1  2  3

 
Belgium, SMD (Service de la Métrologie Belge)
Complete CMCs in Thermometry for Belgium (.PDF file)
Temperature. Water triple point cell, 0.01 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.25
Direct comparison
Temperature-controlled bath
Approved on 20 January 2010
Internal NMI service identifier: SMD/1.3
 
Temperature. Long stem SPRT, 0.01 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.3
Measurement at water triple point
Temperature-controlled bath
Approved on 20 January 2010
Internal NMI service identifier: SMD/2.3
 
Temperature. Argon cell, -189.3442 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.8
Direct comparison
Adiabatic calorimeter
Approved on 20 May 2010
Internal NMI service identifier: SMD/1.1
 
Temperature. Mercury cell, -38.8344 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.6
Direct comparison
Temperature controlled bath
Approved on 20 May 2010
Internal NMI service identifier: SMD/1.2
 
Temperature. Gallium cell, 29.7646 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.6
Direct comparison
Air flow furnace
Approved on 20 May 2010
Internal NMI service identifier: SMD/1.4
 
Temperature. Indium cell, 156.5985 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 1.7
Direct comparison
Three zone furnace
Approved on 20 May 2010
Internal NMI service identifier: SMD/1.5
 
Temperature. Zinc cell, 419.527 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 2.2
Direct comparison
Three zone furnace
Approved on 20 May 2010
Internal NMI service identifier: SMD/1.7
 
Temperature. Long stem SPRT, -189.3442 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 1
Measurement at Argon point
Adiabatic calorimeter
Approved on 20 May 2010
Internal NMI service identifier: SMD/2.1
 
Temperature. Long stem SPRT, -38.8344 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.7
Measurement at mercury point
Temperature controlled bath
Approved on 20 May 2010
Internal NMI service identifier: SMD/2.2
 
Temperature. Long stem SPRT, 29.7646 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.7
Measurement at gallium point
Air flow furnace
Approved on 20 May 2010
Internal NMI service identifier: SMD/2.4
 
Temperature. Long stem SPRT, 156.5985 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 1.8
Measurement at Indium point
Three zone furnace
Approved on 20 May 2010
Internal NMI service identifier: SMD/2.5
 
Temperature. Long stem SPRT, 419.527 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 2.6
Measurement at zinc melting point
Three zone furnace
Approved on 20 May 2010
Internal NMI service identifier: SMD/2.7
 
Temperature. Tin cell, 231.928 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 1.8
Direct comparison
Three zone furnace
Approved on 17 January 2013
Internal NMI service identifier: SMD/1.6
 
Temperature. Long stem SPRT, 231.928 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 1.9
Measurement at tin point
Three zone furnace
Approved on 17 January 2013
Internal NMI service identifier: SMD/2.6
 
Temperature. Long stem SPRT, -189 °C to 0 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 1.4
Measurement at fixed points
Ambient temperature: (22 ± 1) °C
Three zone furnace
Approved on 17 January 2013
Internal NMI service identifier: SMD/3.1
 
Temperature. Long stem SPRT, -38 °C to 29 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.8
Measurement at fixed points
Ambient temperature: (22 ± 1) °C
Three zone furnace
Approved on 17 January 2013
Internal NMI service identifier: SMD/3.2
 
Temperature. Long stem SPRT, 0 °C to 29 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.5 to 0.8
Measurement at fixed points
Ambient temperature: (22 ± 1) °C
Three zone furnace
Approved on 17 January 2013
Internal NMI service identifier: SMD/3.3
 
Temperature. Long stem SPRT, 0 °C to 156 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 0.7 to 1.8
Measurement at fixed points
Ambient temperature: (22 ± 1) °C
Three zone furnace
Approved on 17 January 2013
Internal NMI service identifier: SMD/3.4
 
Temperature. Long stem SPRT, 0 °C to 231 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 2.1
Measurement at fixed points
Ambient temperature: (22 ± 1) °C
Three zone furnace
Approved on 17 January 2013
Internal NMI service identifier: SMD/3.5
 
Temperature. Long stem SPRT, 0 °C to 419 °C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in mK: 1.6 to 2.9
Measurement at fixed points
Ambient temperature: (22 ± 1) °C
Three zone furnace
Approved on 17 January 2013
Internal NMI service identifier: SMD/3.6