In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Chile, INN (Instituto Nacional de Normalización)
Mass , Conventional mass : 2 mg
Standard weights
Absolute expanded uncertainty : 2.0E-3 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID : SIM-M-CL-0000055L-1
 
Mass , Conventional mass : 5 g
Standard weights
Absolute expanded uncertainty : 1.5E-2 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID : SIM-M-CL-00000557-1
 
Mass , Conventional mass : 200 g
Standard weights
Absolute expanded uncertainty : 0.1 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID : SIM-M-CL-0000055C-1
 
Mass , Conventional mass : 5 kg
Standard weights
Absolute expanded uncertainty : 2.5 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID : SIM-M-CL-0000055G-1
 
Mass , Conventional mass : 50 g
Standard weights
Absolute expanded uncertainty : 3.0E-2 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID : SIM-M-CL-0000055A-1
 
Mass , Conventional mass : 1 g
Standard weights
Absolute expanded uncertainty : 1.0E-2 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID : SIM-M-CL-0000055T-1
 
Mass , Conventional mass : 10 kg
Standard weights
Absolute expanded uncertainty : 5 mg
Substitution weighing with buoyancy correction
Laboratory temperature : 18 °C to 27 °C, the allowed temperature change in 1 hour is equal to ± 0.5 ºC, in 4 hours: ± 0.7 ºC and in 24 hours: ± 1 ºC
Humidity : 40 % to 60 %, the allowed humidity change in 4 hours is equal to ± 10 %
Service provided by CESMEC (Centro de Estudios, Medición y Certificación de Calidad).
Approved on 04 May 2006
Institute service identifier : INN/131-750
CMC ID : SIM-M-CL-0000055H-1
 
Force , Force : 20 kN to 200 kN
STFR 500 kN
Relative expanded uncertainty : 5.0E-2 %
Force reference transducer system
Temperature : (22 ± 1) ºC
Humidity : (50 ± 10) %
Service provided by IDIC (Instituto de Investigaciones y Control).
Approved on 04 May 2006
CMC ID : SIM-M-CL-00000560-1
 
Force , Force : 50 kN to 500 kN
STFR 500 kN
Relative expanded uncertainty : 5.0E-2 %
Force reference transducer system
Temperature : (22 ± 1) ºC
Humidity : (50 ± 10) %
Service provided by IDIC (Instituto de Investigaciones y Control).
Approved on 04 May 2006
CMC ID : SIM-M-CL-00000561-1
 
Chile, ISP (Instituto de Salud Pública)
Matrix or material Analyte or component Dissemination range of measurement capability Range of certified values in reference materials
Mass fraction in mg/g Relative expanded uncertainty in % Mass fraction in mg/g Relative expanded uncertainty in %
water ethanol 0.5 to 1.5 4.4 to 4.1 0.5 to 1.5 4.4 to 4.1
  • Mechanism for measurement service delivery : CRM ETOH G1
  • Approved on 20 July 2020
  • Institute service identifier : CRM
  • CMC ID : SIM-QM-CL-00000LQG-1
     
