Results for: Chile
Filters
CMC Quick search Filters close;
CMC Area
To filter, either use the tick boxes or click on the underscored item
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Chile, INN (Instituto Nacional de Normalización)
Mass ,
Conventional mass
:
2 mg
Standard weights
Absolute expanded uncertainty
:
2.0E-3 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055L-1
Mass ,
Conventional mass
:
5 g
Standard weights
Absolute expanded uncertainty
:
1.5E-2 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-00000557-1
Mass ,
Conventional mass
:
200 g
Standard weights
Absolute expanded uncertainty
:
0.1 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055C-1
Mass ,
Conventional mass
:
5 kg
Standard weights
Absolute expanded uncertainty
:
2.5 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055G-1
Mass ,
Conventional mass
:
50 g
Standard weights
Absolute expanded uncertainty
:
3.0E-2 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055A-1
Mass ,
Conventional mass
:
1 g
Standard weights
Absolute expanded uncertainty
:
1.0E-2 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055T-1
Mass ,
Conventional mass
:
10 kg
Standard weights
Absolute expanded uncertainty
:
5 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID :
SIM-M-CL-0000055H-1
Force ,
Force
:
20 kN to 200 kN
STFR 500 kN
Relative expanded uncertainty
:
5.0E-2 %
Force reference transducer system
Temperature : (22 ± 1) ºC
Humidity : (50 ± 10) %
Service provided by IDIC (Instituto de Investigaciones y Control).
Approved on 04 May 2006
CMC ID :
SIM-M-CL-00000560-1
Force ,
Force
:
50 kN to 500 kN
STFR 500 kN
Relative expanded uncertainty
:
5.0E-2 %
Force reference transducer system
Temperature : (22 ± 1) ºC
Humidity : (50 ± 10) %
Service provided by IDIC (Instituto de Investigaciones y Control).
Approved on 04 May 2006
CMC ID :
SIM-M-CL-00000561-1
Chile, ISP (Instituto de Salud Pública)
Matrix or material |
Analyte or component |
Dissemination range of measurement capability |
Range of certified values in reference materials |
Mass fraction in mg/kg |
Relative expanded uncertainty in % |
Mass fraction in mg/kg |
Relative expanded uncertainty in % |
water |
arsenic |
0.02 to 50 |
1.5E1 to 1.0E1 |
|
|
-
Mechanism for measurement service delivery :
Assination valeu of PT/ Assign value to samples (RM-SP3 WATER)
-
Service of aptitude tests for the laboratory for the control of drink water quality
-
Approved on
05 September 2024
-
Institute service identifier : PEEC SP3
-
CMC ID :
SIM-QM-CL-00000PLA-1
Matrix or material |
Analyte or component |
Dissemination range of measurement capability |
Range of certified values in reference materials |
Mass fraction in mg/kg |
Relative expanded uncertainty in % |
Mass fraction in mg/kg |
Relative expanded uncertainty in % |
sea food |
arsenic |
0.2 to 50 |
9.1 to 1.5E1 |
|
|
-
Mechanism for measurement service delivery :
Reference material for proficiency testing value assignment
-
Service of aptitude tests for the food laboratory for the control of nutrients and surveillance of the nutritional labeling of foods in Chile
-
Approved on
12 August 2024
-
Institute service identifier : PEEC SP1
-
CMC ID :
SIM-QM-CL-00000PLB-1
Matrix or material |
Analyte or component |
Dissemination range of measurement capability |
Range of certified values in reference materials |
Mass fraction in mg/kg |
Relative expanded uncertainty in % |
Mass fraction in mg/kg |
Relative expanded uncertainty in % |
sea food |
cadmium |
0.04 to 10 |
9.3 to 1.5E1 |
|
|
-
Mechanism for measurement service delivery :
Reference material for proficiency testing value assignment
-
Service of aptitude tests for the food laboratory for the control of nutrients and surveillance of the nutritional labeling of foods in Chile
-
Approved on
12 August 2024
-
Institute service identifier : PEEC SP1
-
CMC ID :
SIM-QM-CL-00000PLC-1
Matrix or material |
Analyte or component |
Dissemination range of measurement capability |
Range of certified values in reference materials |
Mass fraction in mg/kg |
Relative expanded uncertainty in % |
Mass fraction in mg/kg |
Relative expanded uncertainty in % |
milk |
copper |
2.5 to 100 |
6 to 1.5E1 |
|
|
-
Mechanism for measurement service delivery :
Reference material for proficiency testing value assignment
-
Service of aptitude tests for the food laboratory for the control of nutrients and surveillance of the nutritional labeling of foods in Chile
-
Approved on
12 August 2024
-
Institute service identifier : PEEC SP4
-
CMC ID :
SIM-QM-CL-00000PL3-1
Matrix or material |
Analyte or component |
Dissemination range of measurement capability |
