In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Bulgaria, BIM (Bulgarian Institute of Metrology)
Ionizing radiation
Kerma/rate
,
Gamma irradiator
: 0.1 Gy/h to 1 Gy/h
Relative expanded uncertainty 1.5 %
Calibration against a secondary standard free in air
Co-60, no further specifications
Reference standard : Secondary standard with ionization chambers type ND1001
Source of traceability : BIPM
Approved on 13 February 2007
Institute service identifier : EUR-RAD-BIM-1004
CMC ID :
EURAMET-RI-BG-00000H6D-1
Ionizing radiation
Surface emission rate
,
Single nuclide extended area source
: 20 s-1 to 4400 s-1
Relative expanded uncertainty 1.8 %
Large area windowless 2p proportional counter
Pm-147, no further specifications
Reference standard : 2π windowless proportional counter
Source of traceability : BIM
Approved on 16 March 2023
Institute service identifier : EUR-RAD-BIM-2006
CMC ID :
EURAMET-RI-BG-00000O1W-1
Ionizing radiation
Surface emission rate
,
Single nuclide extended area source
: 20 s-1 to 4400 s-1
Relative expanded uncertainty 1.8 %
Large area windowless 2p proportional counter
C-14, no further specifications
Reference standard : 2π windowless proportional counter
Source of traceability : BIM
Approved on 16 March 2023
Institute service identifier : EUR-RAD-BIM-2005
CMC ID :
EURAMET-RI-BG-00000O1T-1
Ionizing radiation
Surface emission rate
,
Single nuclide extended area source
: 20 s-1 to 4400 s-1
Relative expanded uncertainty 1.8 %
Large area windowless 2p proportional counter
Sr-90, no further specifications
Reference standard : 2π windowless proportional counter
Source of traceability : BIM
Approved on 16 March 2023
Institute service identifier : EUR-RAD-BIM-2007
CMC ID :
EURAMET-RI-BG-00000O1X-1
Ionizing radiation
Kerma/rate
,
Dosemeter
: 6.00E-6 Gy/h to 0.1 Gy/h
Relative expanded uncertainty 2.3 %
Calibration against a secondary standard free in air
Cs-137, no further specifications
Reference standard : Secondary standard ionization chamber
Source of traceability : PTB
Approved on 13 February 2007
Institute service identifier : EUR-RAD-BIM-1007
CMC ID :
EURAMET-RI-BG-00000H6G-2
Ionizing radiation
Kerma/rate
,
Dosemeter
: 0.1 Gy/h to 1 Gy/h
Relative expanded uncertainty 1.5 %
Calibration against a secondary standard free in air
Cs-137, no further specifications
Reference standard : Secondary standard ionization chamber
Source of traceability : PTB
Approved on 16 March 2023
Institute service identifier : EUR-RAD-BIM-1006
CMC ID :
EURAMET-RI-BG-00000H6F-2
Ionizing radiation
Surface emission rate
,
Single nuclide extended area source
: 4 s-1 to 4400 s-1
Relative expanded uncertainty 1.8 %
Large area windowless 2p proportional counter
Am-241, no further specifications
Reference standard : 2π windowless proportional counter
Source of traceability : BIM
Approved on 16 March 2023
Institute service identifier : EUR-RAD-BIM-2008
CMC ID :
EURAMET-RI-BG-00000O1Y-1
Ionizing radiation
Kerma/rate
,
Dosemeter ionization chamber
: 0.1 Gy to 1 Gy
Relative expanded uncertainty 1.5 %
Calibration against a secondary standard free in air
Co-60, Beam size 15 cm by 15 cm, field at 1 m distance. Air kerma rate 0.2 mGy/s
Reference standard : Secondary standard with ionization chambers type ND1001
Source of traceability : BIPM
Approved on 13 February 2007
Institute service identifier : EUR-RAD-BIM-1002
CMC ID :
EURAMET-RI-BG-00000H6B-1
Ionizing radiation
Kerma/rate
,
Gamma irradiator
: 6.00E-6 Gy/h to 0.1 Gy/h
Relative expanded uncertainty 2.3 %
Calibration against a secondary standard free in air
Co-60, no further specifications
Reference standard : Secondary standard with ionization chambers type ND 1000
Source of traceability : PTB
Approved on 13 February 2007
Institute service identifier : EUR-RAD-BIM-1005
CMC ID :
EURAMET-RI-BG-00000H6E-1
Ionizing radiation
Kerma/rate
,
Dosemeter
: 6.00E-6 Gy to 0.1 Gy
Relative expanded uncertainty 2.3 %
Calibration against a secondary standard free in air
Co-60, no further specifications
Reference standard : Secondary standard ionization chamber
Source of traceability : PTB
Approved on 13 February 2007
Institute service identifier : EUR-RAD-BIM-1003
CMC ID :
EURAMET-RI-BG-00000H6C-2