Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (112)
CMCs Ionizing Radiation (22)
PHYSICS
Temperature (44)
DC voltage, current, and resistance (15)
Fluid flow (13)
AC voltage, current, and power (13)
Mass, mass standards (9)
Dimensional metrology (9)
Laser frequencies (4)
Properties of detectors and sources (3)
Humidity (2)
IONIZING RADIATION
Air kerma rate (8)
Ambient dose equivalent rate (3)
Directional dose equivalent rate (3)
Personal dose equivalent rate, penetrating (3)
Absorbed dose rate to water (1)
Personal dose equivalent rate, superficial (1)
Absorbed dose to water (1)
Air kerma (1)
Personal dose equivalent, rate, superficial (1)
GEOGRAPHIC LOCATION
EURAMET (134)
Norway (134)
Result of the search
Your query 'Norway' produced 134 results New search
1  2  3 [Next >>]

 
Norway, JV (Justervesenet)
Complete CMCs in Photometry and Radiometry for Norway (.PDF file)
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 4.8
Double grating monochromator, cut -in filters
Wavelength range: 260 nm to 400 nm
Bandwith: 4 nm
Power: 0.4 µW to 2.1 µW
Approved on 29 September 2001
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.63
Double grating monochromator, cut -in filters
Wavelength range: 400 nm to 930 nm
Bandwith: 4 nm
Power: 0.3 µW to 1.5 µW
Approved on 29 September 2001
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.4
Double grating monochromator, cut -in filters
Wavelength range: 930 nm to 1600 nm
Bandwith: 4 nm
Power: 1 µW to 6 µW
Approved on 29 September 2001
 
Mass. Mass standard, 1 mg to 100 mg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 0.6 to 1.5
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known.
Approved on 02 December 2003
 
Mass. Mass standard, 0.1 g to 1 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 1.5 to 2.1
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known.
Approved on 02 December 2003
 
Mass. Mass standard, 1 g to 10 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 2.1 to 9
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known.
Approved on 02 December 2003
 
Mass. Mass standard, 10 g to 100 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 9 to 14
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known.
Approved on 02 December 2003
 
Mass. Mass standard, 0.1 kg to 1 kg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 14 to 70
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known.
Approved on 02 December 2003
 
Mass. Mass standard, 1 kg to 10 kg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mg: 0.07 to 1.3
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known.
Approved on 02 December 2003
 
Mass. Mass standard, 10 kg to 50 kg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mg: 1.3 to 22
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known.
Approved on 02 December 2003
 
Mass. Mass standard, 50 kg to 100 kg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mg: 22 to 170
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known.
Approved on 02 December 2003
 
Mass. Mass standard, 100 kg to 500 kg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in g: 0.17 to 1.1
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known.
Approved on 02 December 2003
 
Mass flowrate. Flow meter, 0.1 kg/s to 25 kg/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.20
Coriolis, pulse or other electrical output
Liquidwater
Temperature: 5 °C to 65 °C
Pipe size: DN 10 - 150
Approved on 02 December 2003
Internal NMI service identifier: NO1
 
Mass flowrate. Flow meter, 0.1 kg/s to 25 kg/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.20
Coriolis, visual reading
Liquidwater
Temperature: 5 °C to 65 °C
Pipe size: DN 10 - 150
Approved on 02 December 2003
Internal NMI service identifier: NO2
 
Volume flowrate. Flow meter, 0.1 L/s to 25 L/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.22
Positive displacement, turbine, electromagnetic, ultrasonic, pulse or other electrical output
Liquidwater
Temperature: 5 °C to 65 °C
Pipe size: DN 10 - 150
Approved on 02 December 2003
Internal NMI service identifier: NO3
 
Volume flowrate. Flow meter, 0.1 L/s to 25 L/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.22
Positive displacement, turbine, electromagnetic, ultrasonic, visual reading
Liquidwater
Temperature: 5 °C to 65 °C
Pipe size: DN 10 - 150
Approved on 02 December 2003
Internal NMI service identifier: NO4
 
Mass. Mass meter, 1 kg to 5000 kg
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.10
Coriolis
Liquidwater
Temperature: 5 °C to 65 °C
Pipe size: DN 10 - 150
Approved on 02 December 2003
Internal NMI service identifier: NO5
 
Volume. Volume meter, 1 L to 5000 L
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.12
Positive displacement, turbine, electromagnetic, ultrasonic
Liquidwater
Temperature: 5 °C to 65 °C
Pipe size: DN 10 - 150
Approved on 02 December 2003
Internal NMI service identifier: NO6
 
Volume flowrate. Water-flow meter, 0.01 L/s to 4 L/s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.20
Positive displacement, turbine, electromagnetic, ultrasonic
Liquidwater
Temperature: 5 °C to 15 °C
Pipe size: DN 5 - 50
Approved on 02 December 2003
Internal NMI service identifier: NO7
 
Volume. Water meter, 10 L to 500 L
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.10
Positive displacement, turbine, electromagnetic, ultrasonic
Liquidwater
Temperature: 5 °C to 15 °C
Pipe size: DN 5 - 50
Approved on 02 December 2003
Internal NMI service identifier: NO8