometer
Relative expanded uncertainty
:
0.6 %
ISO 16063-11
Frequency : >5 kHz to 10 kHz
Approved on 17 January 2018
KCDB ID :
EURAMET-AUV-DK-000009Q5-1
Linear vibration ,
Voltage sensitivity (phase shift)
:
-180 ° to 180 °
Acceleration measuring chain
Absolute expanded uncertainty
:
0.3 °
ISO 16063-11
Frequency : 0.1 Hz to 5 kHz
Approved on 17 January 2018
KCDB ID :
EURAMET-AUV-DK-000009P4-1
Linear vibration ,
Charge sensitivity (phase shift)
:
-180 ° to 180 °
Accelerometer
Absolute expanded uncertainty
:
0.3 °
ISO 16063-11
Frequency : 0.2 Hz to 5 kHz
Approved on 17 January 2018
KCDB ID :
EURAMET-AUV-DK-000009P7-1
Measurement microphones ,
Pressure sensitivity level
Measurement microphone type LS2
Absolute expanded uncertainty
:
4.0E-2 dB
IEC 61094-2:1992
Frequency : 10.1 kHz to 12.6 kHz
Approved on 27 December 2007
KCDB ID :
EURAMET-AUV-DK-000009PK-1
Measurement microphones ,
Pressure sensitivity level
Measurement microphone type LS1
Absolute expanded uncertainty
:
5.0E-2 dB
IEC 61094-2:1992
Frequency : 4.1 kHz to 8.0 kHz
Approved on 27 December 2007
KCDB ID :
EURAMET-AUV-DK-000009PE-1