Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (12)
PHYSICS
Time interval (12)
Time interval (12)
GEOGRAPHIC LOCATION
APMP (12)
Korea, Republic of (12)
Result of the search
Your query 'kriss' produced 12 results New search

 
Korea, Republic of, KRISS (Korea Research Institute of Standards and Science)
Complete CMCs in Time and Frequency for Korea, Republic of (.PDF file)
Time interval. Rise/fall time meter, 5E-09 s to 1 s
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in ns: 2
Direct rise/fall time measurement
Number of measurements: 100
Amplitude: > 0.5 V (50 )
Slew rate: > 0.5 V/ns
DUT effects excluded
Approved on 13 February 2013
Internal NMI service identifier: C301062
 
Time interval. Pulse width meter, 5E-09 s to 1 s
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in ns: 2
Direct pulse width measurement
Number of measurements: 100
Amplitude: > 0.5 V (50 )
Slew rate: > 0.5 V/ns
DUT effects excluded
Approved on 13 February 2013
Internal NMI service identifier: C301062
 
Time interval. Time difference meter, 5E-09 s to 100 s
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in ns: 2
Direct time interval measurement
Number of measurements: 100
Amplitude: > 0.5 V (50 )
Slew rate: > 0.5 V/ns
DUT effects excluded
Approved on 13 February 2013
Internal NMI service identifier: C301062
 
Time interval. Delay meter, 5E-09 s to 100 s
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in ns: 3
Direct time interval measurement
Number of measurements: 100
Amplitude: > 0.5 V (50 )
Slew rate: > 0.5 V/ns
DUT effects excluded
Approved on 13 February 2013
Internal NMI service identifier: C301062
 
Time interval. Rise/fall time source, 5E-09 s to 1 s
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in ns: 2
Direct rise/fall time measurement
Number of measurements: 100
Slew rate: > 0.5 V/ns
Amplitude: > 0.5 V (50 )
DUT effects excluded
Approved on 13 February 2013
Internal NMI service identifier: C301051
 
Time interval. Pulse width source, 5E-09 s to 1 s
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in ns: 2
Direct pulse width measurement
Number of measurements: 100
Amplitude: > 0.5 V (50 )
Slew rate: > 0.5 V/ns
DUT effects excluded
Approved on 13 February 2013
Internal NMI service identifier: C301051
 
Time interval. Time difference source, 5E-09 s to 100 s
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in ns: 2
Direct time interval measurement
Number of measurements: 100
Amplitude: > 0.5 V (50 )
Slew rate: > 0.5 V/ns
DUT effects excluded
Approved on 13 February 2013
Internal NMI service identifier: C301051
 
Time interval. Delay source, 5E-09 s to 100 s
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in ns: 3
Direct time interval measurement
Number of measurements: 100
Amplitude: > 0.5 V (50 )
Slew rate: > 0.5 V/ns
DUT effects excluded
Approved on 13 February 2013
Internal NMI service identifier: C301051
 
Time interval. Period meter, 0.01 s to 10 s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in s/s: 1E-11
Direct period measurement
Gate time: 100 s
Number of measurements: 100
Amplitude: > 0.5 V (50 )
DUT effects excluded
Approved on 13 February 2013
Internal NMI service identifier: C301043
 
Time interval. Period source, 5E-09 s to 0.01 s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in s/s: 1E-12
Direct period measurement
Gate time: 100 s
Number of measurements: 100
Amplitude: > 0.5 V (50 )
DUT effects excluded
Approved on 13 February 2013
Internal NMI service identifier: C301031
 
Time interval. Period source, 0.01 s to 10 s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in s/s: 1E-11
Direct period measurement
Gate time: 100 s
Number of measurements: 100
Amplitude: > 0.5 V (50 )
DUT effects excluded
Approved on 13 February 2013
Internal NMI service identifier: C301031
 
Time interval. Period meter, 5E-09 s to 0.01 s
Relative expanded uncertainty (k = 2, level of confidence 95 %) in s/s: 1E-12
Direct period measurement
Gate time: 100 s
Number of measurements: 100
Amplitude: > 0.5 V (50 )
DUT effects excluded
Approved on 13 February 2013
Internal NMI service identifier: C301043