Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (202)
CMCs Ionizing Radiation (48)
CMCs Chemistry (43)
PHYSICS >
DC voltage, current, and resistance (43)
Mass, mass standards (29)
Impedance up to the MHz range (17)
Temperature (17)
Density (16)
Pressure (13)
AC voltage, current, and power (12)
Force (11)
Dimensional metrology (11)
Frequency (7)
High voltage and current (6)
Sound in air (5)
IONIZING RADIATION
Activity per unit mass (25)
Activity (23)
CHEMICAL MATERIAL >
methanol, multi-component methanol solutions (6)
sediment (6)
soil (6)
whole fat soybean powder (4)
acid solution (3)
CHEMICAL ANALYTE >
zinc (6)
mercury (6)
chromium (5)
cadmium (4)
lead (4)
GEOGRAPHIC LOCATION
SIM (293)
Argentina (293)
Result of the search
Your query 'Argentina' produced 293 results New search
1  2  3 [Next >>]

 
Argentina, INTI (Instituto Nacional de Tecnologia Industrial)
Complete CMCs in Thermometry for Argentina (.PDF file)
Temperature. PRT, -25 C to 10 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: 0.1
Comparison with SPRT
Bath: temperature controlled ethylene glycol and water bath
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC08 or PEC13
 
Temperature. PRT, 10 C to 80 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: 0.02
Comparison with SPRT
Bath: temperature controlled water bath
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC08 or PEC13
 
Temperature. PRT, 80 C to 200 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: 0.03
Comparison with SPRT
Bath: temperature controlled siliconed oil bath
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC08 or PEC13
 
Temperature. PRT, 200 C to 400 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: 0.1
Comparison with SPRT
Bath: temperature controlled alumina powder-in air fluid bath
Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the Calibration Report
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC08 or PEC13
 
Temperature. Mercury-in-glass thermometer 0.1 C graduation, -25 C to 10 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: 0.1
Total immersion, comparison with SPRT
Bath: temperature controlled ethylene glycol and water bath
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC10
 
Temperature. Mercury-in-glass thermometer 0.1 C graduation, 10 C to 80 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: 0.04
Total immersion, comparison with SPRT
Bath: temperature controlled water bath
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC10
 
Temperature. Mercury-in-glass thermometer 0.1 C graduation, 80 C to 200 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: 0.05
Total immersion, comparison with SPRT
Bath: temperature controlled siliconed oil bath
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC10
 
Temperature. Mercury-in-glass thermometer 0.2 C graduation, -25 C to 10 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: 0.12
Total immersion, comparison with SPRT
Bath: temperature controlled ethylene glycol and water bath
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC10
 
Temperature. Mercury-in-glass thermometer 0.2 C graduation, 10 C to 80 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: 0.07
Total immersion, comparison with SPRT
Bath: temperature controlled water bath
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC10
 
Temperature. Mercury-in-glass thermometer 0.2 C graduation, 80 C to 200 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: 0.07
Total immersion, comparison with SPRT
Bath: temperature controlled siliconed oil bath
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC10
 
Temperature. Mercury-in-glass thermometer 0.5 C graduation, -25 C to 10 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: 0.2
Total immersion, comparison with SPRT
Bath: temperature controlled ethylene glycol and water bath
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC10
 
Temperature. Mercury-in-glass thermometer 0.5 C graduation, 10 C to 80 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: 0.2
Total immersion, comparison with SPRT
Bath: temperature controlled water bath
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC10
 
Temperature. Mercury-in-glass thermometer 0.5 C graduation, 80 C to 200 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: 0.2
Total immersion, comparison with SPRT
Bath: temperature controlled siliconed oil bath
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC10
 
Temperature. type E thermocouple, 0 C to 900 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: [0.5 + 0.0009 t /(C)]
Comparison with thermocouples
Isothermal block in temperature controlled furnace
Pre-determined value of inhomogeneity included in the CMC entry
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC09
 
Temperature. type J thermocouple, 0 C to 700 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: [0.5 + 0.0009 t /(C)]
Comparison with thermocouples
Isothermal block in temperature controlled furnace
Pre-determined value of inhomogeneity included in the CMC entry
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC09
 
Temperature. type K or N thermocouple, 0 C to 1100 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: [0.5 + 0.0009 t /(C)]
Comparison with thermocouples
Isothermal block in temperature controlled furnace
Pre-determined value of inhomogeneity included in the CMC entry
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC09
 
Temperature. type T thermocouple, 0 C to 400 C
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in C: [0.5 + 0.0009 t /(C)]
Comparison with thermocouples
Isothermal block in temperature controlled furnace
Pre-determined value of inhomogeneity included in the CMC entry
Approved on 12 July 2007
Internal NMI service identifier: INTI/PEC09
 
Pressure sensitivity level. LS1P microphone. Measurand unit, dB (reference: 1 V/Pa)
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.05
Pressure sensitivity level by reciprocity IEC 61094-2
Frequency: 63 Hz to 2 kHz
Internal NMI service identifier: INTI/02.01E.03.002
 
Pressure sensitivity level. LS1P microphone. Measurand unit, dB (reference: 1 V/Pa)
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.1
Pressure sensitivity level by reciprocity IEC 61094-3
Frequency: 4 kHz to 8 kHz
Internal NMI service identifier: INTI/02.01E.03.002
 
Pressure sensitivity level. LS1P microphone. Measurand unit, dB (reference: 1 V/Pa)
Absolute expanded uncertainty (k = 2, level of confidence 95%) in dB: 0.2
Pressure sensitivity level by comparison
Frequency: 63 Hz to 8 kHz
At reference ambient conditions
Internal NMI service identifier: INTI/02.01E.03.001