Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs Chemistry (459)
CMCs General Physics (360)
CMCs Ionizing Radiation (187)
PHYSICS >
Impedance up to the MHz range
Materials
Photometry
Properties of materials
Temperature (56)
Properties of detectors and sources (55)
Radio frequency measurements (50)
Dimensional metrology (45)
AC voltage, current, and power (20)
Electric and magnetic fields (18)
Fluid flow (14)
Sound in water (11)
DC voltage, current, and resistance (11)
Pressure (11)
Laser frequencies (10)
High voltage and current (9)
IONIZING RADIATION >
Activity (76)
Activity per unit mass (31)
Absorbed dose to water (15)
Fluence rate (14)
Emission rate (10)
Air kerma (10)
Surface emission rate (7)
Ambient dose equivalent rate (6)
Personal dose equivalent rate (6)
Emission anisotropy (6)
CHEMICAL MATERIAL >
nitrogen (183)
methane (44)
synthetic air (27)
nitrogen or air (17)
monoelemental solution (14)
CHEMICAL ANALYTE >
methane (21)
carbon dioxide (13)
carbon monoxide (12)
propane (12)
ethane (9)
GEOGRAPHIC LOCATION
EURAMET (1006)
United Kingdom (999)
Denmark (7)
Result of the search
Your query 'United Kingdom' produced more than 1000 results New search
[Prev] 2  3  4 [Next >>]

 
United Kingdom, NPL (National Physical Laboratory)
Complete CMCs in Photometry and Radiometry for United Kingdom (.PDF file)
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.6 to 3.2, depending on wavelength
Double grating monochromator
Wavelength: 200 nm to 210 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.0
Double grating monochromator
Wavelength: 211 nm to 239 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.7 at 240 nm, 0.5 above 240 nm
Double grating monochromator
Wavelength: 240 nm to 315 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.3
Double grating monochromator
Wavelength: 316 nm to 404 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.1
Double grating monochromator
Wavelength: 405 nm to 919 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.3
Double grating monochromator
Wavelength: 920 nm to 1000 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.3
Double grating monochromator
Wavelength: 1001 nm to 1400 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.3 to 0.4 (depending on wavelength)
Double grating monochromator
Wavelength: 1401 nm to 1800 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Approved on 20 April 2017
 
Responsivity, spectral, irradiance. Broad band detector. Measurand unit, A/(W m-2), V/(W m-2), reading/(W m-2)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.2
Double grating monochromator
Wavelength: 200 nm to 210 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Other types of detector can also be calibrated
Approved on 20 April 2017
 
Responsivity, spectral, irradiance. Broad band detector. Measurand unit, A/(W m-2), V/(W m-2), reading/(W m-2)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1
Double grating monochromator and reference detectors
Wavelength: 211 nm to 239 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Other types of detector can also be calibrated
Approved on 20 April 2017
 
Responsivity, spectral, irradiance. Broad band detector. Measurand unit, A/(W m-2), V/(W m-2), reading/(W m-2)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.7 at 240 nm, 0.5 above 240 nm
Double grating monochromator and reference detectors
Wavelength: 240 nm to 315 nm
Bandwidth: < 10 nm
Power level: < 20 µW
Other types of detector can also be calibrated
Approved on 20 April 2017
 
Irradiance, spectral. Tungsten lamp, 5E-04 W/(m2 nm) to 0.25 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.5 to 0.9 (with wavelength)
Spectroradiometer
Wavelength: 400 nm to 500 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017
 
Irradiance, spectral. Tungsten lamp, 6E-03 W/(m2 nm) to 0.27 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.9 to 0.8 (with wavelength)
Spectroradiometer
Wavelength: 500 nm to 800 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017
 
Irradiance, spectral. Tungsten lamp, 1.5E-03 W/(m2 nm) to 0.29 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.5
Spectroradiometer
Wavelength: 801 nm to 1600 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017
 
Irradiance, spectral. Tungsten lamp, 1.5E-03 W/(m2 nm) to 0.29 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.8 to 1.2 (with wavelength)
Spectroradiometer
Wavelength: 1601 nm to 2400 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017
 
Irradiance, spectral. Tungsten lamp, 1.5E-03 W/(m2 nm) to 0.29 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.8
Spectroradiometer
Wavelength: 2401 nm to 2500 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017
 
Radiance, spectral. Tungsten lamp, 1E-05 W/(m2 sr nm) to 0.1 W/(m2 sr nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.5 to 1.9 (with wavelength)
Spectroradiometer
Wavelength range: 250 nm to 300 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017
 
Radiance, spectral. Tungsten lamp, 1E-05 W/(m2 sr nm) to 0.1 W/(m2 sr nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.9
Spectroradiometer
Wavelength range: 300 nm to 315 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017
 
Radiance, spectral. Tungsten lamp, 1E-05 W/(m2 sr nm) to 0.1 W/(m2 sr nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.9 to 1.2 (with wavelength)
Spectroradiometer
Wavelength range: 315 nm to 430 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017
 
Radiance, spectral. Tungsten lamp, 1E-04 W/(m2 sr nm) to 1 W/(m2 sr nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.2 to 0.8, with wavelength
Spectroradiometer
Wavelength range: 430 nm to 600 nm
Bandwidth: < 20 nm
Other types of source can also be measured
Approved on 20 April 2017