Results for: Morocco "LPEE/LNM"
Filters
CMC Quick search Filters close;
?
Added keywords act as an additional filter. Keywords separated by a space will be treated by the logical function AND automatically. You may also use OR. Keyword searches also find close matches. Placing quotation marks (") before and after the search word, e.g. "word", will allow for exact matches.

In case of multiple search, please use ADVANCED SEARCH.
20 results
Reset all
CMC Area
To filter, either use the tick boxes or click on the underscored item
Geographical location
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Morocco, LPEE-LNM (Laboratoire National de Métrologie du Laboratoire Public d'Essais et d'Études)
Items for disseminating ITS-90 and PLTS 2000 , Temperature : -40 °C to 230 °C
Temperature sensor with display unit
Absolute expanded uncertainty : 3.0E-2 °C
PRT
Calibration by comparison in alcohol, water and oil bath
Approved on 18 October 2021
CMC ID : AFRIMETS-T-MA-00000MDB-1
 
DC resistance , DC resistance standards and sources: low values : 1.00E-3 Ω to 1 Ω
Fixed resistor, resistance box
Absolute expanded uncertainty : 0.27 µΩ to 5.6E1 µΩ
{"row_1":{"col_1":"Range","col_2":"Expanded uncertainty /&nbsp;Ω"},"row_2":{"col_1":"1 mΩ to 10 mΩ","col_2":"(9.0E-05 <em>R</em> + 18E-08)"},"row_3":{"col_1":"10 mΩ to 100 mΩ","col_2":"(8.1E-05 <em>R</em> + 13E-07)"},"row_4":{"col_1":"100 mΩ to 1 Ω","col_2":"(4.5E-05 <em>R</em> + 11E-06)"}}
Potential difference between terminals of standard resistance
Current : up to 10 A
Temperature : (23 ± 1.5)°C
Hygrometry : (50 ± 10)%
Approved on 23 June 2022
CMC ID : AFRIMETS-EM-MA-00000N7X-1
 
High voltage and current , High AC voltage: AC high voltage: meters : 1 kV to 20 kV
High AC voltage measuring systems, AC voltage sources, AC voltmeters
Absolute expanded uncertainty : 2.0E-2 kV to 0.4 kV
Comparison to a voltmeter with a resistive divider
Frequency : 50 Hz
new CMCs
Approved on 13 May 2022
CMC ID : AFRIMETS-EM-MA-00000N7W-1
 
DC current (up to 100 A) , DC current sources: intermediate values : 0.1 A to 10 A
Current generator, multifunction calibrator
Absolute expanded uncertainty : 2.9E-2 mA to 3.2 mA
{"row_1":{"col_1":"Range","col_2":"Expanded uncertainty (mA)<br>The current <em>I</em> in the uncertainty equations below is in mA.<br type=\"_moz\">"},"row_2":{"col_1":"100 mA to 1 A","col_2":"8.5E-05&nbsp;<em>I</em>​​​​​​​ + 2E-02"},"row_3":{"col_1":"1 A to 5 A","col_2":"8.5E-05&nbsp;<em>I</em>​​​​​​​ + 2E-01"},"row_4":{"col_1":"5 A to 10 A","col_2":"1.2E-04&nbsp;<em>I</em>​​​​​​​ + 2"}}
Potential difference between terminals of standard resistance
Temperature : (23 ± 1.5) °C
Hygrometry : (50 ± 10) %
Approved on 12 January 2024
CMC ID : AFRIMETS-EM-MA-00000P5W-1
 
DC voltage (up to 1100 V) , DC voltage sources, low value ranges : 1.00E-5 V to 10 V
DC voltage source, multifunction calibrator
Absolute expanded uncertainty : 1.2 µV to 3.1E1 µV
{"row_1":{"col_1":"Range","col_2":"Expanded uncertainty (V)"},"row_2":{"col_1":"10 µv to 220 mV","col_2":"1.5E-06&nbsp;<em>U</em>​​​​​​​ + 1.2E-06"},"row_3":{"col_1":"220 mV to 2.2 V","col_2":"2.5E-06&nbsp;<em>U</em>​​​​​​​ + 1E-06"},"row_4":{"col_1":"2.2 V to 10 V","col_2":"2.7E-06&nbsp;<em>U</em>​​​​​​​ + 4E-06"}}
Opposition
Temperature : (23 ± 1.5)°C
Hygrometry : (50 ± 10)%
New CMC
Approved on 12 January 2024
CMC ID : AFRIMETS-EM-MA-00000P5S-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 230 °C to 260 °C
Temperature sensor with display unit
Absolute expanded uncertainty : 6.0E-2 °C
PRT
Calibration by comparison in alcohol, water and oil bath
Approved on 18 October 2021
CMC ID : AFRIMETS-T-MA-00000MDC-1
 
