Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs General Physics (5)
PHYSICS
Time interval (5)
Time interval (5)
GEOGRAPHIC LOCATION
COOMET (5)
Belarus (5)
Result of the search
Your query 'belarus' produced 5 results New search

 
Belarus, BelGIM (Belarussian State Institute for Metrology)
Complete CMCs in Time and Frequency for Belarus (.PDF file)
Time interval. Period source, 1E-08 s to 2 s
Absolute expanded uncertainty (k = 2, level of confidence 95%) in s: 2E-10
Time interval meter
Slewing rate: >= 0.5 V/ns
Approved on 23 April 2010
Internal NMI service identifier: BelGIM/3.1.1.1
 
Time interval. Rise/fall time source, 2.5E-11 s to 1E-08 s
Absolute expanded uncertainty (k = 2, level of confidence 95%) in s: (8E-12 + 1E-03T), T time interval in s
Oscilloscope
Pulse amplitude: - 1 V to +1 V
Approved on 23 April 2010
Internal NMI service identifier: BelGIM/3.2.1.1
 
Time interval. Rise/fall time source, 5E-09 s to 1E-03 s
Absolute expanded uncertainty (k = 2, level of confidence 95%) in s: (2E-09 + 1E-03T), T time interval in s
Oscilloscope
Pulse amplitude: - 60 V to +60 V
Approved on 23 April 2010
Internal NMI service identifier: BelGIM/3.2.1.2
 
Time interval. Pulse width source, 1E-08 s to 2 s
Absolute expanded uncertainty (k = 2, level of confidence 95%) in s: 2E-09
Time interval meter
Slewing rate: >= 0.5 V/ns
Approved on 23 April 2010
Internal NMI service identifier: BelGIM/3.2.2.1
 
Time interval. Rise/fall time meter, 2.5E-11 s to 5E-07 s
Relative expanded uncertainty (k = 2, level of confidence 95%) in s/s: 8E-02
Calibrated rise/fall time source
Pulse amplitude: 5 mV to 3 V
Approved on 23 April 2010
Internal NMI service identifier: BelGIM/3.4.2.1