Chile, UDEC (Universidad de Concepción (UDEC))
DC current (up to 100 A) , DC current meters: intermediate values : 1.00E-4 A to 20 A
Multimeter
Relative expanded uncertainty : 5 µA/A to 3.4E2 µA/A
{"row_1":{"col_1":"","col_2":"relative uncertainty (<i>k</i> = 2)"},"row_2":{"col_1":"&gt;0.1 mA to &lt;0.2 mA","col_2":"78 to 130"},"row_3":{"col_1":"0.2 mA","col_2":"8.0"},"row_4":{"col_1":"&gt;0.2 mA to 0.22 mA","col_2":"73 to 77"},"row_5":{"col_1":"&gt;0.22 mA to &lt;1 mA","col_2":"33 to 57"},"row_6":{"col_1":"1 mA","col_2":"5.0"},"row_7":{"col_1":"&gt;1 mA to &lt;2 mA","col_2":"29 to 32"},"row_8":{"col_1":"2 mA","col_2":"7.0"},"row_9":{"col_1":"&gt;2 mA to 2.2 mA","col_2":"29"},"row_10":{"col_1":"&gt;2.2 mA to &lt;10 mA","col_2":"30 to 44"},"row_11":{"col_1":"10 mA","col_2":"5.0"},"row_12":{"col_1":"&gt;10 mA to &lt;20 mA","col_2":"28 to 29"},"row_13":{"col_1":"20 mA","col_2":"9.0"},"row_14":{"col_1":"&gt;20 mA to 22 mA","col_2":"27"},"row_15":{"col_1":"&gt;22 mA to &lt;30 mA","col_2":"59 to 67"},"row_16":{"col_1":"30 mA","col_2":"10"},"row_17":{"col_1":"&gt;30 mA to &lt;50 mA","col_2":"50 to 59"},"row_18":{"col_1":"50 mA","col_2":"10"},"row_19":{"col_1":"&gt;50 mA to &lt;100 mA","col_2":"43 to 49"},"row_20":{"col_1":"100 mA","col_2":"9.0"},"row_21":{"col_1":"&gt;100 mA to &lt;200 mA","col_2":"39 to 42"},"row_22":{"col_1":"200 mA","col_2":"11"},"row_23":{"col_1":"&gt;200 mA to 220 mA","col_2":"39"},"row_24":{"col_1":"&gt;220 mA to &lt;300 mA","col_2":"91 to 110"},"row_25":{"col_1":"300 mA","col_2":"24"},"row_26":{"col_1":"&gt;300 mA to &lt;500 mA","col_2":"75 to 90"},"row_27":{"col_1":"500 mA","col_2":"24"},"row_28":{"col_1":"&gt;500 mA to &lt;1 A","col_2":"63 to 74"},"row_29":{"col_1":"1 A","col_2":"21"},"row_30":{"col_1":"&gt;1 A to &lt;2 A","col_2":"57 to 62"},"row_31":{"col_1":"2 A","col_2":"32"},"row_32":{"col_1":"&gt;2 A to 2.2 A","col_2":"56"},"row_33":{"col_1":"&gt;2.2 A to &lt;3 A","col_2":"320 to 340"},"row_34":{"col_1":"3 A","col_2":"32"},"row_35":{"col_1":"&gt;3 A to &lt;5 A","col_2":"290 to 320"},"row_36":{"col_1":"5 A","col_2":"32"},"row_37":{"col_1":"&gt;5 A to &lt;10 A","col_2":"270 to 290"},"row_38":{"col_1":"10 A","col_2":"26"},"row_39":{"col_1":"&gt;10 A to &lt;20 A","col_2":"260 to 270"},"row_40":{"col_1":"20 A","col_2":"40"}}
The expanded uncertainties are expressed in µA/A
Direct comparison
Room temperature : 23 °C
Measurement lower limit for values >1E-04 A.
Approved on 21 April 2016
Institute service identifier : UDEC/3.2.2
CMC ID : SIM-EM-CL-000007PO-1
 
DC resistance , DC resistance standards and sources: intermediate values : 0.01 kΩ to 1000 kΩ
Fixed resistor, resistance box
Relative expanded uncertainty : 3.5 µΩ/Ω to 1.0E1 µΩ/Ω
{"row_1":{"col_1":"","col_2":"relative uncertainty (<i>k</i> = 2)"},"row_2":{"col_1":"10 Ω","col_2":"3.5"},"row_3":{"col_1":"100 Ω","col_2":"3.5"},"row_4":{"col_1":"1 kΩ","col_2":"3.5"},"row_5":{"col_1":"10 kΩ","col_2":"3.5"},"row_6":{"col_1":"100 kΩ","col_2":"5.0"},"row_7":{"col_1":"1 MΩ","col_2":"10"},"row_8":{"col_1":"10 MΩ","col_2":"12"},"row_9":{"col_1":"100 MΩ","col_2":"25"}}
The expanded uncertainties are expressed in µΩ/Ω
Indirect comparison using a high accuracy voltmeter as transfer standard
Room temperature : 23 °C
Decade values of resistance
Approved on 21 April 2016
Institute service identifier : UDEC/2.1.2
CMC ID : SIM-EM-CL-000007PG-1
 
