France, LNE-LNHB (Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel)
Ionizing radiation
Personal dose equivalent/rate, superficial
,
Personal dosimeter
: 2.00E-5 Sv to 5 Sv
Relative expanded uncertainty 6.5 %
Calibration in a calibrated field using a phantom
Beta radiation, ISO 6980, Pm-147, Kr-85, Sr-90/Y-90
Reference standard : Primary extrapolation ionisation chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1002
CMC ID :
EURAMET-RI-FR-00000KZH-1
Ionizing radiation
Personal dose equivalent/rate, penetrating
,
Personal dosimeter
: 5.00E-6 Sv/s to 0.01 Sv/s
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 10 kV to 50 kV, 10 kV to 50 kV, CCRI series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1069
CMC ID :
EURAMET-RI-FR-00000KZ5-1
Ionizing radiation
Absorbed dose/rate to water
,
Dosemeter
: 0.2 Gy to 1800 Gy
Relative expanded uncertainty 0.5 %
Calibration in a calibrated field in water phantom
Co-60, therapy level
Reference standard : Absorbed dose to graphite by calorimetry and transfer to absorbed dose to water based on calculations and absorbed dose to water by calorimetry
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1014
CMC ID :
EURAMET-RI-FR-00000KZU-1
Ionizing radiation
Absorbed dose/rate to water
,
Dosemeter
: 0.3 Gy to 30 Gy
Relative expanded uncertainty 2.7 %
Calibration against a transfer standard in a water phantom
Electrons, medical type accelerator: 9 MeV, 12 MeV and 18 MeV (R50 from 3.5 g/cm2 to 7.3 g/cm2)
Reference standard : Co-60 absorbed dose to water and TRS 398
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1016
CMC ID :
EURAMET-RI-FR-00000KZR-1
Ionizing radiation
Kerma/rate
,
Dosemeter
: 1.50E-8 Gy to 200 Gy
Relative expanded uncertainty 1 %
Calibration in a calibrated field free in air
Co-60, ISO 4037
Reference standard : Primary standard graphite air cavity ionisation chamber and transfer chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1020
CMC ID :
EURAMET-RI-FR-00000L0G-1
Ionizing radiation
Personal dose equivalent/rate, superficial
,
Personal dosimeter
: 5.00E-6 Sv/s to 0.01 Sv/s
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 50 kV to 420 kV, 60 kV to 300 kV, ISO 4037 high dose rate series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1077
CMC ID :
EURAMET-RI-FR-00000KZO-1
Ionizing radiation
Personal dose equivalent/rate, superficial
,
Personal dosimeter
: 1.00E-5 Sv to 0.15 Sv
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 50 kV to 420 kV, 50 kV to 150 kV, IEC 61267 RQR series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1081
CMC ID :
EURAMET-RI-FR-00000KZQ-1
Ionizing radiation
Personal dose equivalent/rate, superficial
,
Personal dosimeter
: 1.00E-5 Sv to 0.1 Sv
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 10 kV to 50 kV, 10 kV to 50 kV, IEC 61267 RQR-M series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1082
CMC ID :
EURAMET-RI-FR-00000KZM-1
Ionizing radiation
Personal dose equivalent/rate, superficial
,
Personal dosimeter
: 5.00E-6 Sv/s to 0.01 Sv/s
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 50 kV to 420 kV, 100 kV to 250 kV, CCRI series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1079
CMC ID :
EURAMET-RI-FR-00000KZP-1
Ionizing radiation
Personal dose equivalent/rate, superficial
,
Personal dosimeter
: 1.00E-5 Sv to 0.15 Sv
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 10 kV to 50 kV, 40 kV, IEC 61267 RQR series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1080
CMC ID :
EURAMET-RI-FR-00000KZL-1
Ionizing radiation
Kerma/rate
,
Dosemeter
: 5.00E-6 Gy/s to 0.01 Gy/s
Relative expanded uncertainty 0.7 %
Free air chamber
X-ray, 50 kV to 420 kV, 100 kV to 250 kV, CCRI series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1045
CMC ID :
EURAMET-RI-FR-00000L09-1
Ionizing radiation
Activity
,
Single nuclide solution
: 1.00E4 Bq to 5.00E7 Bq
Relative expanded uncertainty 1.5 %
Secondary standard ionisation chamber
Lu-177, glass ampoule
Reference standard : 4πβγ-anticoincidence counting, TDCR liquid scintillation counting.
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 21 September 2016
Institute service identifier : EUR-RAD-LNE-LNHB-2119
CMC ID :
EURAMET-RI-FR-00000L1W-1
Ionizing radiation
Activity
,
Single nuclide solution
: 1.00E4 Bq to 1.00E10 Bq
Relative expanded uncertainty 0.6 %
Secondary standard ionisation chamber
Sb-124, glass ampoule
Reference standard : 4πβγ-anticoincidence counting
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 21 September 2016
Institute service identifier : EUR-RAD-LNE-LNHB-2163
CMC ID :
EURAMET-RI-FR-00000L20-1
Ionizing radiation
Activity
,
Single nuclide solution
: 1.00E4 Bq to 5.00E7 Bq
Relative expanded uncertainty 0.5 %
Secondary standard ionisation chamber
Na-24, glass ampoule
Reference standard : 4πβγ-coincidence counting
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 21 September 2016
Institute service identifier : EUR-RAD-LNE-LNHB-2134
CMC ID :
EURAMET-RI-FR-00000L1L-1
Ionizing radiation
Kerma/rate
,
Dosemeter
: 0.01 Gy to 300 Gy
Relative expanded uncertainty 1 %
Calibration in a calibrated field free in air
Cs-137, therapy level
Reference standard : Primary standard graphite air cavity ionisation chamber and transfer chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1012
CMC ID :
EURAMET-RI-FR-00000L0K-1
Ionizing radiation
Ambient dose equivalent/rate
,
Dosemeter
: 1.00E-5 Sv to 0.15 Sv
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 50 kV to 420 kV, 50 kV to 150 kV, IEC 61267 RQR series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1055
CMC ID :
EURAMET-RI-FR-00000L0X-1
Ionizing radiation
Directional dose equivalent/rate
,
Dosemeter
: 5.00E-6 Sv/s to 0.01 Sv/s
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 50 kV to 420 kV, 100 kV to 250 kV, CCRI series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1062
CMC ID :
EURAMET-RI-FR-00000L1C-1
Ionizing radiation
Ambient dose equivalent/rate
,
Dosemeter
: 5.00E-4 Sv to 0.1 Sv
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 10 kV to 50 kV, 10 kV to 50 kV, IEC 61267 RQR-M series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1056
CMC ID :
EURAMET-RI-FR-00000L0T-1
Ionizing radiation
Directional dose equivalent/rate
,
Dosemeter
: 5.00E-6 Sv/s to 0.01 Sv/s
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 10 kV to 50 kV, 10 kV to 50 kV, CCRI series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1061
CMC ID :
EURAMET-RI-FR-00000L17-1
Ionizing radiation
Ambient dose equivalent/rate
,
Dosemeter
: 1.00E-5 Sv to 0.15 Sv
Relative expanded uncertainty 4.1 %
Free air chamber
X-ray, 10 kV to 50 kV, 40 kV. IEC 61267 RQR series
Reference standard : Free air chamber
Source of traceability : LNE-LNHB
Service provided by the LNE-LNHB
Approved on 11 September 2013
Institute service identifier : EUR-RAD-LNE-LNHB-1054
CMC ID :
EURAMET-RI-FR-00000L0S-1