Results for: Chinese Taipei "NRSL/INER"
Filters
CMC Quick search Filters close;
?
Added keywords act as an additional filter. Keywords separated by a space will be treated by the logical function AND automatically. You may also use OR. Keyword searches also find close matches. Placing quotation marks (") before and after the search word, e.g. "word", will allow for exact matches.

In case of multiple search, please use ADVANCED SEARCH.
89 results
Reset all
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Chinese Taipei, NRSL/INER (Institute of Nuclear Energy Research)
Ionizing radiation
Activity per unit mass , Single nuclide, liquid : 1.00E5 Bq/g to 2.50E7 Bq/g
Relative expanded uncertainty 0.8 %
4πe(X)-γ coincidence counting system, balance
Y-88, 2 g solution
Reference standard : 4πe(X)-γ direct measurement, set of standard weights
Source of traceability : INER
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2015
KCDB ID : APMP-RI-TW-00000GOT-1
 
Ionizing radiation
Activity per unit mass , Single nuclide, liquid : 1.00E5 Bq/g to 2.50E7 Bq/g
Relative expanded uncertainty 0.9 %
4πβ-γ coincidence counting system, balance
Ir-192, 2 g solution
Reference standard : 4πβ-γ direct measurement, set of standard weights
Source of traceability : INER
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2013
KCDB ID : APMP-RI-TW-00000GOR-1
 
Ionizing radiation
Activity , Single nuclide, liquid : 3.67E5 Bq to 7.33E8 Bq
Relative expanded uncertainty 2.1 %
High pressure well type ionization chamber, electrometer
Mo-99, 3 mL solution in NRSL standard 5 mL glass ampoule
Reference standard : High pressure well type ionization chamber
Source of traceability : NPL
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2051
KCDB ID : APMP-RI-TW-00000GP5-1
 
Ionizing radiation
Activity , Single nuclide, liquid : 2.33E5 Bq to 4.67E8 Bq
Relative expanded uncertainty 1 %
High pressure well type ionization chamber, electrometer
Au-198, 3 mL solution in NRSL standard 5 mL glass ampoule
Reference standard : High pressure well type ionization chamber
Source of traceability : NPL
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2017
KCDB ID : APMP-RI-TW-00000GOV-1
 
Ionizing radiation
Activity , Single nuclide, liquid : 3.92E4 Bq to 7.83E7 Bq
Relative expanded uncertainty 1.5 %
High pressure well type ionization chamber, electrometer
Br-82, 3 mL solution in NRSL standard 5 mL glass ampoule
Reference standard : High pressure well type ionization chamber
Source of traceability : NPL
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2021
KCDB ID : APMP-RI-TW-00000GOZ-1
 
Ionizing radiation
Activity , Single nuclide, liquid : 1.26E5 Bq to 2.51E8 Bq
Relative expanded uncertainty 1.2 %
High pressure well type ionization chamber, electrometer
K-42, 3 mL solution in NRSL standard 5 mL glass ampoule
Reference standard : High pressure well type ionization chamber
Source of traceability : NPL
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2048
KCDB ID : APMP-RI-TW-00000GP2-1
 
Ionizing radiation
Activity , Single nuclide, liquid : 6.60E4 Bq to 1.32E8 Bq
Relative expanded uncertainty 1 %
High pressure well type ionization chamber, electrometer
Mn-56, 3 mL solution in NRSL standard 5 mL glass ampoule
Reference standard : High pressure well type ionization chamber
Source of traceability : NPL
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2050
KCDB ID : APMP-RI-TW-00000GP4-1
 
Ionizing radiation
Emission rate , Single nuclide, extended area source : 200 s-1 to 1.00E4 s-1
Relative expanded uncertainty 3 %
2π windowless proportional counter
Co-60, electroplate, active area between 20 cm2 and 240 cm2
Reference standard : 2π proportional counter
Source of traceability : INER
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2078
KCDB ID : APMP-RI-TW-00000GOL-1
 
Ionizing radiation
Kerma/rate , Ionization chamber : 2.67E-4 Gy/s to 4.27E-3 Gy/s
Relative expanded uncertainty 1 %
Irradiation in a calibrated field in air
Co-60, 60 TBq (2004), ISO-4037-1
Reference standard : Primary standard ionization chamber
Source of traceability : INER
Approved on 03 October 2006
Institute service identifier : APM-RAD-INER-1001
KCDB ID : APMP-RI-TW-00000GOM-1
 
Ionizing radiation
Activity , Single nuclide, liquid : 1.44E6 Bq to 2.89E9 Bq
Relative expanded uncertainty 2.1 %
High pressure well type ionization chamber, electrometer
Hg-197, 3 mL solution in NRSL standard 5 mL glass ampoule
Reference standard : High pressure well type ionization chamber
Source of traceability : NPL
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2039
KCDB ID : APMP-RI-TW-00000GQM-1
 