DC voltage (up to 1100 V) , DC voltage meters: intermediate values : 0.022 V to 1000 V
DC voltmeter, multimeter
Relative expanded uncertainty : 2.1 µV/V to 3.9E1 µV/V
{"row_1":{"col_1":"","col_2":"Expanded Uncertainty µV/V","col_3":"Uncertaity formula (<i>U</i> in V)"},"row_2":{"col_1":"0.1 V","col_2":"6.5","col_3":"-"},"row_3":{"col_1":"0.2 V","col_2":"6.5","col_3":"-"},"row_4":{"col_1":"0.6 V","col_2":"4.0","col_3":"-"},"row_5":{"col_1":"1 V","col_2":"2.1","col_3":"-"},"row_6":{"col_1":"2 V","col_2":"3.5","col_3":"-"},"row_7":{"col_1":"6 V","col_2":"3.5","col_3":"-"},"row_8":{"col_1":"10 V","col_2":"2.0","col_3":"-"},"row_9":{"col_1":"20 V","col_2":"3.5","col_3":"-"},"row_10":{"col_1":"60 V","col_2":"3.5","col_3":"-"},"row_11":{"col_1":"100 V","col_2":"2.0","col_3":"-"},"row_12":{"col_1":"200 V","col_2":"4.5","col_3":"-"},"row_13":{"col_1":"600 V","col_2":"6.5","col_3":"-"},"row_14":{"col_1":"1000 V","col_2":"3.0","col_3":"-"},"row_15":{"col_1":"0.022 V to 0.22 V","col_2":"39 to 11","col_3":"(0.7/<i>U</i> + 7)"},"row_16":{"col_1":"&gt;0.22 V to 2.2 V","col_2":"9.1 to 5.5","col_3":"(0.9/<i>U</i> + 5)"},"row_17":{"col_1":"&gt;2.2 V to 22 V","col_2":"5.8 to 3.8","col_3":"(5/<i>U</i> + 3.5)"},"row_18":{"col_1":"&gt;22 V to 220 V","col_2":"7.1 to 5.3","col_3":"(45/<i>U</i> + 5)"},"row_19":{"col_1":"&gt;220 V to 1000 V","col_2":"10","col_3":"(400/<i>U</i> + 8)"}}
Direct comparison
Room temperature : 23 °C
Approved on 21 April 2016
Institute service identifier : UDEC/1.2.2
CMC ID : SIM-EM-CL-000007PE-1
 
