In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a 2 + b 2]1/2
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
Sweden, RISE (Research Institutes of Sweden AB)
Matrix or material Dissemination range of measurement capability Range of certified values in reference materials
Electrolytic conductivity in S/m Absolute expanded uncertainty in % Electrolytic conductivity in Absolute expanded uncertainty in
aqueous solution 3.00E-3 to 1.5 2 to 0.4
  • Mechanism for measurement service delivery : Calibration
  • Temperature range: 15 °C to 40 °C
  • Approved on 06 September 2012
  • Institute service identifier : F02
  • CMC ID : EURAMET-QM-SE-000000ZP-1
     
Matrix or material Dissemination range of measurement capability Range of certified values in reference materials
Electrolytic conductivity in µS/m Absolute expanded uncertainty in % Electrolytic conductivity in Absolute expanded uncertainty in
aqueous solution 20 to 3000 3 to 2
  • Mechanism for measurement service delivery : Calibration
  • Temperature range: 15 °C to 40 °C
  • Approved on 19 June 2014
  • Institute service identifier : F01
  • CMC ID : EURAMET-QM-SE-000000ZQ-1
     
Fluid flow , Volume flowrate : 1.70E-3 L/s to 1.5 L/s
Flow meter
Relative expanded uncertainty : 0.1 %
Pulse, analog or visual output
Liquid : water
Temperature : 30 °C to 90 °C
Pipe size : DN 25
Approved on 19 December 2012
Institute service identifier : SE29
CMC ID : EURAMET-M-SE-000008GY-1
 
Fluid flow , Mass flowrate : 0.015 kg/s to 6.5 kg/s
Flow meter
Relative expanded uncertainty : 6.0E-2 %
Pulse, analog or visual output
Liquid : water
Temperature : 10 °C to 90 °C
Pipe size : DN 40
Approved on 19 December 2012
Institute service identifier : SE32
CMC ID : EURAMET-M-SE-000008H4-1
 
Fluid flow , Volume : 0.1 L to 1 L
Glassware
Relative expanded uncertainty : 3.0E-3 %
Any type of instrument
Temperature : room temperature
Contained / delivered volume
Approved on 19 December 2012
Institute service identifier : SE2
CMC ID : EURAMET-M-SE-000008HB-1
 
Density , Volume of solid : 1.2 cm3 to 12.5 cm3
Mass standard: 10 g to 0.1 kg
Absolute expanded uncertainty : 0.3 mm3 to 0.6 mm3
Hydrostatic weighing
Reference temperature : 21 °C
Approved on 21 October 2001
CMC ID : EURAMET-M-SE-000008GL-1
 
Measurement microphones , Pressure sensitivity level
Measurement microphone type LS2
Absolute expanded uncertainty : 0.1 dB
IEC 61094-2:1992
Frequency : 12.5 kHz
Approved on 21 October 2001
CMC ID : EURAMET-AUV-SE-000008HY-1
 
Linear vibration , Charge sensitivity (magnitude)
Accelerometer
Relative expanded uncertainty : 0.5 %
ISO 16063-11
Frequency : 630 Hz to 1 kHz
Approved on 21 October 2001
CMC ID : EURAMET-AUV-SE-000008I9-1
 
Force , Force: compression, tension : 1 N to 2000 N
Force measuring device
Relative expanded uncertainty : 1.0E-2 %
Deadweight
Approved on 21 October 2001
CMC ID : EURAMET-M-SE-000008HL-1
 
Pressure , Absolute pressure : 7.00E5 Pa to 7.40E6 Pa
Pressure balance
Absolute expanded uncertainty
4.6 + 3.0E-05p

p pressure in Pa

The uncertainty is expressed in Pa

Gas
Uncertainty values range from 2.6E+01 Pa to 2.3E+02 Pa
Approved on 02 July 2008
CMC ID : EURAMET-M-SE-000008GW-1
 
Pressure , Absolute pressure : 5.00E-6 Pa to 5.00E-4 Pa
Vacuum gauge
Absolute expanded uncertainty
12E-02p

p pressure in Pa

The uncertainty is expressed in Pa

Gas
Uncertainty values range from 6E-07 Pa to 6E-05 Pa
Approved on 02 July 2008
CMC ID : EURAMET-M-SE-000008GQ-1
 
Items for defining ITS-90 , Temperature : 962 °C to 1600 °C
Radiation thermometer
Absolute expanded uncertainty : 1.5 °C
Calibration against Gold fixed point blackbody
Working wavelength : 651 nm
Approved on 18 May 2004
Institute service identifier : 25
CMC ID : EURAMET-T-SE-00000DHI-1
 
Items for disseminating ITS-90 and PLTS 2000 , Temperature : -196 °C
Platinum resistance thermometers
Absolute expanded uncertainty : 5 mK
Comparison
Ambient temperature : (23 ± 1) °C
Approved on 03 November 2009
Institute service identifier : 27
CMC ID : EURAMET-T-SE-00000DHK-1
 