Ionizing radiation
Activity , Single nuclide, liquid : 2.12E6 Bq to 4.24E9 Bq
Relative expanded uncertainty 4 %
High pressure well type ionization chamber, electrometer
I-125, 3 mL solution in NRSL standard 5 mL glass ampoule
Reference standard : High pressure well type ionization chamber
Source of traceability : INER
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2043
KCDB ID : APMP-RI-TW-00000GQQ-1
 
Ionizing radiation
Activity , Single nuclide, liquid : 3.06E5 Bq to 6.11E8 Bq
Relative expanded uncertainty 1.5 %
High pressure well type ionization chamber, electrometer
Yb-169, 3 mL solution in NRSL standard 5 mL glass ampoule
Reference standard : High pressure well type ionization chamber
Source of traceability : NPL
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2072
KCDB ID : APMP-RI-TW-00000GOF-1
 
Ionizing radiation
Emission rate , Single nuclide, extended area source : 200 s-1 to 1.00E4 s-1
Relative expanded uncertainty 3 %
2π windowless proportional counter
C-14, electroplate, active area between 20 cm2 and 240 cm2
Reference standard : 2π proportional counter
Source of traceability : INER
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2075
KCDB ID : APMP-RI-TW-00000GOI-1
 
Ionizing radiation
Activity per unit mass , Single nuclide, liquid : 1.00E5 Bq/g to 2.00E6 Bq/g
Relative expanded uncertainty 0.8 %
4πα-γ coincidence counting system, balance
Am-241, 2 g solution
Reference standard : 4πα-γ direct measurement, set of standard weights
Source of traceability : INER
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2001
KCDB ID : APMP-RI-TW-00000GPK-1
 
Ionizing radiation
Activity , Single nuclide, liquid : 1.28E6 Bq to 2.56E9 Bq
Relative expanded uncertainty 2.1 %
High pressure well type ionization chamber, electrometer
Ce-141, 3 mL solution in NRSL standard 5 mL glass ampoule
Reference standard : High pressure well type ionization chamber
Source of traceability : NPL
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2025
KCDB ID : APMP-RI-TW-00000GPW-1
 
Ionizing radiation
Activity , Single nuclide, liquid : 5.83E5 Bq to 1.17E9 Bq
Relative expanded uncertainty 1.2 %
High pressure well type ionization chamber, electrometer
Ce-139, 3 mL solution in NRSL standard 5 mL glass ampoule
Reference standard : High pressure well type ionization chamber
Source of traceability : INER
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2024
KCDB ID : APMP-RI-TW-00000GPV-1
 
Ionizing radiation
Activity , Single nuclide, liquid : 4.49E4 Bq to 8.98E7 Bq
Relative expanded uncertainty 0.8 %
High pressure well type ionization chamber, electrometer
Co-60, 3 mL solution in NRSL standard 5 mL glass ampoule
Reference standard : High pressure well type ionization chamber
Source of traceability : INER
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2029
KCDB ID : APMP-RI-TW-00000GQ0-1
 
Ionizing radiation
Activity , Single nuclide, liquid : 9.33E4 Bq to 1.87E8 Bq
Relative expanded uncertainty 1.4 %
High pressure well type ionization chamber, electrometer
Fe-59, 3 mL solution in NRSL standard 5 mL glass ampoule
Reference standard : High pressure well type ionization chamber
Source of traceability : INER
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2036
KCDB ID : APMP-RI-TW-00000GQJ-1
 
Ionizing radiation
Activity , Single nuclide, liquid : 1.10E6 Bq to 2.20E9 Bq
Relative expanded uncertainty 1 %
High pressure well type ionization chamber, electrometer
Rb-86, 3 mL solution in NRSL standard 5 mL glass ampoule
Reference standard : High pressure well type ionization chamber
Source of traceability : NPL
Approved on 14 February 2005
Institute service identifier : APM-RAD-INER-2056
KCDB ID : APMP-RI-TW-00000GPA-1
 
Ionizing radiation
Ambient dose equivalent rate , Neutron survey meter : 2.70E-6 Sv/h to 1.10E-5 Sv/h
Relative expanded uncertainty 5 %
Calibration with a calibrated neutron source
Am-241/Be-9 source, Am-241/Be-9 source, ISO 8529-1, 3
Reference standard : Calibrated Am-Be source
Source of traceability : NPL
Approved on 03 October 2006
Institute service identifier : APM-RAD-INER-3003
KCDB ID : APMP-RI-TW-00000GPI-1
 

Select Metrology Area

General physics
Acoustics, Ultrasound, Vibration

This metrology area covers the fields of acoustics, ultrasound and vibration.

Electricity and Magnetism

This metrology area covers the fields of DC and AC measurements, impedance, electric and magnetic fields, radiofrequencies and measurements on materials.

Length

This metrology area covers the fields of laser frequencies and dimensional metrology.

Mass and related quantities

This metrology area covers the fields of mass standards, force, pressure, density, hardness, torque, gravity and viscosity.

Chemistry and Biology
Chemistry and Biology

This metrology area covers all fields relevant to chemistry and biology.

Ionizing radiation
Ionizing Radiation

This metrology area covers the fields of dosimetry, radioactivity and neutron measurements.