AC voltage (up to the MHz range) , AC voltage (up to the MHz range): AC voltage up to 1100 V, meters : 0.1 V to 1100 V
Multimeter
Relative expanded uncertainty : 4.8E1 µV/V to 3.2E3 µV/V
{"row_1":{"col_1":"","col_2":"20 Hz to 40 Hz","col_3":"50 Hz to 1 kHz","col_4":"50 Hz to 20 kHz","col_5":"50 kHz to 100 kHz","col_6":"500 kHz to 1 MHz","col_7":"uncertaity formula (<i>U</i> in V) at 20 Hz to 40 Hz","col_8":"uncertaity formula (<i>U</i> in V) at 50 Hz to 1 kHz","col_9":"uncertaity formula (<i>U</i> in V) at 50 Hz to 20 kHz","col_10":"uncertaity formula (<i>U</i> in V) at 50 kHz to 100 kHz","col_11":"uncertaity formula (<i>U</i> in V) at 500 kHz to 1 MHz"},"row_2":{"col_1":"&gt;0.1 V to 0.22 V","col_2":"-","col_3":"-","col_4":"120 to 150","col_5":"540 to 630","col_6":"3000 to 3200","col_7":"-","col_8":"-","col_9":"1E06*(7µV+80E-06*<i>U</i>)/<i>U</i>","col_10":"1E06*(17µV+460E-06*<i>U</i>)/<i>U</i>","col_11":"1E06*(45µV+2.7E-03*<i>U</i>)/<i>U</i>"},"row_3":{"col_1":"&gt;0.22 V to 2.2 V","col_2":"-","col_3":"-","col_4":"49 to 82","col_5":"130 to 250","col_6":"1900 to 3100","col_7":"-","col_8":"-","col_9":"1E06*(8µV+45E-06*<i>U</i>)/<i>U</i>","col_10":"1E06*(30µV+110E-06*<i>U</i>)/<i>U</i>","col_11":"1E06*(0.3mV+1.7E-03*<i>U</i>)/<i>U</i>"},"row_4":{"col_1":"&gt;2.2 V to 22 V","col_2":"-","col_3":"-","col_4":"48 to 68","col_5":"110 to 200","col_6":"-","col_7":"-","col_8":"-","col_9":"1E06*(50µV+45E-06*<i>U</i>)/<i>U</i>","col_10":"1E06*(200µV+100E-06*<i>U</i>)/<i>U</i>","col_11":"-"},"row_5":{"col_1":"&gt;2.2 V to 10 V","col_2":"-","col_3":"-","col_4":"-","col_5":"-","col_6":"1900 to 3000","col_7":"-","col_8":"-","col_9":"-","col_10":"-","col_11":"1E06*(3.2mV+1.5E-03*<i>U</i>)/<i>U</i>"},"row_6":{"col_1":"&gt;22 V to 220 V","col_2":"100 to 160","col_3":"-","col_4":"55 to 80","col_5":"1600 to 1700","col_6":"-","col_7":"1E06*(1.5mV+90E-06*<i>U</i>)/<i>U</i>","col_8":"-","col_9":"1E06*(0.6mV+52E-06*<i>U</i>)/<i>U</i>","col_10":"1E06*(2.5mV+150E-06*<i>U</i>)/<i>U</i>","col_11":"-"},"row_7":{"col_1":"&gt;220 V to 1000 V","col_2":"-","col_3":"74 to 90","col_4":"-","col_5":"-","col_6":"-","col_7":"-","col_8":"1E06*(3.5mV+70E-06*<i>U</i>)/<i>U</i>","col_9":"-","col_10":"-","col_11":"-"}}
The expanded uncertainties are expressed in µV/V
Direct comparison
Frequency : 20 Hz to 1 MHz
Approved on 21 April 2016
Institute service identifier : UDEC/5.2.2
CMC ID : SIM-EM-CL-000007PS-1
 
DC voltage (up to 1100 V) , DC voltage sources: low values ranges : 0.2 V to 6 V
DC voltage source, multifunction calibrator
Relative expanded uncertainty : 2 µV/V to 4 µV/V
{"row_1":{"col_1":"","col_2":"Expanded Uncertainty µV/V"},"row_2":{"col_1":"0.1 V","col_2":"2.3"},"row_3":{"col_1":"0.2 V","col_2":"4.0"},"row_4":{"col_1":"0.6 V","col_2":"2.0"},"row_5":{"col_1":"1 V","col_2":"1.2"},"row_6":{"col_1":"2 V","col_2":"2.0"},"row_7":{"col_1":"6 V","col_2":"2.0"},"row_8":{"col_1":"10 V","col_2":"1.2"},"row_9":{"col_1":"20 V","col_2":"2.0"},"row_10":{"col_1":"60 V","col_2":"2.0"},"row_11":{"col_1":"100 V","col_2":"1.2"},"row_12":{"col_1":"200 V","col_2":"2.0"},"row_13":{"col_1":"600 V","col_2":"3.0"},"row_14":{"col_1":"1000 V","col_2":"1.4"}}
Indirect comparison using: a high accuracy voltmeter, two voltage standard, a reference divider
Fixed values : 0.2 V, 0.6 V, 2 V, 6 V
Approved on 21 April 2016
Institute service identifier : UDEC/1.1.2
CMC ID : SIM-EM-CL-000007PA-1
 
