Database of higher-order reference materials,
measurement methods/procedures and services

logo_BIPM
logo_JCTLM
> You are here : JCTLM-DB > Reference measurement services > List
Increase text size Normal text size Decrease text size

Result of the search: list of reference measurement services

RfB-DGKL, Germany
Phone: +49 (0)511 532 3862 Contact person: Dr. D. Grote-Koska, Prof. Dr. G. Schumann
Fax: +49 (0)511 532 8523 Email: Grote-Koska.Denis@mh-hannover.de
Web: http://www.rfb.bio/cgi/switchLang?lang=en
Analyte glucose
Material or matrix whole blood, blood serum, blood plasma, other, urine
Applicable material or matrix whole blood, blood serum, blood plasma, urine, cerebro-spinal fluid
Quantity Amount-of-substance concentration
Service measurement range 0.5 mmol/l to 25 mmol/l
Determination of higher concentrations after predilution of serum
Expanded uncertainty
(level of confidence 95%)
1.0 %
Interlaboratory comparison results RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at
http://www.dgkl-rfb.de:81/index.shtml
Measurement principle Isotope dilution gas chromatography mass spectrometry (ID/GC/MS)
JCTLM reference measurement method/procedure NIST reference method for serum glucose

1