High voltage and current , High DC voltage: DC voltage meter : 1 kV to 100 kV
High DC voltage measuring systems, DC voltage sources, DC voltmeters
Absolute expanded uncertainty : 3.0E-3 kV to 0.3 kV
Comparison to a voltmeter with a resistive divider
new CMCs
Approved on 13 May 2022
CMC ID : AFRIMETS-EM-MA-00000N7V-1
 
Mass , Mass standards : 1.00E-3 g to 5.00E4 g
Mass standards
Absolute expanded uncertainty : 2.0E-3 mg to 7.5E2 mg
{"row_1":{"col_1":"MASS (g)<br>","col_2":"Expanded uncertainty (mg)<br>"},"row_2":{"col_1":"0.001","col_2":"0.002<br>"},"row_3":{"col_1":"0.002","col_2":"0.002"},"row_4":{"col_1":"0.005<br>","col_2":"0.002<br>"},"row_5":{"col_1":"0.01","col_2":"0.002<br>"},"row_6":{"col_1":"0.02<br>","col_2":"0.003<br>"},"row_7":{"col_1":"0.05<br>","col_2":"0.004"},"row_8":{"col_1":"0.1<br>","col_2":"0.005"},"row_9":{"col_1":"0.2","col_2":"0.006"},"row_10":{"col_1":"0,5<br>","col_2":"0.008"},"row_11":{"col_1":"1<br>","col_2":"0.010"},"row_12":{"col_1":"2<br>","col_2":"0.012<br>"},"row_13":{"col_1":"5","col_2":"0.015<br>"},"row_14":{"col_1":"10<br>","col_2":"0.020"},"row_15":{"col_1":"20","col_2":"0.025"},"row_16":{"col_1":"50<br>","col_2":"0.030"},"row_17":{"col_1":"100<br>","col_2":"0.050"},"row_18":{"col_1":"200<br>","col_2":"0.10<br>"},"row_19":{"col_1":"500<br>","col_2":"0.25"},"row_20":{"col_1":"1000<br>","col_2":"0.50"},"row_21":{"col_1":"2000<br>","col_2":"1.0<br>"},"row_22":{"col_1":"5000<br>","col_2":"2.5"},"row_23":{"col_1":"10000<br>","col_2":"5.0"},"row_24":{"col_1":"20000","col_2":"300"},"row_25":{"col_1":"50000","col_2":"750"}}
1mg to 10kg class E2
20kg & 50kg class M1
Direct comparison method
OIML R 111
New CMC
Approved on 25 October 2022
CMC ID : AFRIMETS-M-MA-00000NYA-1
 
Fluid flow , Volume of liquid : 10 µL to 1.00E4 µL
Pipettes
Relative expanded uncertainty : 1.0E-2 % to 0.25 %
{"row_1":{"col_1":"Range (mL)","col_2":"Expanded uncertainty (%)"},"row_2":{"col_1":"0.01 to 0.02","col_2":"0.25"},"row_3":{"col_1":"0.02 to 0.1","col_2":"0.08"},"row_4":{"col_1":"0.1 to 10","col_2":"0.01"}}
Gravimetric
EURAMET Cg19, ISO 8655-6
Approved on 02 February 2022
CMC ID : AFRIMETS-M-MA-00000LAP-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : -40 °C to 230 °C
Industrial Platinum Resistance Thermometer
Absolute expanded uncertainty : 2.5E-2 °C
Calibration by comparison in alcohol, water and oil bath
Approved on 06 July 2021
CMC ID : AFRIMETS-T-MA-00000MD2-1
 
