Results for tag: LNHB-XHE
In the CMC uncertainty statements, Q[a,b] = [a² + b²]
Unless otherwise stated the expanded uncertainties given below correspond to k = 2 (at a 95 % level of confidence)
France, LNE-LNHB (Commissariat à l'énergie atomique / Laboratoire National Henri Becquerel)
Ionizing radiation
Absorbed dose/rate , Passive dosimeter : 0.3 Gy to 30 Gy
Relative expanded uncertainty 0.89 % to 0.93 %
{"row_1":{"col_1":"High energy photon beam<br type=\"_moz\">","col_2":"Expanded uncertainty&nbsp; (<em>k</em> = 2) in %"},"row_2":{"col_1":"6 MV","col_2":"0.93"},"row_3":{"col_1":"10 MV","col_2":"0.92"},"row_4":{"col_1":"15 MV","col_2":"0.90"},"row_5":{"col_1":"20 MV","col_2":"0.89"}}
Irradiation in a water phantom, in a field calibrated against a transfer standard
Photons, high energy, Medical type accelerator: 6 MV, 10 MV, 15 MV and 20 MV (TPR from 0.66 to 0.79)
Reference standard : Graphite calorimeter and Monte Carlo calculations to convert from graphite to water
Source of traceability : LNE-LNHB
Approved on 11 January 2025
Institute service identifier : EUR-RAD-LNE-LNHB-1083
CMC ID : EURAMET-RI-FR-00000Q1H-1
 
Ionizing radiation
Absorbed dose/rate , Ionization chamber : 0.01 Gy/s to 0.08 Gy/s
Relative expanded uncertainty 1.1 %
Calibration against a transfer standard in a water phantom
Photons, high energy, Medical type accelerator: 6 MV, 10 MV, 15 MV et 20 MV (TPR from 0.66 to 0.79) and extrapolation to TPR from 0.60 to 0.81 using IAEA TRS 398 
Reference standard : Graphite calorimeter and Monte Carlo calculations to convert from graphite to water
Source of traceability : LNE-LNHB
Approved on 08 January 2025
Institute service identifier : EUR-RAD-LNE-LNHB-1015
CMC ID : EURAMET-RI-FR-00000KZS-2