CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter A

WAVELENGTH      380 nm

 

CCPR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

SIM.PR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report, combine with Monte Carlo calculations, see p. 7 of the SIM.PR-K6.2010 Final Report.

Spectral transmittance
Di
Participating institutes

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report and SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter A

WAVELENGTH      600 nm

 

CCPR-K6.2010

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

SIM.PR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report, combine with Monte Carlo calculations, see p. 7 of the SIM.PR-K6.2010 Final Report.

Spectral transmittance
Di
Participating institutes

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report and SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter A

WAVELENGTH      1000 nm

 

CCPR-K6.2010

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

SIM.PR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report, combine with Monte Carlo calculations, see p. 7 of the SIM.PR-K6.2010 Final Report.

Spectral transmittance
Di
Participating institutes

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report and SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter A

WAVELENGTH      1000 nm

 

Close up of graph.

 

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter C

WAVELENGTH      380 nm

 

CCPR-K6.2010

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

SIM.PR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report, combine with Monte Carlo calculations, see p. 7 of the SIM.PR-K6.2010 Final Report.

Spectral transmittance
Di
Participating institutes

This is the possibility to place a comment.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter C

WAVELENGTH      600 nm

 

CCPR-K6.2010

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

SIM.PR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report, combine with Monte Carlo calculations, see p. 7 of the SIM.PR-K6.2010 Final Report.

Spectral transmittance
Di
Participating institutes

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report and SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter C

WAVELENGTH      600 nm

 

Close up of graph.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter C

WAVELENGTH      1000 nm

 

CCPR-K6.2010

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

SIM.PR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report, combine with Monte Carlo calculations, see p. 7 of the SIM.PR-K6.2010 Final Report.

Spectral transmittance
Di
Participating institutes

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report and SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter C

WAVELENGTH      1000 nm

 

Close up of graph.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter E

WAVELENGTH      380 nm

 

CCPR-K6.2010

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

SIM.PR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report, combine with Monte Carlo calculations, see p. 7 of the SIM.PR-K6.2010 Final Report.

Spectral transmittance
Di
Participating institutes

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report and SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter E

WAVELENGTH      380 nm

 

Close up of graph.

CCPR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter E

WAVELENGTH      600 nm

 

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

Spectral transmittance
Di
Participating institutes

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report.

CCPR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter E

WAVELENGTH      1000 nm

 

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

Spectral transmittance
Di
Participating institutes

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter A

WAVELENGTH      380 nm

 

CCPR-K6.2010

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

SIM.PR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report, combine with Monte Carlo calculations, see p. 7 of the SIM.PR-K6.2010 Final Report.

LABi Di Uneg,i Upos,i
-
-
-
NMC A*STAR 4.4E-04 5.1E-04
KRISS 6.3E-04 2.9E-03
LNE-LNHB -2.3E-04 6.6E-04
MKEH 1.7E-03 2.0E-03
NIST -8.1E-04 9.6E-04
NMIJ AIST 1.0E-03 8.2E-04
NMISA -6.9E-04 7.7E-03
NPL -1.5E-04 3.3E-04
NRC -3.5E-05 3.2E-04
PTB 7.3E-06 6.1E-04
VNIIOFI -5.9E-05 9.9E-04
MSL -2.1E-04 1.7E-04
CENAM 1.8E-03 3.8E-03
INM -2.3E-03 2.0E-03
CMS/ITRI -1.9E-03 2.2E-03
NIM -1.6E-03 1.4E-03
NIMT 2.9E-03 6.3E-03
Participating institutes
CCPR-K6.2010
SIM.PR-K6.2010

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report and SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter A

WAVELENGTH      600 nm

 

CCPR-K6.2010

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

SIM.PR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report, combine with Monte Carlo calculations, see p. 7 of the SIM.PR-K6.2010 Final Report.

