Serbia, VINS (Vinca Institute of Nuclear Sciences)
Ionizing radiation
Personal dose equivalent/rate, penetrating
,
Active or passive personal dosimeter
: 9.60E-6 Sv/h to 6.80E-3 Sv/h
Relative expanded uncertainty 4.6 %
Irradiation on phantom in a calibrated field
Cs-137, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1007
CMC ID :
EURAMET-RI-RS-00000K6V-1
Ionizing radiation
Ambient dose equivalent/rate
,
Survey meter, area monitor
: 9.60E-6 Sv/h to 6.80E-3 Sv/h
Relative expanded uncertainty 4.6 %
Irradiation in a calibrated field in air
Cs-137, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1005
CMC ID :
EURAMET-RI-RS-00000K6T-1
Ionizing radiation
Kerma/rate
,
Protection ionisation chamber
: 5.00E-7 Gy to 0.15 Gy
Relative expanded uncertainty 2 %
Calibration against a secondary standard free in air
Cs-137, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1004
CMC ID :
EURAMET-RI-RS-00000K6S-1
Ionizing radiation
Personal dose equivalent/rate, penetrating
,
Active or passive personal dosimeter
: 5.00E-7 Sv to 0.15 Sv
Relative expanded uncertainty 4.6 %
Irradiation on phantom in a calibrated field
Cs-137, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1008
CMC ID :
EURAMET-RI-RS-00000K6W-1
Ionizing radiation
Ambient dose equivalent/rate
,
Survey meter, area monitor
: 2.50E-6 Sv/h to 3 Sv/h
Relative expanded uncertainty 4.6 %
Irradiation in a calibrated field in air
Co-60, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1013
CMC ID :
EURAMET-RI-RS-00000K71-1
Ionizing radiation
Kerma/rate
,
Protection ionisation chamber
: 2.50E-7 Gy to 30 Gy
Relative expanded uncertainty 2 %
Calibration against a secondary standard free in air
Co-60, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1012
CMC ID :
EURAMET-RI-RS-00000K70-1
Ionizing radiation
Personal dose equivalent/rate, superficial
,
Active or passive personal dosimeter
: 2.50E-6 Sv/h to 3 Sv/h
Relative expanded uncertainty 4.6 %
Irradiation on phantom in a calibrated field
Co-60, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1017
CMC ID :
EURAMET-RI-RS-00000K95-1
Ionizing radiation
Personal dose equivalent/rate, superficial
,
Active or passive personal dosimeter
: 9.00E-6 Sv/h to 6.80E-3 Sv/h
Relative expanded uncertainty 4.6 %
Irradiation on phantom in a calibrated field
Cs-137, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1009
CMC ID :
EURAMET-RI-RS-00000K6X-1
Ionizing radiation
Personal dose equivalent/rate, penetrating
,
Active or passive personal dosimeter
: 2.50E-7 Sv to 30 Sv
Relative expanded uncertainty 4.6 %
Irradiation on phantom in a calibrated field
Co-60, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1016
CMC ID :
EURAMET-RI-RS-00000K74-1
Ionizing radiation
Absorbed dose/rate to water
,
Therapy ionisation chamber
: 5.00E-4 Gy/s to 4.40E-3 Gy/s
Relative expanded uncertainty 1.2 %
Calibration against a secondary standard in a water phantom
Co-60, IAEA TRS-398
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1002
CMC ID :
EURAMET-RI-RS-00000K6Q-1
Ionizing radiation
Ambient dose equivalent/rate
,
Survey meter, area monitor
: 5.00E-7 Sv to 0.15 Sv
Relative expanded uncertainty 4.6 %
Irradiation in a calibrated field in air
Cs-137, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1006
CMC ID :
EURAMET-RI-RS-00000K6U-1
Ionizing radiation
Kerma/rate
,
Therapy ionisation chamber
: 5.00E-4 Gy/s to 4.40E-3 Gy/s
Relative expanded uncertainty 1 %
Calibration against a secondary standard free in air
Co-60, IAEA TRS-277
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1001
CMC ID :
EURAMET-RI-RS-00000K6P-1
Ionizing radiation
Personal dose equivalent/rate, superficial
,
Active or passive personal dosimeter
: 2.50E-7 Sv to 30 Sv
Relative expanded uncertainty 4.6 %
Irradiation on phantom in a calibrated field
Co-60, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1018
CMC ID :
EURAMET-RI-RS-00000K96-1
Ionizing radiation
Personal dose equivalent/rate, superficial
,
Active or passive personal dosimeter
: 5.00E-7 Sv to 0.15 Sv
Relative expanded uncertainty 4.6 %
Irradiation on phantom in a calibrated field
Cs-137, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1010
CMC ID :
EURAMET-RI-RS-00000K6Y-1
Ionizing radiation
Kerma/rate
,
Protection ionisation chamber
: 8.00E-3 mGy/h to 5.7 mGy/h
Relative expanded uncertainty 2 %
Calibration against a secondary standard free in air
Cs-137, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1003
CMC ID :
EURAMET-RI-RS-00000K6R-1
Ionizing radiation
Personal dose equivalent/rate, penetrating
,
Active or passive personal dosimeter
: 2.50E-6 Sv/h to 3 Sv/h
Relative expanded uncertainty 4.6 %
Irradiation on phantom in a calibrated field
Co-60, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1015
CMC ID :
EURAMET-RI-RS-00000K73-1
Ionizing radiation
Ambient dose equivalent/rate
,
Survey meter, area monitor
: 2.50E-7 Sv to 30 Sv
Relative expanded uncertainty 4.6 %
Irradiation in a calibrated field in air
Co-60, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1014
CMC ID :
EURAMET-RI-RS-00000K72-1
Ionizing radiation
Kerma/rate
,
Protection ionisation chamber
: 2.50E-3 mGy/h to 3000 mGy/h
Relative expanded uncertainty 2 %
Calibration against a secondary standard free in air
Co-60, ISO 4037
Reference standard : Secondary standard ionization chamber
Source of traceability : IAEA
Approved on 27 November 2019
Institute service identifier : EUR-RAD-VINS-1011
CMC ID :
EURAMET-RI-RS-00000K6Z-1