DC resistance , DC resistance standards and sources: intermediate values : 1 Ω to 1.00E6 Ω
Fixed resistor, resistance box
Relative expanded uncertainty : 1.0E1 µΩ/Ω to 2.1E1 µΩ/Ω
{"row_1":{"col_1":"Range","col_2":"Expanded uncertainty (µΩ/Ω)"},"row_2":{"col_1":"1&nbsp;Ω to 10&nbsp;Ω","col_2":"21"},"row_3":{"col_1":"10 Ω to 100&nbsp;Ω","col_2":"14"},"row_4":{"col_1":"100 Ω to 1 kΩ","col_2":"12"},"row_5":{"col_1":"1 kΩ to 10 kΩ","col_2":"10"},"row_6":{"col_1":"10 kΩ to 100 kΩ","col_2":"12"},"row_7":{"col_1":"100 kΩ to 1 MΩ<br>","col_2":"12"}}
Resistance bridge with KV divider
Voltage : up to 50 V
Temperature : (23 ± 1.5)°C
Hygrometry : (50 ± 10)%
Approved on 23 June 2022
CMC ID : AFRIMETS-EM-MA-00000N7Y-1
 
DC resistance , DC resistance standards and sources: high values : 1.00E6 Ω to 1.00E7 Ω
Fixed resistor, resistance box
Relative expanded uncertainty : 3.0E1 µΩ/Ω
Resistance bridge with KV divider
Voltage : up to 100 V
Temperature : (23 ± 1.5)°C
Hygrometry : (50 ± 10)%
new CMCs
Approved on 23 June 2022
CMC ID : AFRIMETS-EM-MA-00000N82-1
 
DC voltage (up to 1100 V) , DC voltage sources, single values : 1.018 V to 10 V
DC solid state voltage standard
Absolute expanded uncertainty : 3 µV to 1.5E1 µV
{"row_1":{"col_1":"Fixed values","col_2":"Expanded uncertainty"},"row_2":{"col_1":"1.018 V","col_2":"3 µV"},"row_3":{"col_1":"10 V","col_2":"15 µV"}}
Substitution comparison with refrence standard of same value
new CMCs
Approved on 26 April 2022
CMC ID : AFRIMETS-EM-MA-00000N7M-1
 
DC voltage (up to 1100 V) , DC voltage sources: intermediate values : 10 V to 1000 V
DC voltage source, multifunction calibrator
Absolute expanded uncertainty : 3.8E-2 mV to 3.3 mV
{"row_1":{"col_1":"Range","col_2":"Expanded uncertainty (V)"},"row_2":{"col_1":"10 V to 22 V","col_2":"2.8E-06&nbsp;<em>U</em>​​​​​​​ + 7E-06"},"row_3":{"col_1":"22 V to 100 V","col_2":"2.8E-06&nbsp;<em>U</em>​​​​​​​ + 3E-05"},"row_4":{"col_1":"100 V to 220 V","col_2":"3.0E-06&nbsp;<em>U</em>​​​​​​​ + 7E-05"},"row_5":{"col_1":"220 V to 1000 V","col_2":"3.0E-06&nbsp;<em>U</em>​​​​​​​ + 3E-04"}}
Opposition
Temperature : (23 ± 1.5)°C
Hygrometry : (50 ± 10)%
new CMC
Approved on 12 January 2024
CMC ID : AFRIMETS-EM-MA-00000P5T-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : 230 °C to 260 °C
Industrial Platinum Resistance Thermometer
Absolute expanded uncertainty : 5.0E-2 °C
Calibration by comparison in alcohol, water and oil bath
Approved on 18 October 2021
CMC ID : AFRIMETS-T-MA-00000MD9-1
 
DC resistance , DC resistance standards and sources: high values : 1.00E7 Ω to 1.00E12 Ω
Fixed resistor, resistance box
Relative expanded uncertainty : 8.0E-2 mΩ/Ω to 4 mΩ/Ω
{"row_1":{"col_1":"Range","col_2":"Expanded uncertainty (mΩ/Ω)"},"row_2":{"col_1":"10 MΩ to 1 GΩ","col_2":"0.08"},"row_3":{"col_1":"1 GΩ to 100 GΩ","col_2":"0.3"},"row_4":{"col_1":"100 GΩ to 1 TΩ","col_2":"4"}}
Expanded uncertainty may be increased for other calibration voltage
2 DC voltage sources method
Voltage : 100 V
Temperature : (23 ± 1.5)°C
Hygrometry : (50 ± 10)%
new CMCs
Approved on 23 June 2022
CMC ID : AFRIMETS-EM-MA-00000N83-1
 