LABi Di Uneg,i Upos,i
-
-
-
NMC A*STAR 4.5E-04 5.6E-04
KRISS 3.5E-04 2.5E-03
LNE-LNHB -1.7E-04 2.1E-04
MKEH 2.7E-03 1.8E-03
NIST 3.2E-04 9.6E-04
NMIJ AIST 6.5E-06 5.4E-04
NMISA -1.1E-03 8.2E-03
NPL -2.3E-05 2.7E-04
NRC -8.9E-05 2.3E-04
PTB 2.0E-04 5.6E-04
VNIIOFI -6.1E-05 2.0E-03
MSL -8.8E-05 1.2E-04
CENAM 1.1E-03 3.9E-03
INM -7.9E-04 1.7E-03
INMETRO 2.6E-03 2.2E-03
CMS/ITRI -1.4E-03 2.1E-03
NIM -2.6E-04 1.5E-03
NIMT 3.0E-03 6.5E-03
Participating institutes
CCPR-K6.2010
SIM.PR-K6.2010

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report and SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter A

WAVELENGTH      1000 nm

 

CCPR-K6.2010

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

SIM.PR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report, combine with Monte Carlo calculations, see p. 7 of the SIM.PR-K6.2010 Final Report.

LABi Di Uneg,i Upos,i
-
-
-
NMC A*STAR 7.3E-04 6.1E-04
KRISS 1.4E-05 2.5E-03
LNE-LNHB -2.1E-05 2.8E-04
MKEH 3.8E-03 2.8E-03
NIST 8.6E-04 9.6E-04
NMIJ AIST -2.9E-05 5.3E-04
NMISA 4.2E-03 4.3E-02
NPL -2.1E-05 2.7E-04
NRC -1.4E-04 2.6E-04
PTB 2.7E-04 5.8E-04
VNIIOFI 6.2E-04 2.3E-03
MSL -3.1E-04 1.3E-04
INMETRO 3.1E-03 2.2E-03
CMS/ITRI 2.4E-04 2.2E-03
NIM 2.8E-04 1.3E-03
NIMT 4.0E-03 6.3E-03
Participating institutes
CCPR-K6.2010
SIM.PR-K6.2010

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report and SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter C

WAVELENGTH      380 nm

 

CCPR-K6.2010

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

SIM.PR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report, combine with Monte Carlo calculations, see p. 7 of the SIM.PR-K6.2010 Final Report.

LABi Di Uneg,i Upos,i
-
-
-
NMC A*STAR 2.8E-05 4.4E-04
KRISS 8.7E-04 8.6E-04
LNE-LNHB -6.1E-05 1.5E-04
MKEH 7.4E-05 1.1E-03
NIST 6.0E-04 8.6E-04
NMIJ AIST 7.8E-04 4.7E-04
NMISA -4.3E-05 1.0E-02
NPL -8.4E-05 4.9E-04
NRC -2.1E-04 1.7E-04
PTB -9.1E-05 1.0E-04
VNIIOFI -9.3E-06 2.3E-03
MSL 9.6E-06 1.6E-04
CENAM -7.5E-04 1.4E-03
INM -5.7E-04 1.4E-03
CMS/ITRI -8.0E-04 1.4E-03
NIM -5.5E-04 1.4E-03
NIMT 5.9E-04 2.9E-03
Participating institutes
CCPR-K6.2010
SIM.PR-K6.2010

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report and SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter C

WAVELENGTH      600 nm

 

CCPR-K6.2010

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

SIM.PR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report, combine with Monte Carlo calculations, see p. 7 of the SIM.PR-K6.2010 Final Report.

