Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (103)
CMCs Ionizing Radiation (15)
PHYSICS >
Temperature (14)
Vibration (13)
DC voltage, current, and resistance (12)
Sound in air (8)
Pressure (8)
Mass, mass standards (7)
AC voltage, current, and power (7)
Photometry (4)
Radio frequency measurements (4)
Impedance up to the MHz range (4)
High voltage and current (4)
Laser frequencies (4)
IONIZING RADIATION
Air kerma rate (6)
Ambient dose equivalent rate (4)
Absorbed dose rate to tissue (2)
Personal dose equivalent rate, penetrating (2)
Absorbed dose rate to water (1)
GEOGRAPHIC LOCATION
APMP (118)
Malaysia (118)
Result of the search
Your query 'Malaysia' produced 118 results New search
1  2  3 [Next >>]

 
Malaysia, NMIM (National Metrology Institute of Malaysia)
Complete CMCs in Time and Frequency for Malaysia (.PDF file)
Frequency. General frequency source, 1E-06 MHz to 225 MHz
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in Hz: 7.0E-03
Direct frequency measurements
Measurement time: 10 s
Number of measurements: 100
Excluded DUT's effect
Approved on 22 February 2008
Internal NMI service identifier: NMIM/CP-446-205
 
Time interval. Time difference source, 0.01 s to 100000 s
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in s: 2.0E-05
Time interval measurement
Number of measurements: 10
Included DUT's effect
Approved on 22 February 2008
Internal NMI service identifier: NMIM/CP-446-208
 
Frequency. General frequency source, 0.2 MHz to 225 MHz
Relative expanded uncertainty (k = 2, level of confidence 95 %) in Hz/Hz: 3.0E-12
Direct frequency measurement
Averaging time: 14400 s
Gate time: 1 s
Excluded DUT's effect
Approved on 22 February 2008
Internal NMI service identifier: NMIM/CP-446-203
 
Frequency. Local frequency standard, 1 MHz to 10 MHz
Relative expanded uncertainty (k = 2, level of confidence 95 %) in s/s: 4.0E-13
Phase comparison
Measurement time: 259200 s
Excluded DUT's effect
Approved on 22 February 2008
Internal NMI service identifier: NMIM/CP-446-207
 
Mass. Mass standard, 0.001 g to 0.5 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 0.6 to 1.0
Comparison in air
Internal NMI service identifier: NMIM/1
 
Mass. Mass standard, 1 g to 5 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 1.2 to 5.0
Comparison in air
Internal NMI service identifier: NMIM/2
 
Mass. Mass standard, 10 g to 100 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 7.0 to 20
Comparison in air
Internal NMI service identifier: NMIM/3
 
Mass. Mass standard, 200 g to 500 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 35 to 75
Comparison in air
Internal NMI service identifier: NMIM/4
 
Mass. Mass standard, 1000 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 150
Comparison in air
Internal NMI service identifier: NMIM/5
 
Mass. Mass standard, 2000 g to 10000 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 330 to 1600
Comparison in air
Internal NMI service identifier: NMIM/6
 
Mass. Mass standard, 20000 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 6000
Comparison in air
Internal NMI service identifier: NMIM/7
 
Absolute pressure: gas medium. Pressure measuring device, pressure balance, 10 kPa to 175 kPa
Relative expanded uncertainty (k = 2, level of confidence 95 %): 2.5E-05
Direct comparison with standard, cross-float
Internal NMI service identifier: NMIM/8
 
Absolute pressure: gas medium. Pressure measuring device, pressure balance, 175 kPa to 700 kPa
Relative expanded uncertainty (k = 2, level of confidence 95 %): 3.5E-05
Direct comparison with standard, cross-float
Internal NMI service identifier: NMIM/9
 
Absolute pressure: gas medium. Pressure measuring device, pressure balance, 0.7 MPa to 7 MPa
Relative expanded uncertainty (k = 2, level of confidence 95 %): 4.5E-05
Direct comparison with standard, cross-float
Internal NMI service identifier: NMIM/10
 
Gauge pressure: gas medium. Pressure measuring device, pressure balance, 10 kPa to 175 kPa
Relative expanded uncertainty (k = 2, level of confidence 95 %): 2.5E-05
Direct comparison with standard, cross-float
Internal NMI service identifier: NMIM/11
 
Gauge pressure: gas medium. Pressure measuring device, pressure balance, 175 kPa to 700 kPa
Relative expanded uncertainty (k = 2, level of confidence 95 %): 3.5E-05
Direct comparison with standard, cross-float
Internal NMI service identifier: NMIM/12
 
Gauge pressure: gas medium. Pressure measuring device, pressure balance, 0.7 MPa to 7 MPa
Relative expanded uncertainty (k = 2, level of confidence 95 %): 4.5E-05
Direct comparison with standard, cross-float
Internal NMI service identifier: NMIM/13
 
Gauge pressure: liquid medium. Pressure measuring device, pressure balance, 0.345 MPa to 50 MPa
Relative expanded uncertainty (k = 2, level of confidence 95 %): 5.0E-05
Direct comparison with standard, cross-float, oil
Internal NMI service identifier: NMIM/14
 
Gauge pressure: liquid medium. Pressure measuring device, pressure balance, 50 MPa to 138 MPa
Relative expanded uncertainty (k = 2, level of confidence 95 %): 6.0E-05
Direct comparison with standard, cross-float, oil
Internal NMI service identifier: NMIM/15
 
DC voltage sources: single values. Zener reference, standard cell, 1 V to 10 V
Absolute expanded uncertainty (k = 2, level of confidence 95%) in µV: 0.2 to 0.4
Voltage difference
Temperature: 23 °C
Approved on 10 November 2014
Internal NMI service identifier: NMIM/1.1.1
Uncertainty table