Items for defining ITS-90 , Temperature : 660.323 °C
Long-stem SPRT and HTSPRT
Absolute expanded uncertainty : 4 mK
Fixed point (aluminium)
Ambient temperature : 23 °C
Approved on 18 May 2004
Institute service identifier : 17
CMC ID : EURAMET-T-SE-00000DHG-1
 
Items for defining ITS-90 , Temperature : 961.78 °C
Long-stem HTSPRT
Absolute expanded uncertainty : 1.0E1 mK
Fixed point (silver)
Ambient temperature : 23 °C
Approved on 18 May 2004
Institute service identifier : 18
CMC ID : EURAMET-T-SE-00000DHH-1
 
Items for defining ITS-90 , Temperature : 156.5985 °C
Long-stem SPRT
Absolute expanded uncertainty : 1.5 mK
Fixed point (indium)
Ambient temperature : 23 °C
Approved on 18 May 2004
Institute service identifier : 14
CMC ID : EURAMET-T-SE-00000DHD-1
 
Items for defining ITS-90 , Temperature : 419.527 °C
Long-stem SPRT
Absolute expanded uncertainty : 2 mK
Fixed point (zinc)
Ambient temperature : 23 °C
Approved on 18 May 2004
Institute service identifier : 16
CMC ID : EURAMET-T-SE-00000DHF-1
 
Items for defining ITS-90 , Temperature : 660.323 °C
Aluminium point
Absolute expanded uncertainty : 4 mK
Direct comparison
Ambient temperature : 23 °C
Approved on 18 May 2004
Institute service identifier : 8
CMC ID : EURAMET-T-SE-00000DH9-1
 
Items for defining ITS-90 , Temperature : 419.527 °C
Zinc point
Absolute expanded uncertainty : 1.5 mK
Direct comparison
Ambient temperature : 23 °C
Approved on 20 May 2010
Institute service identifier : 7
CMC ID : EURAMET-T-SE-00000DHV-1
 
DC resistance , DC resistance standards and sources: intermediate values : 1 Ω to 1.00E4 Ω
Fixed resistor
Relative expanded uncertainty : 0.25 µΩ/Ω
{"row_1":{"col_1":"","col_2":"Voltage","col_3":"Expanded uncertainty"},"row_2":{"col_1":"0.01 mΩ to 0.1 mΩ","col_2":"-","col_3":"20"},"row_3":{"col_1":"0.1 mΩ to 1 mΩ","col_2":"-","col_3":"7"},"row_4":{"col_1":"1 mΩ to 10 mΩ","col_2":"-","col_3":"4"},"row_5":{"col_1":"10 mΩ to 100 mΩ","col_2":"-","col_3":"2"},"row_6":{"col_1":"100 mΩ to 1 Ω","col_2":"-","col_3":"0.5"},"row_7":{"col_1":"1 Ω to 10 kΩ","col_2":"-","col_3":"0.25"},"row_8":{"col_1":"10 kΩ to 100 kΩ","col_2":"< 10 V","col_3":"2"},"row_9":{"col_1":"100 kΩ to 1 MΩ","col_2":"< 10 V","col_3":"4"},"row_10":{"col_1":"1 MΩ to 10 MΩ","col_2":"< 10 V","col_3":"5"},"row_11":{"col_1":"10 Ω to 100 MΩ","col_2":"1 V to 100 V","col_3":"7"},"row_12":{"col_1":"100 MΩ to 1 GΩ","col_2":"1 V to 1000 V","col_3":"15"},"row_13":{"col_1":"1 GΩ to 10 GΩ","col_2":"1 V to 1000 V","col_3":"50"},"row_14":{"col_1":"10 GΩ to 100 GΩ","col_2":"10 V to 1000 V","col_3":"100"},"row_15":{"col_1":"100 GΩ to 1 TΩ","col_2":"50 V to 1000 V","col_3":"300"},"row_16":{"col_1":"1 TΩ to 10 TΩ","col_2":"100 V to 1000 V","col_3":"500"},"row_17":{"col_1":"10 TΩ to 100 TΩ","col_2":"100 V to 1000 V","col_3":"1000"}}
The expanded uncertainties given in this table are expressed in µΩ/Ω
DCC bridge
Power : 10 mW
Approved on 06 August 2013
Institute service identifier : 15
CMC ID : EURAMET-EM-SE-00000EW8-1
 

Select Metrology Area

General physics
Acoustics, Ultrasound, Vibration

This metrology area covers the fields of acoustics, ultrasound and vibration.

Electricity and Magnetism

This metrology area covers the fields of DC and AC measurements, impedance, electric and magnetic fields, radiofrequencies and measurements on materials.

Length

This metrology area covers the fields of laser frequencies and dimensional metrology.

Mass and related quantities

This metrology area covers the fields of mass standards, force, pressure, density, hardness, torque, gravity and viscosity.

Chemistry and Biology
Chemistry and Biology

This metrology area covers all fields relevant to chemistry and biology.

Ionizing radiation
Ionizing Radiation

This metrology area covers the fields of dosimetry, radioactivity and neutron measurements.