LABi Di Uneg,i Upos,i
-
-
-
NMC A*STAR 6.7E-05 1.2E-04
KRISS -2.8E-05 3.3E-04
LNE-LNHB -2.8E-05 1.3E-04
MKEH 6.1E-05 3.2E-04
NIST 4.0E-05 1.2E-04
NMIJ AIST 1.2E-04 1.3E-04
NMISA -5.9E-04 9.2E-03
NPL -3.7E-06 4.1E-05
NRC -1.3E-06 4.1E-05
PTB -1.5E-05 3.5E-05
VNIIOFI 2.0E-04 2.0E-04
MSL -4.8E-05 3.5E-05
CENAM 2.6E-05 4.6E-04
INM 1.4E-05 6.2E-04
INMETRO 6.7E-05 2.6E-04
CMS/ITRI -2.5E-05 8.4E-04
NIM -5.1E-05 5.2E-04
NIMT 2.5E-04 3.2E-03
Participating institutes
CCPR-K6.2010
SIM.PR-K6.2010

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report and SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter C

WAVELENGTH      1000 nm

 

CCPR-K6.2010

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

SIM.PR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report, combine with Monte Carlo calculations, see p. 7 of the SIM.PR-K6.2010 Final Report.

LABi Di Uneg,i Upos,i
-
-
-
NMC A*STAR 6.9E-05 1.0E-04
KRISS -8.3E-05 2.7E-04
LNE-LNHB -1.7E-06 1.3E-04
MKEH -9.2E-05 2.9E-04
NIST 7.5E-05 3.2E-05
NMIJ AIST 4.2E-05 1.1E-04
NMISA 4.6E-03 4.3E-02
NPL -5.4E-05 4.5E-05
NRC -2.6E-05 3.5E-05
PTB -4.3E-05 8.7E-05
VNIIOFI 1.2E-05 2.3E-04
MSL 2.0E-06 2.2E-05
INMETRO 1.3E-04 1.9E-04
CMS/ITRI -6.6E-05 1.1E-03
NIM -9.3E-05 3.8E-04
NIMT -4.7E-04 2.8E-03
Participating institutes
CCPR-K6.2010
SIM.PR-K6.2010

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report and SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter E

WAVELENGTH      380 nm

 

CCPR-K6.2010

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

SIM.PR-K6.2010

The degree of equivalence of laboratory i relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report, combine with Monte Carlo calculations, see p. 7 of the SIM.PR-K6.2010 Final Report.

LABi Di Uneg,i Upos,i
-
-
-
NMC A*STAR -5.4E-08 1.0E-06
KRISS 2.1E-06 2.3E-06
LNE-LNHB 4.3E-07 9.8E-07
MKEH -7.4E-07 3.5E-06
NIST 1.4E-05 3.8E-05
NMIJ AIST 2.3E-06 6.8E-06
NRC -2.8E-07 8.7E-07
PTB -2.3E-07 8.4E-07
MSL 5.6E-07 5.6E-07
CENAM 6.3E-06 9.6E-05
CMS/ITRI -7.1E-06 1.3E-05
NIM -3.1E-06 1.3E-05
NIMT 2.7E-05 3.2E-03
Participating institutes
CCPR-K6.2010
SIM.PR-K6.2010

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report and SIM.PR-K6.2010 Final Report.

CCPR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter E

WAVELENGTH      600 nm

 

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

LABi Di Uneg,i Upos,i
-
-
-
NMC A*STAR 5.7E-06 3.3E-06
KRISS 3.4E-06 7.7E-06
LNE-LNHB -1.0E-06 4.2E-06
MKEH 6.7E-06 7.8E-06
NIST 6.3E-06 1.9E-05
NMIJ AIST 8.0E-06 1.2E-05
NPL -1.2E-06 1.9E-06
NRC -1.1E-06 1.2E-06
PTB -1.8E-07 1.5E-06
VNIIOFI 6.0E-06 2.1E-05
MSL -9.0E-07 1.3E-06
Participating institutes
CCPR-K6.2010

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report.

CCPR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter E

WAVELENGTH      1000 nm

 

The degree of equivalence of laboratory relative to the key comparison reference value is given by a pair of terms: Di and its expanded uncertainty (k = 2) Ui, computed according to the description given on pp. 7 to 10 of the CCPR-K6.2010 Final Report.