AC power , AC power and energy: three phase : 0.0125 Wh to 2880 Wh
Energy meter
Relative expanded uncertainty : 1.2E2 µWh (VAh)-1
Comparison
Frequency : 50 Hz to 60 Hz
Voltage : 60 V to 240 V
Current : >0.05 A to 120 A
Power Factor : 0.5<=cosfind<=1, 0.5<=cosfcap<=1
Measuring time : 10 s to 120 s
Approved on 21 April 2016
Institute service identifier : UDEC/7.1.3
CMC ID : SIM-EM-CL-000007PW-1
 
AC voltage (up to the MHz range) , AC voltage (up to the MHz range): AC voltage up to 1100 V, sources : 0.1 V to 1000 V
Multifunction calibrator
Relative expanded uncertainty : 2.4E1 µV/V to 1.2E3 µV/V
{"row_1":{"col_1":"","col_2":"50 Hz to 20 kHz","col_3":"50 kHz to 100 kHz","col_4":"500 kHz to 1 MHz","col_5":"uncertaity formula (<i>U</i> in V) at 50 Hz to 20 kHz","col_6":"uncertaity formula (<i>U</i> in V) at 50 kHz to 100 kHz","col_7":"uncertaity formula (<i>U</i> in V) at 500 kHz to 1 MHz"},"row_2":{"col_1":"&gt;0.1 V to 0.22 V","col_2":"45 to 53","col_3":"180 to 190","col_4":"1100","col_5":"1E06*(1.5µV+38E-06*<i>U</i>)/<i>U</i>","col_6":"1E06*(2.5µV+0.16E-03*<i>U</i>)/<i>U</i>","col_7":"1E06*(8.0µV+1.0E-03*<i>U</i>)/<i>U</i>"},"row_3":{"col_1":"&gt;0.22 V to 0.7 V","col_2":"36 to 40","col_3":"83 to 91","col_4":"1000","col_5":"1E06*(1.5µV+33E-06*<i>U</i>)/<i>U</i>","col_6":"1E06*(2.5µV+79E-06*<i>U</i>)/<i>U</i>","col_7":"1E06*(8.0µV+0.96E-03*<i>U</i>)/<i>U</i>"},"row_4":{"col_1":"&gt;0.7 V to 2.2 V","col_2":"24","col_3":"71","col_4":"900","col_5":"-","col_6":"-","col_7":"-"},"row_5":{"col_1":"&gt;2.2 V to 7 V","col_2":"24","col_3":"81","col_4":"1200","col_5":"-","col_6":"-","col_7":"-"},"row_6":{"col_1":"&gt;7 V to 22 V","col_2":"27","col_3":"81","col_4":"1200","col_5":"-","col_6":"-","col_7":"-"},"row_7":{"col_1":"&gt;22 V to 70 V","col_2":"32","col_3":"94","col_4":"1200","col_5":"-","col_6":"-","col_7":"-"},"row_8":{"col_1":"&gt;70 V to 220 V","col_2":"38","col_3":"98","col_4":"-","col_5":"-","col_6":"-","col_7":"-"},"row_9":{"col_1":"&gt;220 V to 700 V","col_2":"47","col_3":"-","col_4":"-","col_5":"-","col_6":"-","col_7":"-"},"row_10":{"col_1":"&gt;700 V to 1000 V","col_2":"44","col_3":"-","col_4":"-","col_5":"-","col_6":"-","col_7":"-"}}
The expanded uncertainties are expressed in µV/V
Direct comparison
Frequency : 50 Hz to 1 MHz
Measurement lower limit for values >0.01 V.
Approved on 21 April 2016
Institute service identifier : UDEC/5.2.1
CMC ID : SIM-EM-CL-000007PQ-1
 
