Results for: Chile "INN"
Filters
CMC Quick search Filters close;
CMC Area
To filter, either use the tick boxes or click on the underscored item
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Chile, INN (Instituto Nacional de Normalización)
Mass ,
Conventional mass
:
10 mg
Standard weights
Absolute expanded uncertainty
:
2.0E-3 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055N-1
Mass ,
Conventional mass
:
20 mg
Standard weights
Absolute expanded uncertainty
:
3.0E-3 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055O-1
Mass ,
Conventional mass
:
1 mg
Standard weights
Absolute expanded uncertainty
:
2.0E-3 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055K-1
Mass ,
Conventional mass
:
20 kg
Standard weights
Absolute expanded uncertainty
:
1.0E1 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055I-1
Mass ,
Conventional mass
:
50 kg
Standard weights
Absolute expanded uncertainty
:
7.5E1 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055J-1
Mass ,
Conventional mass
:
2 kg
Standard weights
Absolute expanded uncertainty
:
1 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055F-1
Force ,
Force
:
0.1 MN to 1 MN
STFR 3 MN
Relative expanded uncertainty
:
5.0E-2 %
Force reference transducer system
Temperature : (22 ± 1) ºC
Humidity : (50 ± 10) %
Service provided by IDIC (Instituto de Investigaciones y Control).
Approved on 04 May 2006
CMC ID :
SIM-M-CL-00000562-1
Force ,
Force
:
1 kN to 10 kN
STFR 50 kN
Relative expanded uncertainty
:
5.0E-2 %
Force reference transducer system
Temperature : (22 ± 1) ºC
Humidity : (50 ± 10) %
Service provided by IDIC (Instituto de Investigaciones y Control).
Approved on 04 May 2006
CMC ID :
SIM-M-CL-0000055W-1
Mass ,
Conventional mass
:
20 g
Standard weights
Absolute expanded uncertainty
:
2.5E-2 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-00000559-1
Mass ,
Conventional mass
:
1 kg
Standard weights
Absolute expanded uncertainty
:
0.5 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055E-1
Force ,
Force
:
0.3 MN to 3 MN
STFR 3 MN
Relative expanded uncertainty
:
0.1 %
Force reference transducer system
Temperature : (22 ± 1) ºC
Humidity : (50 ± 10) %
Service provided by IDIC (Instituto de Investigaciones y Control).
Approved on 04 May 2006
CMC ID :
SIM-M-CL-00000563-1
Mass ,
Conventional mass
:
100 mg
Standard weights
Absolute expanded uncertainty
:
5.0E-3 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055Q-1
Mass ,
Conventional mass
:
50 mg
Standard weights
Absolute expanded uncertainty
:
4.0E-3 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055P-1
Mass ,
Conventional mass
:
10 g
Standard weights
Absolute expanded uncertainty
:
2.0E-2 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-00000558-1
Force ,
Force
:
0.5 kN to 5 kN
STFR 50 kN
Relative expanded uncertainty
:
5.0E-2 %
Force reference transducer system
Temperature : (22 ± 1) ºC
Humidity : (50 ± 10) %
Service provided by IDIC (Instituto de Investigaciones y Control).
Approved on 04 May 2006
CMC ID :
SIM-M-CL-0000055V-1
Force ,
Force
:
2 kN to 20 kN
STFR 50 kN
Relative expanded uncertainty
:
5.0E-2 %
Force reference transducer system
Temperature : (22 ± 1) ºC
Humidity : (50 ± 10) %
Service provided by IDIC (Instituto de Investigaciones y Control).
Approved on 04 May 2006
CMC ID :
SIM-M-CL-0000055X-1
Force ,
Force
:
10 kN to 100 kN
STFR 500 kN
Relative expanded uncertainty
:
5.0E-2 %
Force reference transducer system
Temperature : (22 ± 1) ºC
Humidity : (50 ± 10) %
Service provided by IDIC (Instituto de Investigaciones y Control).
Approved on 04 May 2006
CMC ID :
SIM-M-CL-0000055Z-1
Mass ,
Conventional mass
:
200 mg
Standard weights
Absolute expanded uncertainty
:
6.0E-3 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055R-1
Mass ,
Conventional mass
:
500 g
Standard weights
Absolute expanded uncertainty
:
8.0E-3 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055S-1
Mass ,
Conventional mass
:
2 g
Standard weights
Absolute expanded uncertainty
:
1.2E-2 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055U-1