LABi Di Uneg,i Upos,i
-
-
-
NMC A*STAR 2.1E-05 1.3E-05
KRISS 2.0E-06 2.5E-05
LNE-LNHB 2.4E-06 2.0E-05
MKEH 1.7E-05 3.3E-05
NIST 4.1E-06 1.9E-05
NMIJ AIST -9.6E-06 7.0E-05
NPL -5.5E-06 6.1E-06
NRC 8.1E-07 5.2E-06
PTB -1.1E-05 1.1E-05
VNIIOFI -9.7E-06 3.0E-05
MSL 1.1E-06 3.9E-06
Participating institutes
CCPR-K6.2010

Results are here displayed only for 380 nm, 600 nm and 1000 nm. For data and graphs at other wavelengths, cf. CCPR-K6.2010 Final Report.

Comparison
Comparison close
CC comparison
Linked comparison
SIM.PR-K6.2010
Metrology area, Sub-field Photometry and Radiometry, Properties of materials
Description Spectral regular transmittance
Time of measurements 2016 - 2018
Status Approved for equivalence
Final Reports of the comparisons
References

Appendix H: Relative data

Measurand Spectral regular transmittance
Nominal transmittance values (at 546 nm) of 92 %, 50 %, 10 %, 1 % and 0.1 %

Transfer device 5 neutral colored glass filter plates 50 mm x 50 mm
Comparison type Key Comparison
Consultative Committee CCPR (Consultative Committee for Photometry and Radiometry)
Conducted by SIM (Inter-American Metrology System)
Comments Results published on 17 September 2021.
 
Pilot institute NIST
National Institute of Standards and Technology
United States
Contact person C. Cooksey

+1 301-975-6208
First Name Last Name
wwww@ww.www +356719836 Institute 1 Institute 1 Khmelnitskiy
Pilot laboratory
NIST

National Institute of Standards and Technology, United States, SIM

CENAM

Centro Nacional de Metrologia, Mexico, SIM

INM (CO)

Instituto Nacional de Metrología de Colombia, Colombia, SIM

INMETRO

Instituto Nacional de Metrologia, Qualidade e Tecnologia, Brazil, SIM

NIM

National Institute of Metrology, China, APMP

NIMT

National Institute of Metrology (Thailand), Thailand, APMP

NRC

National Research Council, Canada, SIM

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter A

WAVELENGTH      380 nm

 

The key comparison reference value, xR, is embedded in the caclulation of the Degrees of Equivalence and is not explicitly expressed.

 

Linking SIM.PR-K6.2010 to CCPR-K6.2010

NIST (USA) and NRC (Canada) provides the link between the key comparison SIM.PR-K6.2010 and CCPR-K6.2010, having participated in both comparisons. The linking process is described at see p. 7 of the SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter A

WAVELENGTH      600 nm

 

The key comparison reference value, xR, is embedded in the caclulation of the Degrees of Equivalence and is not explicitly expressed.

 

Linking SIM.PR-K6.2010 to CCPR-K6.2010

NIST (USA) and NRC (Canada) provides the link between the key comparison SIM.PR-K6.2010 and CCPR-K6.2010, having participated in both comparisons. The linking process is described at see p. 7 of the SIM.PR-K6.2010 Final Report.

CCPR-K6

MEASURAND        Spectral regular transmittance

FILTER                    Filter A

WAVELENGTH      1000 nm

 

The key comparison reference value, xR, is embedded in the caclulation of the Degrees of Equivalence and is not explicitly expressed.

 

Linking SIM.PR-K6.2010 to CCPR-K6.2010

NIST (USA) and NRC (Canada) provides the link between the key comparison SIM.PR-K6.2010 and CCPR-K6.2010, having participated in both comparisons. The linking process is described at see p. 7 of the SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter C

WAVELENGTH      380 nm

 

The key comparison reference value, xR, is embedded in the caclulation of the Degrees of Equivalence and is not explicitly expressed.

