In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Spain, CIEMAT (Centro de Investigaciones Energéticas, Medioambientales y Tecnológicas)
Ionizing radiation
Activity per unit mass , Single nuclide solution : 1000 Bq/g to 1.00E6 Bq/g
Relative expanded uncertainty 2 %
Hp-Ge Gamma-ray spectrometer, balance
Ga-67, glass ampoule, 3.6 mL
Reference standard : 4π(xe)-γ coincidence counting, set of standard weights
Source of traceability : LMRI, CEM
Approved on 11 April 2016
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-2063
CMC ID : EURAMET-RI-ES-00000IBN-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 100 Bq/g to 1.00E6 Bq/g
Relative expanded uncertainty 2 %
Hp-Ge Gamma-ray spectrometer, balance
Sr-85, glass ampoule, 3.6 mL
Reference standard : 4π(xe)-γ coincidence counting, set of standard weights
Source of traceability : LMRI, CEM
Approved on 11 April 2016
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-2065
CMC ID : EURAMET-RI-ES-00000IBP-1
 
Ionizing radiation
Activity per unit mass , Multi nuclide solution : 20 Bq/g to 1.00E5 Bq/g
Relative expanded uncertainty 1.5 %
2πα grid ionization chamber, balance
Pu-239/Pu240, glass ampoule, 3.6 mL
Reference standard : 2πα (GIC) α counting, set of standard weights
Source of traceability : LMRI, CEM
Approved on 11 April 2016
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-2086
CMC ID : EURAMET-RI-ES-00000ID3-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 20 Bq/g to 1.00E5 Bq/g
Relative expanded uncertainty 1.5 %
2πα grid ionization chamber, balance
Pu-242, glass ampoule, 3.6 mL
Reference standard : 2πα (GIC) α counting, set of standard weights
Source of traceability : CIEMAT, CEM
Approved on 11 April 2016
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-2085
CMC ID : EURAMET-RI-ES-00000ID4-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 20 Bq/g to 1.00E5 Bq/g
Relative expanded uncertainty 1.5 %
2πα grid ionization chamber, balance
Pu-238, glass ampoule, 3.6 mL
Reference standard : 2πα (GIC) α counting, set of standard weights
Source of traceability : LMRI, CEM
Approved on 11 April 2016
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-2085
CMC ID : EURAMET-RI-ES-00000ID2-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 10 Bq/g to 1.00E5 Bq/g
Relative expanded uncertainty 1.5 %
2πα grid ionization chamber, balance
U-232, glass ampoule, 3.6 mL
Reference standard : 2πα (GIC) α counting, set of standard weights
Source of traceability : LMRI, CEM
Approved on 11 April 2016
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-2082
CMC ID : EURAMET-RI-ES-00000ICM-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 1000 Bq/g to 1.00E6 Bq/g
Relative expanded uncertainty 2 %
CIEMAT/NIST liquid scintillation counting, balance
Tl-204, glass ampoule, 3.6 mL
Reference standard : Tritiated water, set of standard weights
Source of traceability : LMRI, CEM
Approved on 11 April 2016
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-2079
CMC ID : EURAMET-RI-ES-00000ICJ-1
 
Ionizing radiation
Activity per unit mass , Single nuclide solution : 100 Bq/g to 1.00E6 Bq/g
Relative expanded uncertainty 2 %
Gamma spectrometer, balance
Nb-95, glass ampoule, 3.6 mL
Reference standard : 4πβ-γ coincidence counting, set of standard weights
Source of traceability : LMRI, CEM
Approved on 11 April 2016
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-2069
CMC ID : EURAMET-RI-ES-00000IBT-1
 
Ionizing radiation
Personal dose equivalent/rate, penetrating , Personal dosimeter : 0.5 Sv to 1.50E4 Sv
Relative expanded uncertainty 5.5 %
Irradiation or calibration on a ISO PMMA slab phantom
Cs-137, ISO 4037. Irradiation or calibration at 4 m using 2 different beams. Integration times 60-60000 s. Levels on January 1, 2003.
Reference standard : Secondary standard ionization chamber
Source of traceability : PTB
Approved on 21 April 2008
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-1024
CMC ID : EURAMET-RI-ES-00000ID0-1
 
Ionizing radiation
Personal dose equivalent/rate, superficial , Dosimeter : 1.50E-3 Sv to 1 Sv
Relative expanded uncertainty 2.9 %
Irradiation on a water slab or PMMA rod phantoms
Beta radiation, Sr-90/Y-90 with ISO 6980 filter. Calibration at 30 cm. Levels on July 17, 2006. Integration time 120 s - 23 h.
Reference standard : Secondary standard beta source
Source of traceability : PTB
Approved on 21 April 2008
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-1011
CMC ID : EURAMET-RI-ES-00000IC6-1
 
Ionizing radiation
Absorbed dose/rate to tissue , Radiation protection dosemeter : 0.044 Gy/h
Relative expanded uncertainty 2.8 %
Calibration free in air
Beta radiation, Sr-90/Y-90 with ISO 6980 filter. Calibration at 30 cm. Levels on July 17, 2006.
Reference standard : Secondary standard beta source
Source of traceability : PTB
Approved on 21 April 2008
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-1007
CMC ID : EURAMET-RI-ES-00000IC2-1
 
