Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (96)
PHYSICS
Temperature (30)
DC voltage, current, and resistance (15)
Dimensional metrology (13)
Impedance up to the MHz range (8)
Mass, mass standards (7)
Frequency (7)
AC voltage, current, and power (5)
Humidity (5)
Sound in air (4)
High voltage and current (2)
GEOGRAPHIC LOCATION
EURAMET (96)
Ireland (96)
Result of the search
Your query 'Ireland' produced 96 results New search
1  2  3 [Next >>]

 
Ireland, NSAI NML (NSAI National Metrology Laboratory)
Complete CMCs in Mass and related quantities for Ireland (.PDF file)
Mass. Mass standards, 1 mg to 100 mg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 1.5 to 3
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known
 
Mass. Mass standards, 0.1 g to 1 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 3 to 8
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known
 
Mass. Mass standards, 1 g to 10 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 8
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known
 
Mass. Mass standards, 10 g to 100 g
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 8 to 16
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known
 
Mass. Mass standards, 0.1 kg to 1 kg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µg: 16 to 150
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known
 
Mass. Mass standards, 1 kg to 10 kg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mg: 0.15 to 3.4
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known
 
Mass. Mass standards, 10 kg to 20 kg
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mg: 3.4 to 6.8
Comparison in air
Uncertainty scales with measurand level
The volume of the mass standards is known
 
Frequency. General frequency source, 10 Hz to 99 Hz
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µHz: 18 to 6
Frequency counter
Averaging time: 1000 s
Approved on 02 May 2006
Internal NMI service identifier: NSAI NML/4
 
Frequency. General frequency source, 100 Hz to 999 Hz
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µHz: 6 to 2
Frequency counter
Averaging time: 1000 s
Approved on 02 May 2006
Internal NMI service identifier: NSAI NML/5
 
Frequency. General frequency source, 1000 Hz to 9999 Hz
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µHz: 2 to 1
Frequency counter
Averaging time: 1000 s
Approved on 02 May 2006
Internal NMI service identifier: NSAI NML/6
 
Frequency. Local frequency standard, 10 MHz
Relative expanded uncertainty (k = 2, level of confidence 95 %) in Hz/Hz: 5E-13
Phase difference measurement
Averaging time: 1 day
Comparison made with NPL standard using CV method
Approved on 02 May 2006
Internal NMI service identifier: NSAI NML/1
 
Frequency. Local frequency standard, 5 MHz
Relative expanded uncertainty (k = 2, level of confidence 95 %) in Hz/Hz: 5E-13
Phase difference measurement
Averaging time: 1 day
Comparison made with NPL standard using CV method
Approved on 02 May 2006
Internal NMI service identifier: NSAI NML/2
 
Frequency. General frequency source, 1 MHz
Relative expanded uncertainty (k = 2, level of confidence 95 %) in Hz/Hz: 1E-11
Frequency counter
Averaging time: 1 day
Approved on 02 May 2006
Internal NMI service identifier: NSAI NML/3
 
Frequency. General frequency source, 0.01 MHz to 1000 MHz
Relative expanded uncertainty (k = 2, level of confidence 95 %) in Hz/Hz: 1E-10 to 3E-11
Frequency counter
Averaging time: 1000 s
Approved on 02 May 2006
Internal NMI service identifier: NSAI NML/7
 
Sound pressure level. Sound calibrator, single frequency, 1 kHz, 84 dB (reference: 20 µPa)
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in dB: 0.26
Calibrated measurement microphone
Microphone type: WS2P (IEC 61094-4)
Approved on 31 March 2011
Internal NMI service identifier: NMLI_A001
 
Sound pressure level. Sound calibrator, single frequency, 1 kHz, 94 dB (reference: 20 µPa) to 114 dB (reference: 20 µPa)
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in dB: 0.13
Calibrated measurement microphone
Microphone type: WS2P (IEC 61094-4)
Approved on 31 March 2011
Internal NMI service identifier: NMLI_A002
 
Sound pressure level. Pistonphone, single frequency, 250 Hz, 124 dB (reference: 20 µPa)
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in dB: 0.12
Calibrated measurement microphone
Microphone type: WS2P (IEC 61094-4)
Approved on 31 March 2011
Internal NMI service identifier: NMLI_A003
 
Sound pressure response level. Sound level meter, 94 dB (reference: 20 µPa)
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in dB: 0.30
IEC 61672-1, IEC 60651
Frequency: 125 Hz and 1 kHz
Approved on 31 March 2011
Internal NMI service identifier: NMLI_A004
 
CMM artifacts. 1-D artifacts,
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µm: (2.3 + 0.003L), L in mm
Zeiss UMC850 CMM
Maximum length L: 1200 mm
Internal NMI service identifier: NSAI NML/4
 
CMM artifacts. 3-D artifacts,
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in µm: (2.8 + 0.004L), L length in mm
Zeiss UMC850 CMM
Maximum volume: (600x800x1200) mm3