Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs Ionizing Radiation (76)
CMCs General Physics (37)
PHYSICS
DC voltage, current, and resistance (17)
Fluid flow (12)
Temperature (8)
IONIZING RADIATION
Activity per unit mass (51)
Activity (12)
Air kerma (6)
Ambient dose equivalent (2)
Personal dose equivalent penetrating (2)
Absorbed dose to water (1)
Reference air kerma rate (1)
Personal dose equivalent superficial (1)
GEOGRAPHIC LOCATION
COOMET (113)
Cuba (113)
Result of the search
Your query 'Cuba' produced 113 results New search
1  2  3 [Next >>]

 
Cuba, INIMET (Instituto Nacional de Investigaciones en Metrologia)
Complete CMCs in Mass and related quantities for Cuba (.PDF file)
Liquid volume flow. Volumetric flowmeter, 100 L/min to 500 L/min
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.2
Volume comparison (pulsed, electrical, digital and optical outputs)
Fluid: gas-oil, kerosene, gasoline, heavy oil, diesel
Temperature: ambient
Pressure: up to 0.8 MPa
Pipe diameter: 50 mm
Approved on 08 November 2005
Internal NMI service identifier: CU1
 
Liquid volume flow. Volumetric flowmeter, 50 L/min to 5E+03 L/min
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.1
Volume comparison (pulsed, electrical, digital and optical outputs)
Fluid: gas-oil, kerosene
Temperature: ambient
Pressure: up to 0.8 MPa
Pipe: 2 meter runs 50 mm and 100 mm
Number of reference volumes: 4
Approved on 08 November 2005
Internal NMI service identifier: CU2
 
Liquid volume flow. Volumetric flowmeter, 60 L/min to 1.3E+03 L/min
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.2
Volume comparison (pulsed, electrical, digital and optical outputs)
Fluid: water
Temperature: ambient
Pressure: up to 0.8 MPa
Pipe diameter: 75 mm
Master meter
Approved on 08 November 2005
Internal NMI service identifier: CU3
 
Liquid volume flow. Volumetric flowmeter, 50 L/min to 2.5E+03 L/min
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.15
Volume comparison (pulsed, electrical, digital and optical outputs)
Fluid: water
Temperature: ambient
Pressure: up to 0.8 MPa
Pipe diameter: 75 mm
Tank prover
Approved on 08 November 2005
Internal NMI service identifier: CU4
 
Static volume. Glassware, pipettes, automatic pipettes, etc., 2E-03 L to 50 L
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.02 to 0.008
Gravimetric
Liquid: water
Temperature: (20 ± 2)°C
Pressure: atmospheric
Approved on 08 November 2005
Internal NMI service identifier: CU5
 
Static volume. Micropipettes, 2E-02 mL to 5 mL
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1 to 0.09
Gravimetric
Liquid: water
Temperature: (20 ± 2)°C
Pressure: atmospheric
Approved on 08 November 2005
Internal NMI service identifier: CU6
 
Static volume. Metal or glass measure artefacts, 1 L to 5E+04 L
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.008 to 0.04
Gravimetric and volumetric
Liquid: water
Temperature: (20 ± 5)°C
Pressure: atmospheric
Approved on 08 November 2005
Internal NMI service identifier: CU7
 
Static volume. Storage tank (cylindrical standing), 10 m3 to 5E+04 m3
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.1 to 0.05
Geometric and volumetric
Liquid: any products at ambient temperature
Reference temperature: 15 °C or 20 °C
Pressure: atmospheric
Approved on 08 November 2005
Internal NMI service identifier: CU8
 
Static volume. Storage tank underground, road and rail tanks, 1 m3 to 1E+02 m3
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.5 to 1.0
Volumetric
Liquid: water or diesel at ambient temperature
Reference temperature: 15 °C or 20 °C
Pressure: atmospheric
Approved on 08 November 2005
Internal NMI service identifier: CU9
 
Gas volume flow. Gas flow meters, 5E-02 m3/h to 5 m3/h
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.3
Sonic nozzles, turbine, laminar, P.D, rotameters
Gas: dry air
Temperature: ambient
Pressure: up to 0.8 MPa
Approved on 08 November 2005
Internal NMI service identifier: CU10
 
Gas volume flow. Gas flow meters, 5 m3/h to 50 m3/h
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.2
Sonic nozzles, turbine, laminar, P.D, rotameters
Gas: dry air
Temperature: ambient
Pressure: up to 0.8 MPa
Approved on 08 November 2005
Internal NMI service identifier: CU10
 
Liquid volume flow. Volumetric flowmeter, 6.00E-01 m3/h to 6.6 m3/h
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.1
Volume comparison (pulsed, electrical, digital and optical outputs)
Fluid: gas-oil, kerosene, gasoline, heavy oil, diesel
Temperature: ambient
Pressure: up to 0.8 MPa
Pipe diameter: 150 mm
Approved on 08 November 2005
Internal NMI service identifier: CU11
 
Temperature. Water triple point cell, 0.01 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.15
Direct comparison
Thermostat: ice bath
Approved on 06 September 2013
Internal NMI service identifier: INIMET/1
 
Temperature. Long-stem Standard Platinum Resistance Thermometer, 0.01 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.2
Water triple fixed point
Thermostat: ice bath
Approved on 06 September 2013
Internal NMI service identifier: INIMET/2
 
Temperature. Industrial Platinum Resistance Thermometer, 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 10
Ice melting point
Thermostat: ice bath
Approved on 06 September 2013
Internal NMI service identifier: INIMET/3
 
Temperature. Thermistors, 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 10
Ice melting point
Thermostat: ice bath
Approved on 06 September 2013
Internal NMI service identifier: INIMET/4
 
Temperature. Mercury-in-glass, total immersion, 0.1 °C graduation, 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.05
Ice melting point
Thermostat: ice bath
Approved on 06 September 2013
Internal NMI service identifier: INIMET/5
 
Temperature. Mercury-in-glass, total immersion, 0.2 °C graduation, 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.07
Ice melting point
Thermostat: ice bath
Approved on 06 September 2013
Internal NMI service identifier: INIMET/6
 
Temperature. Mercury-in-glass, partial immersion, 0.1 °C graduation, 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.05
Ice melting point
Thermostat: ice bath
Approved on 06 September 2013
Internal NMI service identifier: INIMET/7
 
Temperature. Mercury-in-glass, partial immersion, 0.2 °C graduation, 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in °C: 0.07
Ice melting point
Thermostat: ice bath
Approved on 06 September 2013
Internal NMI service identifier: INIMET/8