In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
France, LNE-LNHB (Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel)
Ionizing radiation
Personal dose equivalent/rate, superficial , Personal dosimeter : 2.00E-5 Sv to 5 Sv
Relative expanded uncertainty 6.5 %
Calibration in a calibrated field using a phantom
Beta radiation, ISO 6980, Pm-147, Kr-85, Sr-90/Y-90
Reference standard : Primary extrapolation ionisation chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1002
CMC ID : EURAMET-RI-FR-00000KZH-1
 
Ionizing radiation
Personal dose equivalent/rate, penetrating , Personal dosimeter : 5.00E-6 Sv/s to 0.01 Sv/s
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 10 kV to 50 kV, 10 kV to 50 kV, CCRI series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1069
CMC ID : EURAMET-RI-FR-00000KZ5-1
 
Ionizing radiation
Absorbed dose/rate to water , Dosemeter : 0.2 Gy to 1800 Gy
Relative expanded uncertainty 0.5 %
Calibration in a calibrated field in water phantom
Co-60, therapy level
Reference standard : Absorbed dose to graphite by calorimetry and transfer to absorbed dose to water based on calculations and absorbed dose to water by calorimetry
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1014
CMC ID : EURAMET-RI-FR-00000KZU-1
 
Ionizing radiation
Absorbed dose/rate to water , Dosemeter : 0.3 Gy to 30 Gy
Relative expanded uncertainty 2.7 %
Calibration against a transfer standard in a water phantom
Electrons, medical type accelerator: 9 MeV, 12 MeV and 18 MeV (R50 from 3.5 g/cm2 to 7.3 g/cm2)
Reference standard : Co-60 absorbed dose to water and TRS 398
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1016
CMC ID : EURAMET-RI-FR-00000KZR-1
 
Ionizing radiation
Kerma/rate , Dosemeter : 1.50E-8 Gy to 200 Gy
Relative expanded uncertainty 1 %
Calibration in a calibrated field free in air
Co-60, ISO 4037
Reference standard : Primary standard graphite air cavity ionisation chamber and transfer chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1020
CMC ID : EURAMET-RI-FR-00000L0G-1
 
Ionizing radiation
Personal dose equivalent/rate, superficial , Personal dosimeter : 5.00E-6 Sv/s to 0.01 Sv/s
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 50 kV to 420 kV, 60 kV to 300 kV, ISO 4037 high dose rate series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1077
CMC ID : EURAMET-RI-FR-00000KZO-1
 
Ionizing radiation
Personal dose equivalent/rate, superficial , Personal dosimeter : 1.00E-5 Sv to 0.15 Sv
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 50 kV to 420 kV, 50 kV to 150 kV, IEC 61267 RQR series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1081
CMC ID : EURAMET-RI-FR-00000KZQ-1
 
Ionizing radiation
Personal dose equivalent/rate, superficial , Personal dosimeter : 1.00E-5 Sv to 0.1 Sv
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 10 kV to 50 kV, 10 kV to 50 kV, IEC 61267 RQR-M series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1082
CMC ID : EURAMET-RI-FR-00000KZM-1
 
Ionizing radiation
Personal dose equivalent/rate, superficial , Personal dosimeter : 5.00E-6 Sv/s to 0.01 Sv/s
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 50 kV to 420 kV, 100 kV to 250 kV, CCRI series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1079
CMC ID : EURAMET-RI-FR-00000KZP-1
 
Ionizing radiation
Personal dose equivalent/rate, superficial , Personal dosimeter : 1.00E-5 Sv to 0.15 Sv
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 10 kV to 50 kV, 40 kV, IEC 61267 RQR series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1080
CMC ID : EURAMET-RI-FR-00000KZL-1
 
Ionizing radiation
Kerma/rate , Dosemeter : 5.00E-6 Gy/s to 0.01 Gy/s
Relative expanded uncertainty 0.7 %
Free air chamber
X-ray, 50 kV to 420 kV, 100 kV to 250 kV, CCRI series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1045
CMC ID : EURAMET-RI-FR-00000L09-1
 