 

Linking SIM.PR-K6.2010 to CCPR-K6.2010

NIST (USA) and NRC (Canada) provides the link between the key comparison SIM.PR-K6.2010 and CCPR-K6.2010, having participated in both comparisons. The linking process is described at see p. 7 of the SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter C

WAVELENGTH      600 nm

 

The key comparison reference value, xR, is embedded in the caclulation of the Degrees of Equivalence and is not explicitly expressed.

 

Linking SIM.PR-K6.2010 to CCPR-K6.2010

NIST (USA) and NRC (Canada) provides the link between the key comparison SIM.PR-K6.2010 and CCPR-K6.2010, having participated in both comparisons. The linking process is described at see p. 7 of the SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter C

WAVELENGTH      1000 nm

 

The key comparison reference value, xR, is embedded in the caclulation of the Degrees of Equivalence and is not explicitly expressed.

 

Linking SIM.PR-K6.2010 to CCPR-K6.2010

NIST (USA) and NRC (Canada) provides the link between the key comparison SIM.PR-K6.2010 and CCPR-K6.2010, having participated in both comparisons. The linking process is described at see p. 7 of the SIM.PR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter E

WAVELENGTH      380 nm

 

The key comparison reference value, xR, is embedded in the caclulation of the Degrees of Equivalence and is not explicitly expressed.

 

Linking SIM.PR-K6.2010 to CCPR-K6.2010

NIST (USA) and NRC (Canada) provides the link between the key comparison SIM.PR-K6.2010 and CCPR-K6.2010, having participated in both comparisons. The linking process is described at see p. 7 of the SIM.PR-K6.2010 Final Report.

CCPR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter E

WAVELENGTH      600 nm

 

The key comparison reference value, xR, is embedded in the caclulation of the Degrees of Equivalence and is not explicitly expressed.

CCPR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter E

WAVELENGTH      1000 nm

 

The key comparison reference value, xR, is embedded in the caclulation of the Degrees of Equivalence and is not explicitly expressed.

CCPR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter A

WAVELENGTH      380 nm

 

Absolute transmission measurements and associated uncertainties of all participants are available in the Appendix D of the CCPR-K6.2010 Final Report.

CCPR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter A

WAVELENGTH      600 nm

 

Absolute transmission measurements and associated uncertainties of all participants are available in the Appendix D of the CCPR-K6.2010 Final Report.

CCPR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter A

WAVELENGTH      1000 nm

 

Absolute transmission measurements and associated uncertainties of all participants are available in the Appendix D of the CCPR-K6.2010 Final Report.

CCPR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter C

WAVELENGTH      380 nm

 

Absolute transmission measurements and associated uncertainties of all participants are available in the Appendix D of the CCPR-K6.2010 Final Report.

CCPR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter C

WAVELENGTH      600 nm

 

Absolute transmission measurements and associated uncertainties of all participants are available in the Appendix D of the CCPR-K6.2010 Final Report.

CCPR-K6.2010 and SIM.PR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter C

WAVELENGTH      1000 nm

 

Absolute transmission measurements and associated uncertainties of all participants are available in the Appendix D of the CCPR-K6.2010 Final Report.

CCPR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter E

WAVELENGTH      380 nm

 

Absolute transmission measurements and associated uncertainties of all participants are available in the Appendix D of the CCPR-K6.2010 Final Report.

CCPR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter E

WAVELENGTH      600 nm

 

Absolute transmission measurements and associated uncertainties of all participants are available in the Appendix D of the CCPR-K6.2010 Final Report.

CCPR-K6.2010

MEASURAND        Spectral regular transmittance

FILTER                    Filter E

WAVELENGTH      1000 nm

 

Absolute transmission measurements and associated uncertainties of all participants are available in the Appendix D of the CCPR-K6.2010 Final Report.