Results for: Azerbaijan
Filters
CMC Quick search Filters close;
CMC Area
To filter, either use the tick boxes or click on the underscored item
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Azerbaijan, AzMI (Azerbaijan Institute of Metrology)
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
-40 °C to 10 °C
Digital thermometer IPRT
Absolute expanded uncertainty
:
4.0E-2 K
Comparison method
Cryostat : ethanol
Approved on 31 March 2020
Institute service identifier : AzMI/17
CMC ID :
COOMET-T-AZ-00000KG6-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
10 °C to 80 °C
Digital thermometer IPRT
Absolute expanded uncertainty
:
1.5E-2 K
Comparison method
Thermostat : water
Approved on 31 March 2020
Institute service identifier : AzMI/18
CMC ID :
COOMET-T-AZ-00000KG7-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
80 °C to 100 °C
Thermistors
Absolute expanded uncertainty
:
5.0E-2 K
Comparison method
Thermostat : silicon oil
Approved on 31 March 2020
Institute service identifier : AzMI/8
CMC ID :
COOMET-T-AZ-00000KFX-1
Mass ,
Mass
:
1 g to 10 g
Mass standards
Absolute expanded uncertainty
:
1.0E1 µg to 2.0E1 µg
Comparison in air
Temperature : 18.0 °C to 23.0 °C
Relative humidity : 40 % to 60 %
Uncertainty scales with measurand level. The volume of the mass standards is known
Approved on 25 October 2020
Institute service identifier : 03-AzMI-04L
CMC ID :
COOMET-M-AZ-00000M7A-1
Mass ,
Mass
:
1 kg to 5 kg
Mass standards
Absolute expanded uncertainty
:
0.5 mg to 2.5 mg
Comparison in air
Temperature : 18.0 °C to 23.0 °C
Relative humidity : 40 % to 60 %
Uncertainty scales with measurand level. The volume of the mass standards is known
Approved on 25 October 2020
Institute service identifier : 03-AzMI-04L
CMC ID :
COOMET-M-AZ-00000M7D-1
Viscosity ,
Kinematic viscosity
:
20 mm2/s to 1000 mm2/s
Newtonian liquids
Relative expanded uncertainty
:
0.22 % to 0.3 %
Reference liquid
Temperature : 20 °C
Calibration is done with traceability of own CRM (measured by standard viscosimetr) and CRM from ZMK.
Approved on 14 June 2018
CMC ID :
COOMET-M-AZ-00000AZQ-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
300 °C to 420 °C
Industrial Platinum
Resistance
thermometers
Absolute expanded uncertainty
:
0.3 K to 5.0E-2 K
Direct comparison
Furnace or blocks : temperature controlled furnaces
Approved on 31 March 2020
Institute service identifier : AzMI/5
CMC ID :
COOMET-T-AZ-00000KFU-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
10 °C to 60 °C
Liquid-in-glass thermometers, total immersion 0.1 °C
Absolute expanded uncertainty
:
3.0E-2 K
Comparison method
Stirred liquid bath : ethanol
Approved on 31 March 2020
Institute service identifier : AzMI/14
CMC ID :
COOMET-T-AZ-00000KG3-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
0 °C to 80 °C
Thermistors
Absolute expanded uncertainty
:
4.0E-2 K
Comparison method
Thermostat : water
Approved on 31 March 2020
Institute service identifier : AzMI/7
CMC ID :
COOMET-T-AZ-00000KFW-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
300 °C to 420 °C
Digital thermometer IPRT
Absolute expanded uncertainty
:
0.3 K to 0.6 K
Comparison method
Approved on 31 March 2020
Institute service identifier : AzMI/20
CMC ID :
COOMET-T-AZ-00000KG9-1
Items for disseminating ITS-90 and PLTS 2000 ,
Temperature
:
300 °C to 420 °C
Digital thermometer Thermocouples
Absolute expanded uncertainty
:
0.8 °C
Comparison method
Approved on 31 March 2020
Institute service identifier : AzMI/24
CMC ID :
COOMET-T-AZ-00000KGD-1
Mass ,
Mass
:
0.1 g to 1 g
Mass standards
Absolute expanded uncertainty
:
5 µg to 1.0E1 µg
Comparison in air
Temperature : 18.0 °C to 23.0 °C
Relative humidity : 40 % to 60 %
Uncertainty scales with measurand level.
Approved on 25 October 2020
Institute service identifier : 03-AzMI-04L
CMC ID :
COOMET-M-AZ-00000M79-1
Mass ,
Mass
:
1 mg to 100 mg
Mass standards
Absolute expanded uncertainty
:
3 µg to 5 µg
Comparison in air
Temperature : 18.0 °C to 23.0 °C
Relative humidity : 40 % to 60 %
Uncertainty scales with measurand level.
Approved on 25 October 2020
Institute service identifier : 03-AzMI-04L
CMC ID :
COOMET-M-AZ-00000M78-1
Density ,
Density of liquid
:
757 kg/m3 to 1615 kg/m3
Density standard liquid
Absolute expanded uncertainty
:
3.1E-2 kg/m3 to 3.6E-2 kg/m3
Reference liquids
Temperature : 20 °C
Pressure : 101.3 kPa
Approved on 20 August 2022
Institute service identifier : AzMI
CMC ID :
COOMET-M-AZ-00000NXD-1
Ionizing radiation
Kerma/rate
,
Protection level ionization chambers
: 12 mGy/h to 120 mGy/h
Relative expanded uncertainty 1.6 % to 1.9 %
Calibration against a secondary standard free in air
X-ray, 50 kV to 300 kV, ISO 4037:2019, Narrow Series, 60 kV to 200 kV
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 24 June 2024
Institute service identifier : COO-RAD-AzMI-1001
CMC ID :