Ionizing radiation
Activity , Single nuclide solution : 1.00E4 Bq to 5.00E7 Bq
Relative expanded uncertainty 1.5 %
Secondary standard ionisation chamber
Lu-177, glass ampoule
Reference standard : 4πβγ-anticoincidence counting, TDCR liquid scintillation counting.
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 21 September 2016
Institute service identifier : EUR-RAD-LNE-LNHB-2119
CMC ID : EURAMET-RI-FR-00000L1W-1
 
Ionizing radiation
Activity , Single nuclide solution : 1.00E4 Bq to 1.00E10 Bq
Relative expanded uncertainty 0.6 %
Secondary standard ionisation chamber
Sb-124, glass ampoule
Reference standard : 4πβγ-anticoincidence counting
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 21 September 2016
Institute service identifier : EUR-RAD-LNE-LNHB-2163
CMC ID : EURAMET-RI-FR-00000L20-1
 
Ionizing radiation
Activity , Single nuclide solution : 1.00E4 Bq to 5.00E7 Bq
Relative expanded uncertainty 0.5 %
Secondary standard ionisation chamber
Na-24, glass ampoule
Reference standard : 4πβγ-coincidence counting
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 21 September 2016
Institute service identifier : EUR-RAD-LNE-LNHB-2134
CMC ID : EURAMET-RI-FR-00000L1L-1
 
Ionizing radiation
Kerma/rate , Dosemeter : 0.01 Gy to 300 Gy
Relative expanded uncertainty 1 %
Calibration in a calibrated field free in air
Cs-137, therapy level
Reference standard : Primary standard graphite air cavity ionisation chamber and transfer chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1012
CMC ID : EURAMET-RI-FR-00000L0K-1
 
Ionizing radiation
Ambient dose equivalent/rate , Dosemeter : 1.00E-5 Sv to 0.15 Sv
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 50 kV to 420 kV, 50 kV to 150 kV, IEC 61267 RQR series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1055
CMC ID : EURAMET-RI-FR-00000L0X-1
 
Ionizing radiation
Directional dose equivalent/rate , Dosemeter : 5.00E-6 Sv/s to 0.01 Sv/s
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 50 kV to 420 kV, 100 kV to 250 kV, CCRI series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1062
CMC ID : EURAMET-RI-FR-00000L1C-1
 
Ionizing radiation
Ambient dose equivalent/rate , Dosemeter : 5.00E-4 Sv to 0.1 Sv
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 10 kV to 50 kV, 10 kV to 50 kV, IEC 61267 RQR-M series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1056
CMC ID : EURAMET-RI-FR-00000L0T-1
 
Ionizing radiation
Directional dose equivalent/rate , Dosemeter : 5.00E-6 Sv/s to 0.01 Sv/s
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 10 kV to 50 kV, 10 kV to 50 kV, CCRI series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1061
CMC ID : EURAMET-RI-FR-00000L17-1
 
Ionizing radiation
Ambient dose equivalent/rate , Dosemeter : 1.00E-5 Sv to 0.15 Sv
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 10 kV to 50 kV, 40 kV. IEC 61267 RQR series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1054
CMC ID : EURAMET-RI-FR-00000L0S-1
 

Select Metrology Area

General physics
Acoustics, Ultrasound, Vibration

This metrology area covers the fields of acoustics, ultrasound and vibration.

Electricity and Magnetism

This metrology area covers the fields of DC and AC measurements, impedance, electric and magnetic fields, radiofrequencies and measurements on materials.

Length

This metrology area covers the fields of laser frequencies and dimensional metrology.

Mass and related quantities

This metrology area covers the fields of mass standards, force, pressure, density, hardness, torque, gravity and viscosity.

Chemistry and Biology
Chemistry and Biology

This metrology area covers all fields relevant to chemistry and biology.

Ionizing radiation
Ionizing Radiation

This metrology area covers the fields of dosimetry, radioactivity and neutron measurements.