DC resistance , DC resistance standards and sources: high values : 1.00E7 Ω
Fixed resistor
Relative expanded uncertainty : 2.0E1 µΩ/Ω
Substitution method
Power : below 10 mW
Temperature : (23 ± 1.5)°C
Hygrometry : (50 ± 10)%
new CMCs
Approved on 23 June 2022
CMC ID : AFRIMETS-EM-MA-00000N81-1
 
DC current (up to 100 A) , DC current sources: intermediate values : 1.00E-4 A to 0.1 A
Current generator, multifunction calibrator
Absolute expanded uncertainty : 6.5E-3 µA to 5 µA
{"row_1":{"col_1":"Range","col_2":"Expanded uncertainty (&micro;A)<br />\nThe current <em>I</em> in the uncertainty equations below is in &micro;A"},"row_2":{"col_1":"100 µA to 1 mA","col_2":"2.3E-05&nbsp;<em>I</em>​​​​​​​ + 4E-03"},"row_3":{"col_1":"1 mA to 10 mA","col_2":"3.0E-05&nbsp;<em>I</em>​​​​​​​ + 2E-01"},"row_4":{"col_1":"10 mA to 100 mA","col_2":"3.0E-05&nbsp;<em>I</em>​​​​​​​ + 2"}}
Potential difference between terminals of standard resistance
Temperature : (23 ± 1.5) °C
Hygrometry : (50 ± 10) %
Approved on 12 January 2024
CMC ID : AFRIMETS-EM-MA-00000P5V-1
 
DC resistance , DC resistance standards and sources: intermediate values : 1 Ω to 1.00E6 Ω
Fixed resistor
Relative expanded uncertainty : 2 µΩ/Ω to 1.0E1 µΩ/Ω
{"row_1":{"col_1":"<br>","col_2":"Expanded uncertainty (µΩ/Ω)"},"row_2":{"col_1":"1&nbsp;Ω","col_2":"4"},"row_3":{"col_1":"10&nbsp;Ω","col_2":"4"},"row_4":{"col_1":"100&nbsp;Ω","col_2":"3"},"row_5":{"col_1":"1 kΩ","col_2":"3"},"row_6":{"col_1":"10 kΩ<br>","col_2":"2"},"row_7":{"col_1":"100 kΩ","col_2":"6"},"row_8":{"col_1":"1 MΩ","col_2":"10"}}
Substitution method
Power : below 10 mW
Temperature : (23 ± 1.5)°C
Hygrometry : (50 ± 10)%
new CMCs
Approved on 23 June 2022
CMC ID : AFRIMETS-EM-MA-00000N7Z-1
 
DC current (up to 100 A) , DC current sources: low values : 1.00E-6 A to 1.00E-4 A
Current generator, multifunction calibrator
Absolute expanded uncertainty : 0.22 nA to 2.3 nA
{"row_1":{"col_1":"Range","col_2":"Expanded uncertainty (µA)"},"row_2":{"col_1":"1 µA to 100 µA","col_2":"(2.1E-05&nbsp;<em>I</em>​​​​​​​ + 2E-04)<br />\nwith<em> I</em>&nbsp; in &micro;A"}}
Potential difference between terminals of standard resistance
Temperature : (23 ± 1.5) °C
Hygrometry : (50 ± 10) %
Approved on 12 January 2024
CMC ID : AFRIMETS-EM-MA-00000P5U-1
 

Select Metrology Area

General physics
Acoustics, Ultrasound, Vibration

This metrology area covers the fields of acoustics, ultrasound and vibration.

Electricity and Magnetism

This metrology area covers the fields of DC and AC measurements, impedance, electric and magnetic fields, radiofrequencies and measurements on materials.

Length

This metrology area covers the fields of laser frequencies and dimensional metrology.

Mass and related quantities

This metrology area covers the fields of mass standards, force, pressure, density, hardness, torque, gravity and viscosity.

Chemistry and Biology
Chemistry and Biology

This metrology area covers all fields relevant to chemistry and biology.

Ionizing radiation
Ionizing Radiation

This metrology area covers the fields of dosimetry, radioactivity and neutron measurements.