Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (208)
CMCs Chemistry (19)
CMCs Ionizing Radiation (14)
PHYSICS >
Temperature (54)
DC voltage, current, and resistance (30)
Impedance up to the MHz range (25)
Radio frequency measurements (21)
Dimensional metrology (21)
AC voltage, current, and power (13)
Pressure (9)
Mass, mass standards (7)
Humidity (7)
Frequency (6)
Fluid flow (5)
Time interval (5)
IONIZING RADIATION
Activity per unit mass (5)
Personal dose equivalent rate penetrating (at 10 mm depth) (5)
Air kerma rate (4)
CHEMICAL MATERIAL >
herb and plant materials (5)
soil (4)
seafood (3)
cosmetic material (2)
yellow gold jewellery alloys (1)
CHEMICAL ANALYTE >
arsenic (4)
zinc (2)
cadmium (2)
iron (2)
Genomic DNA. Copy number ratio of specified intact sequence fragments of 50 to 150 nucleotides length in a single genomic DNA extract. (2)
GEOGRAPHIC LOCATION
EURAMET (241)
Slovenia (241)
Result of the search
Your query 'Slovenia' produced 241 results New search
1  2  3 [Next >>]

 
Slovenia, MIRS/SIQ/Metrology (MIRS/Slovenian Institute of Quality and Metrology/Metrology)
Complete CMCs in Time and Frequency for Slovenia (.PDF file)
Time interval. Time difference source, 1E-06 s to 5 s
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in ns: (0.1 + 1E-09t), t measured time interval in seconds
Time interval counter
Rise time: < 1 ns
Approved on 21 April 2005
 
Time interval. Rise/fall time source, 50 ps to 10000 ps
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in ps: (30 + 0.05tr), tr measured rise time in seconds
Oscilloscope
Repetition frequency: up to 100 kHz
Approved on 21 April 2005
 
Frequency. Local frequency standard, 0.1 MHz
Relative expanded uncertainty (k = 2, level of confidence 95 %): 2E-13
Phase measurement
Averaging time: 24 h
Approved on 21 April 2005
 
Frequency. Local frequency standard, 1 MHz
Relative expanded uncertainty (k = 2, level of confidence 95 %): 2E-13
Phase measurement
Averaging time: 24 h
Approved on 21 April 2005
 
Frequency. Local frequency standard, 5 MHz
Relative expanded uncertainty (k = 2, level of confidence 95 %): 2E-13
Phase measurement
Averaging time: 24 h
Approved on 21 April 2005
 
Frequency. Local frequency standard, 10 MHz
Relative expanded uncertainty (k = 2, level of confidence 95 %): 2E-13
Phase measurement
Averaging time: 24 h
Approved on 21 April 2005
 
Frequency. General frequency source, 0.001 Hz to 1.00E+05 Hz
Relative expanded uncertainty (k = 2, level of confidence 95 %) in Hz/Hz: (3E-06/f + 1.7E-11), f measured frequency in Hz
Frequency counter
Averaging time: 1000 s
Approved on 21 April 2005
 
Frequency. General frequency source, 0.0001 GHz to 26.5 GHz
Relative expanded uncertainty (k = 2, level of confidence 95 %) in Hz/Hz: 1E-12
Frequency counter
Averaging time: 1000 s
Approved on 21 April 2005
 
Time interval. Delay source, 10 ps to 100 ps
Relative expanded uncertainty (k = 2, level of confidence 95 %) in ps/ps: 0.12 to 0.025
Oscilloscope
Source impedance: 50 ohm
Approved on 21 April 2005
 
Time interval. Delay source, 0.1 ns to 1 ns
Relative expanded uncertainty (k = 2, level of confidence 95 %) in ns/ns: 0.026 to 0.008
Oscilloscope
Source impedance: 50 ohm
Approved on 21 April 2005
 
Time interval. Delay source, 1 ns to 1000 ns
Relative expanded uncertainty (k = 2, level of confidence 95 %) in ns/ns: 0.008 to 0.0001
Oscilloscope
Source impedance: 50 ohm
Approved on 21 April 2005
 
Slovenia, MIRS/UL-FE/LMK (MIRS/University of Ljubljana, Faculty of Electrical Engineering/Laboratory of Metrology and Quality)
Complete CMCs in Thermometry for Slovenia (.PDF file)
Temperature. Liquid-in-glass thermometer, 70 °C to 125 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.008
Comparison in Oil bath
Ambient temperature: 23 °C
Approved on 03 November 2009
Internal NMI service identifier: MIRS/UL-FE/LMK/2.4. LIGT
 
Temperature. Liquid-in-glass thermometer, 125 °C to 150 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.04
Comparison in Oil bath
Ambient temperature: 23 °C
Approved on 03 November 2009
Internal NMI service identifier: MIRS/UL-FE/LMK/2.4. LIGT
 
Temperature. Liquid-in-glass thermometer, 150 °C to 250 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.04
Comparison in Oil bath
Ambient temperature: 23 °C
Approved on 03 November 2009
Internal NMI service identifier: MIRS/UL-FE/LMK/2.4. LIGT
 
Temperature. Liquid-in-glass thermometer, 250 °C to 300 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 0.08
Comparison in Oil bath
Ambient temperature: 23 °C
Approved on 03 November 2009
Internal NMI service identifier: MIRS/UL-FE/LMK/2.4. LIGT
 
Temperature. Water triple point cell, 0.01 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.1
Direct comparison
Temperature-controlled bath
Approved on 20 January 2010
Internal NMI service identifier: MIRS/UL-FE/LMK/1.1.1 W
 
Temperature. Long-stem SPRT, 0.01 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.15
Calibration at water triple point
Ambient temperature: 23 °C
Approved on 20 January 2010
Internal NMI service identifier: MIRS/UL-FE/LMK/1.3.2 W
 
Temperature. Melting point of Gallium, 29.7646 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.4
Direct comparison
Temperature-controlled bath
Approved on 20 May 2010
Internal NMI service identifier: MIRS/UL-FE/LMK/1.1.1 Ga
 
Temperature. Freezing point of Tin, 231.928 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 1
Direct comparison
Temperature-controlled bath
Approved on 20 May 2010
Internal NMI service identifier: MIRS/UL-FE/LMK/1.1.1 Sn
 
Temperature. Freezing point of Aluminium, 660.323 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 2.5
Direct comparison
Temperature-controlled bath
Approved on 20 May 2010
Internal NMI service identifier: MIRS/UL-FE/LMK/1.1.1 Al