Results for: Croatia
Filters
CMC Quick search Filters close;
?
Added keywords act as an additional filter. Keywords separated by a space will be treated by the logical function AND automatically. You may also use OR. Keyword searches also find close matches. Placing quotation marks (") before and after the search word, e.g. "word", will allow for exact matches.

In case of multiple search, please use ADVANCED SEARCH.
73 results
Reset all
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Croatia, DHMZ-SOUL (Meteorological and Hydrological Service/Calibration Laboratory)
Matrix or material Analyte or component Dissemination range of measurement capability Range of certified values in reference materials
Amount-of-substance fraction in nmol/mol Absolute expanded uncertainty in nmol/mol Amount-of-substance fraction in % Absolute expanded uncertainty in %
purified air ozone 0 to 1000 1.1 to 2.3E1
  • Uncertainty equation CMC : Q[1.1, 0.02321 x(O3)] The expanded uncertainty is expressed in nmol/mol.
    The CCQM.O3.2019 ozone absorption cross-section value and uncertainty is used.
  • Mechanism for measurement service delivery : Calibration of analysers and calibrators
  • Approved on 29 June 2025
  • CMC ID : EURAMET-QM-HR-00000QHR-1
     
Matrix or material Analyte or component Dissemination range of measurement capability Range of certified values in reference materials
Amount of substance fraction in nmol/mol Absolute expanded uncertainty in nmol/mol Amount of substance fraction in
purified air ozone 0 to 1000 1.1 to 2.3E1
  • Uncertainty equation CMC : Q[1.1, 0.023 x (O3)] O3: concentration of ozone in nmol/mol
  • Mechanism for measurement service delivery : Calibration of analysers and calibrators
  • Approved on 22 October 2020
  • Institute service identifier : SOUL-1
  • CMC ID : EURAMET-QM-HR-00000LWD-1
     
Croatia, DZM (State Office for Metrology)
Mass , Mass : 0.1 g to 1 g
Mass standards
Absolute expanded uncertainty : 1.5 µg to 3 µg
Comparison in air
OIML R 111-1
Temperature : 20°C -23°C 
Humidity : 40% -60%
Approved on 25 February 2023
Institute service identifier : DZM-LMG
CMC ID : EURAMET-M-HR-00000O2U-1
 
Mass , Mass : 0.1 kg to 1 kg
Mass standards
Absolute expanded uncertainty : 1.5E1 µg to 1.5E2 µg
Comparison in air
OIML R 111-1
Temperature : 20°C -23°C 
Humidity : 40% -60%
Approved on 25 February 2023
Institute service identifier : DZM-LMG
CMC ID : EURAMET-M-HR-00000O2X-1
 
Mass , Mass : 1 mg to 100 mg
Mass standards
Absolute expanded uncertainty : 0.6 µg to 1.5 µg
Comparison in air
OIML R 111-1
Temperature : 20°C -23°C 
Humidity : 40% -60%
Approved on 25 February 2023
Institute service identifier : DZM-LMG
CMC ID : EURAMET-M-HR-00000O2T-1
 
Mass , Mass : 1 g to 10 g
Mass standards
Absolute expanded uncertainty : 3 µg to 6 µg
Comparison in air
OIML R 111-1
Temperature : 20°C -23°C 
Humidity : 40% -60%
Approved on 25 February 2023
Institute service identifier : DZM-LMG
CMC ID : EURAMET-M-HR-00000O2V-1
 
Mass , Mass : 10 g to 100 g
Mass standards
Absolute expanded uncertainty : 6 µg to 1.5E1 µg
Comparison in air
OIML R 111-1
Temperature : 20°C -23°C 
Humidity : 40% -60%
Approved on 25 February 2023
Institute service identifier : DZM-LMG
CMC ID : EURAMET-M-HR-00000O2W-1
 
Croatia, FSB-LPM (FSB - Laboratory for Process Measurements)
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 0 °C to 95 °C
Platinum Resistance Thermometer
Absolute expanded uncertainty : 1.0E1 mK
Comparison with SPRT
Water bath
Approved on 03 November 2009
Institute service identifier : FSB-LPM/2.2.2
CMC ID : EURAMET-T-HR-00000D5O-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 0 °C to 95 °C
Noble metal thermocouple
Absolute expanded uncertainty : 0.25 K
Comparison with SPRT
Water bath
Approved on 03 November 2009
Institute service identifier : FSB-LPM/2.3.1
CMC ID : EURAMET-T-HR-00000D5U-1
 
