Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
The BIPM key comparison database
Refine your search
CMC AREA
CMCs Ionizing Radiation (7)
IONIZING RADIATION
Emission rate (7)
Am-241 (1)
Po-210 (1)
Cm-242 (1)
Cm-244 (1)
Pb-210 (1)
GEOGRAPHIC LOCATION
APMP (7)
Japan (7)
Result of the search
Your query 'Japan' produced 7 results New search

 
Japan, NMIJ AIST (National Metrology Institute of Japan)
Complete CMCs in Ionizing Radiation for Japan (.PDF file)
Emission rate. Electroplated source, 2.0E+02 s-1 to 2.00E+04 s-1
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
Si surface barrier detector
Pb-210/Po-210: electroplated source less than 2.5 cm diameter
Reference standard: particle counting with defined solid angle
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2182
Emission rate. Neutron source, 1.0E+03 s-1 to 3.0E+07 s-1
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.2
Measurement of thermal neutron fluence rate in a standard graphite pile relative to a calibrated neutron source
Cf-252 source: sealed neutron source
Reference standard: Calibrated neutron source
Source of traceability: NPL
Approved on 01 June 2011
Internal NMI service identifier: APM-RAD-NMIJ/AIST-3012
Emission rate. Neutron source, 1.0E+03 s-1 to 2.0E+07 s-1
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.0
Measurement of thermal neutron fluence rate in a standard graphite pile relative to a calibrated neutron source
Am-241/Be-9 source: sealed neutron source
Reference standard: Calibrated neutron source
Source of traceability: NPL
Approved on 01 June 2011
Internal NMI service identifier: APM-RAD-NMIJ/AIST-3013
Emission rate. Electroplated source, 2.0E+02 s-1 to 2.00E+04 s-1
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
Si surface barrier detector
Po-210: electroplated source less than 2.5 cm diameter
Reference standard: particle counting with defined solid angle
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2183
Emission rate. Electroplated source, 2.0E+02 s-1 to 2.00E+04 s-1
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
Si surface barrier detector
Am-241: electroplated source less than 2.5 cm diameter
Reference standard: particle counting with defined solid angle
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2184
Emission rate. Electroplated source, 2.0E+02 s-1 to 2.00E+04 s-1
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
Si surface barrier detector
Cm-242: electroplated source less than 2.5 cm diameter
Reference standard: particle counting with defined solid angle
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2185
Emission rate. Electroplated source, 2.0E+02 s-1 to 2.00E+04 s-1
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2
Si surface barrier detector
Cm-244: electroplated source less than 2.5 cm diameter
Reference standard: particle counting with defined solid angle
Source of traceability: NMIJ/AIST
Approved on 14 February 2005
Internal NMI service identifier: APM-RAD-NMIJ/AIST-2186