Range of certified values in reference materials |
Mass fraction in mg/g |
Relative expanded uncertainty in % |
Mass fraction in mg/g |
Relative expanded uncertainty in % |
water |
ethanol |
0.5 to 1.5 |
4.4 to 4.1 |
0.5 to 1.5 |
4.4 to 4.1 |
-
Mechanism for measurement service delivery :
CRM ETOH G1
-
Approved on
20 July 2020
-
Institute service identifier : CRM
-
CMC ID :
SIM-QM-CL-00000LQG-1
Matrix or material |
Analyte or component |
Dissemination range of measurement capability |
Range of certified values in reference materials |
Mass fraction in mg/kg |
Relative expanded uncertainty in % |
Mass fraction in mg/kg |
Relative expanded uncertainty in % |
milk |
iron |
2.5 to 250 |
1.5E1 to 5 |
|
|
-
Mechanism for measurement service delivery :
Reference material for proficiency testing value assignment
-
Service of aptitude tests for the food laboratory for the control of nutrients and surveillance of the nutritional labeling of foods in Chile
-
Approved on
12 August 2024
-
Institute service identifier : PEEC SP4
-
CMC ID :
SIM-QM-CL-00000PL1-1
Matrix or material |
Analyte or component |
Dissemination range of measurement capability |
Range of certified values in reference materials |
Mass fraction in mg/kg |
Relative expanded uncertainty in % |
Mass fraction in mg/kg |
Relative expanded uncertainty in % |
drinking water |
zinc |
2.5 to 150 |
1.8 to 1.0E1 |
|
|
-
Mechanism for measurement service delivery :
Assination valeu of PT/ Assign value to samples (CRM-SP3-WATER)
-
Service of aptitude tests for the food laboratory for the control of drink water quality in Chile.
-
Approved on
05 September 2024
-
Institute service identifier : PEEC SP3
-
CMC ID :
SIM-QM-CL-00000PL4-1
Matrix or material |
Analyte or component |
Dissemination range of measurement capability |
Range of certified values in reference materials |
Mass fraction in mg/kg |
Relative expanded uncertainty in % |
Mass fraction in mg/kg |
Relative expanded uncertainty in % |
milk |
zinc |
2.5 to 150 |
1.8 to 1.0E1 |
|
|
-
Mechanism for measurement service delivery :
Reference material for proficiency testing value assignment
-
Service of aptitude tests for the food laboratory for the control of nutrients and surveillance of the nutritional labeling of foods in Chile
-
Approved on
12 August 2024
-
Institute service identifier : PEEC SP4
-
CMC ID :
SIM-QM-CL-00000PL2-1
Chile, UDEC (Universidad de Concepción (UDEC))
DC current (up to 100 A) ,
DC current meters: intermediate values
:
1.00E-4 A to 20 A
Multimeter
Relative expanded uncertainty
:
5 µA/A to 3.4E2 µA/A
Uncertainty table
{"row_1":{"col_1":"","col_2":"relative uncertainty (<i>k</i> = 2)"},"row_2":{"col_1":">0.1 mA to <0.2 mA","col_2":"78 to 130"},"row_3":{"col_1":"0.2 mA","col_2":"8.0"},"row_4":{"col_1":">0.2 mA to 0.22 mA","col_2":"73 to 77"},"row_5":{"col_1":">0.22 mA to <1 mA","col_2":"33 to 57"},"row_6":{"col_1":"1 mA","col_2":"5.0"},"row_7":{"col_1":">1 mA to <2 mA","col_2":"29 to 32"},"row_8":{"col_1":"2 mA","col_2":"7.0"},"row_9":{"col_1":">2 mA to 2.2 mA","col_2":"29"},"row_10":{"col_1":">2.2 mA to <10 mA","col_2":"30 to 44"},"row_11":{"col_1":"10 mA","col_2":"5.0"},"row_12":{"col_1":">10 mA to <20 mA","col_2":"28 to 29"},"row_13":{"col_1":"20 mA","col_2":"9.0"},"row_14":{"col_1":">20 mA to 22 mA","col_2":"27"},"row_15":{"col_1":">22 mA to <30 mA","col_2":"59 to 67"},"row_16":{"col_1":"30 mA","col_2":"10"},"row_17":{"col_1":">30 mA to <50 mA","col_2":"50 to 59"},"row_18":{"col_1":"50 mA","col_2":"10"},"row_19":{"col_1":">50 mA to <100 mA","col_2":"43 to 49"},"row_20":{"col_1":"100 mA","col_2":"9.0"},"row_21":{"col_1":">100 mA to <200 mA","col_2":"39 to 42"},"row_22":{"col_1":"200 mA","col_2":"11"},"row_23":{"col_1":">200 mA to 220 mA","col_2":"39"},"row_24":{"col_1":">220 mA to <300 mA","col_2":"91 to 110"},"row_25":{"col_1":"300 mA","col_2":"24"},"row_26":{"col_1":">300 mA to <500 mA","col_2":"75 to 90"},"row_27":{"col_1":"500 mA","col_2":"24"},"row_28":{"col_1":">500 mA to <1 A","col_2":"63 to 74"},"row_29":{"col_1":"1 A","col_2":"21"},"row_30":{"col_1":">1 A to <2 A","col_2":"57 to 62"},"row_31":{"col_1":"2 A","col_2":"32"},"row_32":{"col_1":">2 A to 2.2 A","col_2":"56"},"row_33":{"col_1":">2.2 A to <3 A","col_2":"320 to 340"},"row_34":{"col_1":"3 A","col_2":"32"},"row_35":{"col_1":">3 A to <5 A","col_2":"290 to 320"},"row_36":{"col_1":"5 A","col_2":"32"},"row_37":{"col_1":">5 A to <10 A","col_2":"270 to 290"},"row_38":{"col_1":"10 A","col_2":"26"},"row_39":{"col_1":">10 A to <20 A","col_2":"260 to 270"},"row_40":{"col_1":"20 A","col_2":"40"}}
Direct comparison
Room temperature : 23 °C
Measurement lower limit for values >1E-04 A.