AC voltage (up to the MHz range) , AC voltage (up to the MHz range): AC voltage up to 1100 V, meters : 0.1 V to 1000 V
Multimeter
Relative expanded uncertainty : 1.0E1 µV/V to 6.2E1 µV/V
{"row_1":{"col_1":"","col_2":"40 Hz","col_3":"55 Hz","col_4":"1 kHz","col_5":"10 kHz","col_6":"100 kHz","col_7":"1 MHz"},"row_2":{"col_1":"0.1 V","col_2":"28","col_3":"-","col_4":"27","col_5":"27","col_6":"61","col_7":"56"},"row_3":{"col_1":"0.2 V","col_2":"30","col_3":"-","col_4":"19","col_5":"19","col_6":"21","col_7":"54"},"row_4":{"col_1":"0.6 V","col_2":"13","col_3":"-","col_4":"12","col_5":"13","col_6":"17","col_7":"48"},"row_5":{"col_1":"1 V","col_2":"17","col_3":"-","col_4":"12","col_5":"12","col_6":"17","col_7":"62"},"row_6":{"col_1":"2 V","col_2":"18","col_3":"-","col_4":"15","col_5":"17","col_6":"19","col_7":"59"},"row_7":{"col_1":"6 V","col_2":"13","col_3":"-","col_4":"12","col_5":"12","col_6":"15","col_7":"41"},"row_8":{"col_1":"10 V","col_2":"12","col_3":"-","col_4":"10","col_5":"10","col_6":"13","col_7":"61"},"row_9":{"col_1":"20 V","col_2":"15","col_3":"-","col_4":"12","col_5":"12","col_6":"21","col_7":"-"},"row_10":{"col_1":"60 V","col_2":"17","col_3":"-","col_4":"15","col_5":"15","col_6":"30","col_7":"-"},"row_11":{"col_1":"100 V","col_2":"29","col_3":"-","col_4":"19","col_5":"19","col_6":"34","col_7":"-"},"row_12":{"col_1":"200 V","col_2":"28","col_3":"-","col_4":"22","col_5":"21","col_6":"42","col_7":"-"},"row_13":{"col_1":"600 V","col_2":"-","col_3":"41","col_4":"29","col_5":"-","col_6":"-","col_7":"-"},"row_14":{"col_1":"1000 V","col_2":"-","col_3":"40","col_4":"29","col_5":"-","col_6":"-","col_7":"-"}}
The expanded uncertainties are expressed in µV/V
AC/DC transference
Frequency : 40 Hz to 1 MHz
Approved on 21 April 2016
Institute service identifier : UDEC/5.2.2
CMC ID : SIM-EM-CL-000007PR-1
 
DC voltage (up to 1100 V) , DC voltage sources: intermediate values : 20 V to 600 V
DC voltage source, multifunction calibrator
Relative expanded uncertainty : 2 µV/V to 3 µV/V
{"row_1":{"col_1":"","col_2":"Expanded Uncertainty µV/V"},"row_2":{"col_1":"0.1 V","col_2":"2.3"},"row_3":{"col_1":"0.2 V","col_2":"4.0"},"row_4":{"col_1":"0.6 V","col_2":"2.0"},"row_5":{"col_1":"1 V","col_2":"1.2"},"row_6":{"col_1":"2 V","col_2":"2.0"},"row_7":{"col_1":"6 V","col_2":"2.0"},"row_8":{"col_1":"10 V","col_2":"1.2"},"row_9":{"col_1":"20 V","col_2":"2.0"},"row_10":{"col_1":"60 V","col_2":"2.0"},"row_11":{"col_1":"100 V","col_2":"1.2"},"row_12":{"col_1":"200 V","col_2":"2.0"},"row_13":{"col_1":"600 V","col_2":"3.0"},"row_14":{"col_1":"1000 V","col_2":"1.4"}}
Indirect comparison using: a high accuracy voltmeter, two voltage standard, a reference divider
Fixed values : 20 V, 60 V, 200 V, 600 V
Approved on 21 April 2016
Institute service identifier : UDEC/1.1.3
CMC ID : SIM-EM-CL-000007PB-1
 
