Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (190)
CMCs Chemistry (59)
PHYSICS >
Temperature (62)
Dimensional metrology (33)
Sound in air (18)
Pressure (11)
Humidity (10)
Vibration (9)
DC voltage, current, and resistance (9)
Mass, mass standards (8)
Frequency (7)
AC voltage, current, and power (6)
Properties of detectors and sources (5)
Impedance up to the MHz range (5)
CHEMICAL MATERIAL >
drinking water, fresh water (7)
cereal and cereal products (6)
aqueous buffer solution (4)
nitrogen (4)
polypropylene, polyethylene and polyvinyl chloride (4)
CHEMICAL ANALYTE >
calcium
cadmium (8)
zinc (6)
copper (6)
lead (5)
pH (4)
GEOGRAPHIC LOCATION
APMP (249)
Thailand (249)
Result of the search
Your query 'NIMT' produced 249 results New search
1  2  3 [Next >>]

 
Thailand, NIMT (National Institute of Metrology (Thailand))
Complete CMCs in Photometry and Radiometry for Thailand (.PDF file)
Irradiance, spectral. Tungsten lamp, 3E-05 W/(m2 nm) to 4E-02 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 6.6
Monochromator
Wavelength range: 290 nm to 300 nm
Bandwidth: 5 nm
Approved on 16 January 2011
Internal NMI service identifier: NIMT/13030-10101
 
Irradiance, spectral. Tungsten lamp, 3E-05 W/(m2 nm) to 4E-02 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 5.6
Monochromator
Wavelength range: 300 nm to 500 nm (excludes 300 nm)
Bandwidth: 5 nm
Approved on 16 January 2011
Internal NMI service identifier: NIMT/13030-10101
 
Irradiance, spectral. Tungsten lamp, 6E-03 W/(m2 nm) to 0.13 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.3
Monochromator
Wavelength range: 500 nm to 600 nm (excludes 500 nm and 600 nm)
Bandwidth: 5 nm
Approved on 16 January 2011
Internal NMI service identifier: NIMT/13030-10101
 
Irradiance, spectral. Tungsten lamp, 2E-02 W/(m2 nm) to 0.2 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.9
Monochromator
Wavelength range: 600 nm to 830 nm (excludes 830 nm)
Bandwidth: 5 nm
Approved on 16 January 2011
Internal NMI service identifier: NIMT/13030-10101
 
Irradiance, spectral. Tungsten lamp, 5E-03 W/(m2 nm) to 0.2 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 6.5
Monochromator
Wavelength range: 830 nm to 900 nm
Bandwidth: 10 nm
Approved on 16 January 2011
Internal NMI service identifier: NIMT/13030-10101
 
Mass. Mass standard, 1 kg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mg: 0.16
Comparison in air
Approved on 20 August 2013
Internal NMI service identifier: NIMT/10010-10101
 
Mass. Mass standard, 1 mg to 100 mg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 0.6 to 1.6
Comparison in air
Approved on 20 August 2013
Internal NMI service identifier: NIMT/10010-10101
 
Mass. Mass standard, 0.1 g to 1 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 1.6 to 3
Comparison in air
Approved on 20 August 2013
Internal NMI service identifier: NIMT/10010-10101
 
Mass. Mass standard, 1 g to 10 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 3 to 6
Comparison in air
Approved on 20 August 2013
Internal NMI service identifier: NIMT/10010-10101
 
Mass. Mass standard, 10 g to 100 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 6 to 16
Comparison in air
Approved on 20 August 2013
Internal NMI service identifier: NIMT/10010-10101
 
Mass. Mass standard, 0.1 kg to 1 kg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 16 to 160
Comparison in air
Approved on 20 August 2013
Internal NMI service identifier: NIMT/10010-10101
 
Mass. Mass standard, 1 kg to 10 kg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mg: 0.16 to 1.7
Comparison in air
Approved on 20 August 2013
Internal NMI service identifier: NIMT/10010-10101 to 10010-10104
 
Mass. Mass standard, 10 kg to 100 kg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mg: 1.7 to 270
Comparison in air
Approved on 20 August 2013
Internal NMI service identifier: NIMT/10010-10104 to 10011-10402
 
Absolute pressure: gas medium. Pressure gauge, 3.5 kPa to 1750 kPa
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in kPa: (3E-05P), P pressure in kPa, or 0.8 Pa whichever is greater
Direct comparison with pressure standard
Approved on 18 April 2006.
Uncertainty values range from 0.8 Pa to 52.5 Pa
Internal NMI service identifier: NIMT/20
 
Absolute pressure: gas medium. Pressure gauge, 1.75 MPa to 7 MPa
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in MPa: (4E-05P), P pressure in Mpa
Direct comparison with pressure standard
Approved on 18 April 2006.
Uncertainty values range from 70 Pa to 280 Pa
Internal NMI service identifier: NIMT/21
 
Absolute pressure: oil medium. Pressure gauge, 0.1 MPa to 7.1 MPa
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in MPa: (3E-05P), P pressure in MPa, or 21 Pa whichever is greater
Direct comparison with pressure standard
Approved on 18 April 2006.
Uncertainty values range from 21 Pa to 213 Pa
Internal NMI service identifier: NIMT/22
 
Absolute pressure: oil medium. Pressure gauge, 7.1 MPa to 70.1 MPa
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in MPa: (4E-05P), P pressure in Mpa
Direct comparison with pressure standard
Approved on 18 April 2006.
Uncertainty values range from 2.1E+02 Pa to 2.8E+03 Pa
Internal NMI service identifier: NIMT/23
 
Absolute pressure: oil medium. Pressure gauge, 70.1 MPa to 140.1 MPa
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in MPa: (5E-05P), P pressure in Mpa
Direct comparison with pressure standard
Approved on 18 April 2006.
Uncertainty values range from 2.8E+03 Pa to 7.0E+03 Pa
Internal NMI service identifier: NIMT/24
 
Gauge pressure: gas medium. Pressure gauge, -1 kPa to 3 kPa
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in kPa: (3E-04P), P pressure in kPa, or 0.3 Pa whichever is greater
Direct comparison with pressure standard
Approved on 18 April 2006.
Uncertainty values range from 0.3 Pa to 0.9 Pa
Internal NMI service identifier: NIMT/25
 
Gauge pressure: gas medium. Pressure balance, 3.5 kPa to 1750 kPa
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in kPa: (3E-05P), P pressure in kPa, or 0.8 Pa whichever is greater
Direct comparison with pressure standard: crossfloat
Approved on 18 April 2006.
Uncertainty values range from 0.8 Pa to 52.5 Pa
Internal NMI service identifier: NIMT/26