Approved on 21 April 2016
Institute service identifier : UDEC/3.2.2
CMC ID :
SIM-EM-CL-000007PO-1
DC resistance ,
DC resistance standards and sources: intermediate values
:
0.01 kΩ to 1000 kΩ
Fixed resistor, resistance box
Relative expanded uncertainty
:
3.5 µΩ/Ω to 1.0E1 µΩ/Ω
Uncertainty table
{"row_1":{"col_1":"","col_2":"relative uncertainty (<i>k</i> = 2)"},"row_2":{"col_1":"10 Ω","col_2":"3.5"},"row_3":{"col_1":"100 Ω","col_2":"3.5"},"row_4":{"col_1":"1 kΩ","col_2":"3.5"},"row_5":{"col_1":"10 kΩ","col_2":"3.5"},"row_6":{"col_1":"100 kΩ","col_2":"5.0"},"row_7":{"col_1":"1 MΩ","col_2":"10"},"row_8":{"col_1":"10 MΩ","col_2":"12"},"row_9":{"col_1":"100 MΩ","col_2":"25"}}
Indirect comparison using a high accuracy voltmeter as transfer standard
Room temperature : 23 °C
Decade values of resistance
Approved on 21 April 2016
Institute service identifier : UDEC/2.1.2
CMC ID :
SIM-EM-CL-000007PG-1
DC voltage (up to 1100 V) ,
DC voltage meters: intermediate values
:
0.022 V to 1000 V
DC voltmeter, multimeter
Relative expanded uncertainty
:
2.1 µV/V to 3.9E1 µV/V
Uncertainty table
{"row_1":{"col_1":"","col_2":"Expanded Uncertainty µV/V","col_3":"Uncertaity formula (<i>U</i> in V)"},"row_2":{"col_1":"0.1 V","col_2":"6.5","col_3":"-"},"row_3":{"col_1":"0.2 V","col_2":"6.5","col_3":"-"},"row_4":{"col_1":"0.6 V","col_2":"4.0","col_3":"-"},"row_5":{"col_1":"1 V","col_2":"2.1","col_3":"-"},"row_6":{"col_1":"2 V","col_2":"3.5","col_3":"-"},"row_7":{"col_1":"6 V","col_2":"3.5","col_3":"-"},"row_8":{"col_1":"10 V","col_2":"2.0","col_3":"-"},"row_9":{"col_1":"20 V","col_2":"3.5","col_3":"-"},"row_10":{"col_1":"60 V","col_2":"3.5","col_3":"-"},"row_11":{"col_1":"100 V","col_2":"2.0","col_3":"-"},"row_12":{"col_1":"200 V","col_2":"4.5","col_3":"-"},"row_13":{"col_1":"600 V","col_2":"6.5","col_3":"-"},"row_14":{"col_1":"1000 V","col_2":"3.0","col_3":"-"},"row_15":{"col_1":"0.022 V to 0.22 V","col_2":"39 to 11","col_3":"(0.7/<i>U</i> + 7)"},"row_16":{"col_1":">0.22 V to 2.2 V","col_2":"9.1 to 5.5","col_3":"(0.9/<i>U</i> + 5)"},"row_17":{"col_1":">2.2 V to 22 V","col_2":"5.8 to 3.8","col_3":"(5/<i>U</i> + 3.5)"},"row_18":{"col_1":">22 V to 220 V","col_2":"7.1 to 5.3","col_3":"(45/<i>U</i> + 5)"},"row_19":{"col_1":">220 V to 1000 V","col_2":"10","col_3":"(400/<i>U</i> + 8)"}}
Direct comparison
Room temperature : 23 °C
Approved on 21 April 2016
Institute service identifier : UDEC/1.2.2
CMC ID :
SIM-EM-CL-000007PE-1