Ionizing radiation
Activity , Single nuclide solid source : 100 Bq to 1.00E4 Bq
Relative expanded uncertainty 2 %
Defined solid angle counter (alpha particles)
U-232, thin solid source
Reference standard : Absolute counting by defined solid angle
Source of traceability : LMRI
Approved on 11 April 2016
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-2030
CMC ID : EURAMET-RI-ES-00000IEH-1
 
Ionizing radiation
Activity , Single nuclide solution : 1.00E6 Bq to 1.00E9 Bq
Relative expanded uncertainty 1.5 %
High pressure well type ionization chamber
Co-60, glass ampoule, 3.6 mL
Reference standard : 4πβ-γ coincidence counting
Source of traceability : LMRI
Approved on 11 April 2016
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-2007
CMC ID : EURAMET-RI-ES-00000IDV-1
 
Ionizing radiation
Activity , Single nuclide solid source : 1000 Bq to 3.70E5 Bq
Relative expanded uncertainty 2.5 %
Hp-Ge Gamma-ray spectrometer
Ga-67, point source
Reference standard : 4π(xe)-γ coincidence counting
Source of traceability : LMRI
Approved on 11 April 2016
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-2008
CMC ID : EURAMET-RI-ES-00000IDW-1
 
Ionizing radiation
Activity , Single nuclide solid source : 100 Bq to 1.00E6 Bq
Relative expanded uncertainty 3 %
Hp-Ge Gamma-ray spectrometer
Mn-54, point source
Reference standard : 4π(xe)-γ coincidence counting
Source of traceability : LMRI
Approved on 11 April 2016
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-2004
CMC ID : EURAMET-RI-ES-00000IDS-1
 
Ionizing radiation
Activity , Single nuclide solution : 1.00E6 Bq to 1.00E9 Bq
Relative expanded uncertainty 4.2 %
High pressure well type ionization chamber
Eu-152, glass ampoule, 3.6 mL
Reference standard : 4πγ counting
Source of traceability : LMRI
Approved on 11 April 2016
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-2021
CMC ID : EURAMET-RI-ES-00000IE8-1
 
Ionizing radiation
Activity , Single nuclide solid source : 100 Bq to 3.70E5 Bq
Relative expanded uncertainty 2 %
Hp-Ge Gamma-ray spectrometer
Eu-154, point source
Reference standard : 4πγ counting
Source of traceability : LMRI
Approved on 11 April 2016
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-2022
CMC ID : EURAMET-RI-ES-00000IE9-1
 
Ionizing radiation
Ambient dose equivalent/rate , Radiation protection dosemeter : 1.00E-4 Sv/h to 0.1 Sv/h
Relative expanded uncertainty 4.2 %
Calibration against secondary standard free in air
X-ray, 50 kV to 420 kV, N60 to N300 ISO 4037-1 quality
Reference standard : Secondary standard ionization chamber
Source of traceability : PTB
Approved on 21 April 2008
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-1030
CMC ID : EURAMET-RI-ES-00000IDI-1
 
Ionizing radiation
Ambient dose equivalent/rate , Radiation protection dosemeter : 3.00E-4 Sv/h to 0.3 Sv/h
Relative expanded uncertainty 4.2 %
Calibration against secondary standard free in air
X-ray, 10 kV to 50 kV, N25 to N40 ISO 4037-1 quality
Reference standard : Secondary standard ionization chamber
Source of traceability : PTB
Approved on 21 April 2008
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-1029
CMC ID : EURAMET-RI-ES-00000IDH-1
 
Ionizing radiation
Kerma/rate , Ionization chamber : 3.00E-7 Gy/s to 5.00E-4 Gy/s
Relative expanded uncertainty 1.2 %
Calibration against secondary standard free in air
X-ray, 50 kV to 420 kV, W60 to W300 ISO 4037-1 quality
Reference standard : Secondary standard ionization chamber
Source of traceability : PTB
Approved on 21 April 2008
Institute service identifier : EUR-RAD-CIEMAT(LMRI)-1028
CMC ID : EURAMET-RI-ES-00000IDG-1
 

Select Metrology Area

General physics
Acoustics, Ultrasound, Vibration

This metrology area covers the fields of acoustics, ultrasound and vibration.

Electricity and Magnetism

This metrology area covers the fields of DC and AC measurements, impedance, electric and magnetic fields, radiofrequencies and measurements on materials.

Length

This metrology area covers the fields of laser frequencies and dimensional metrology.

Mass and related quantities

This metrology area covers the fields of mass standards, force, pressure, density, hardness, torque, gravity and viscosity.

Chemistry and Biology
Chemistry and Biology

This metrology area covers all fields relevant to chemistry and biology.

Ionizing radiation
Ionizing Radiation

This metrology area covers the fields of dosimetry, radioactivity and neutron measurements.