Pressure , Absolute pressure : 1500 Pa to 1.40E5 Pa
Absolute pressure gauge
Absolute expanded uncertainty
6.1 + 4E-05p

p pressure in Pa

The uncertainty is expressed in Pa

Gas medium
Uncertainty values range from 6.2 Pa to 12 Pa
Approved on 28 April 2008
CMC ID : EURAMET-M-HR-00000D4Q-1
 
Pressure , Absolute pressure : 1.40E5 Pa to 1.40E6 Pa
Absolute pressure gauge
Absolute expanded uncertainty
4 + 5E-05p

p pressure in Pa

The uncertainty is expressed in Pa

Gas medium
Uncertainty values range from 11 Pa to 80 Pa
Approved on 28 April 2008
CMC ID : EURAMET-M-HR-00000D4R-1
 
Pressure , Gauge pressure : 1.50E5 Pa to 7.00E6 Pa
Pressure balance
Absolute expanded uncertainty
25 + 5E-05p

p pressure in Pa

The uncertainty is expressed in Pa

Oil medium
Uncertainty values range from 32 Pa to 3.5E+02 Pa
Approved on 28 April 2008
CMC ID : EURAMET-M-HR-00000D4W-1
 
Items for defining ITS-90 , Temperature : 660.323 °C
SPRT (long stem)
Absolute expanded uncertainty : 1.2E1 mK
Fixed point
Ambient temperature : 23 °C
Approved on 17 January 2013
Institute service identifier : FSB-LPM/1.3.2 Al
CMC ID : EURAMET-T-HR-00000D6H-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 90 °C to 180 °C
Temperature sensor with display unit
Absolute expanded uncertainty : 1.2E1 mK
Comparison with SPRT
Oil bath
Approved on 03 November 2009
Institute service identifier : FSB-LPM/2.7
CMC ID : EURAMET-T-HR-00000D68-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : -40 °C to 0 °C
Base metal thermocouple
Absolute expanded uncertainty : 0.2 K
Comparison with SPRT
Glycol bath
Approved on 03 November 2009
Institute service identifier : FSB-LPM/2.3.2
CMC ID : EURAMET-T-HR-00000D5Z-1
 
Items for defining ITS-90 , Temperature : 660.323 °C
Freezing point of Aluminium
Absolute expanded uncertainty : 1.2E1 mK
Direct comparison
Approved on 17 January 2013
Institute service identifier : FSB-LPM/1.1.1 Al
CMC ID : EURAMET-T-HR-00000D6G-1
 
Hygrometers , Humidity : -70 °C to -50 °C
Dew-point hygrometer
Absolute expanded uncertainty : 0.1 °C
Dew-point generator
Ambient temperature
Approved on 19 March 2012
Institute service identifier : FSB-LPM/3.1
CMC ID : EURAMET-T-HR-00000D6C-1
 
Pressure , Gauge pressure : 1.40E6 Pa to 1.70E7 Pa
Pressure balance
Absolute expanded uncertainty
7E-05p

p pressure in Pa

The uncertainty is expressed in Pa

Gas medium
Uncertainty values range from 98 Pa to 1.2E+03 Pa
Approved on 28 April 2008
CMC ID : EURAMET-M-HR-00000D4V-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 180 °C to 660 °C
Temperature sensor with display unit
Absolute expanded uncertainty : 1.0E2 mK
Comparison with SPRT
Fluidised bath
Approved on 03 November 2009
Institute service identifier : FSB-LPM/2.7
CMC ID : EURAMET-T-HR-00000D69-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 1084.62 °C
Pure metal thermocouple, Pt Pd
Absolute expanded uncertainty : 0.2 K
Calibration at Cu freezing point
3-zone furnace
Approved on 03 November 2009
Institute service identifier : FSB-LPM/2.3.3 Cu
CMC ID : EURAMET-T-HR-00000D64-1
 

Select Metrology Area

General physics
Acoustics, Ultrasound, Vibration

This metrology area covers the fields of acoustics, ultrasound and vibration.

Electricity and Magnetism

This metrology area covers the fields of DC and AC measurements, impedance, electric and magnetic fields, radiofrequencies and measurements on materials.

Length

This metrology area covers the fields of laser frequencies and dimensional metrology.

Mass and related quantities

This metrology area covers the fields of mass standards, force, pressure, density, hardness, torque, gravity and viscosity.

Chemistry and Biology
Chemistry and Biology

This metrology area covers all fields relevant to chemistry and biology.

Ionizing radiation
Ionizing Radiation

This metrology area covers the fields of dosimetry, radioactivity and neutron measurements.