DC resistance , DC resistance standards and sources: multiple ranges : 1.00E-6 MΩ to 1000 MΩ
Multifunction calibrator
Relative expanded uncertainty : 4.1 µΩ/Ω to 6.5E2 µΩ/Ω
{"row_1":{"col_1":"","col_2":"relative uncertainty (<i>k</i> = 2)","col_3":"uncertaity formula (<i>U</i> in V)"},"row_2":{"col_1":"1 Ω to 10 Ω","col_2":"8 to 19","col_3":"1E06*(12 µΩ+6.5E-06*<i>R</i>)/<i>R</i>"},"row_3":{"col_1":"&gt;10 Ω to 100 Ω","col_2":"4.8 to 12","col_3":"1E06*(72 µΩ+4.0E-06*<i>R</i>)/<i>R</i>"},"row_4":{"col_1":"&gt;100 Ω to 1000 Ω","col_2":"4.1 to 4.6","col_3":"1E06*(60 µΩ+4.0E-06*<i>R</i>)/<i>R</i>"},"row_5":{"col_1":"&gt;1 kΩ to 10 kΩ","col_2":"5.9 to 6.3","col_3":"1E06*(0.5 mΩ+4.0E-06*<i>R</i>)/<i>R</i>"},"row_6":{"col_1":"&gt;10 kΩ to 100 kΩ","col_2":"6.3","col_3":"1E06*(2.7 mΩ+6.0E-06*<i>R</i>)/<i>R</i>"},"row_7":{"col_1":"&gt;100 kΩ to 1000 kΩ","col_2":"13","col_3":"1E06*(0.1 Ω+12E-06*<i>R</i>)/<i>R</i>"},"row_8":{"col_1":"&gt;1 MΩ to 10 MΩ","col_2":"19 to 25","col_3":"1E06*(6.7 Ω+18E-06*<i>R</i>)/<i>R</i>"},"row_9":{"col_1":"&gt;10 MΩ to 100 MΩ","col_2":"73 to 130","col_3":"1E06*(0.55 kΩ+67E-06*<i>R</i>)/<i>R</i>"},"row_10":{"col_1":"&gt;100 MΩ to 1000 MΩ","col_2":"600 to 650","col_3":"1E06*(6.0 kΩ+0.59E-03*<i>R</i>)/<i>R</i>"}}
The expanded uncertainties are expressed in µΩ/Ω
Indirect comparison using a high accuracy mulltimeter as transfer standard
Room temperature : 23 °C
Approved on 21 April 2016
Institute service identifier : UDEC/2.1.5
CMC ID : SIM-EM-CL-000007PI-1
 

Select Metrology Area

General physics
Acoustics, Ultrasound, Vibration

This metrology area covers the fields of acoustics, ultrasound and vibration.

Electricity and Magnetism

This metrology area covers the fields of DC and AC measurements, impedance, electric and magnetic fields, radiofrequencies and measurements on materials.

Length

This metrology area covers the fields of laser frequencies and dimensional metrology.

Mass and related quantities

This metrology area covers the fields of mass standards, force, pressure, density, hardness, torque, gravity and viscosity.

Chemistry and Biology
Chemistry and Biology

This metrology area covers all fields relevant to chemistry and biology.

Ionizing radiation
Ionizing Radiation

This metrology area covers the fields of dosimetry, radioactivity and neutron measurements.