Your query 'NIMT' produced 5 results New search

 
Thailand, NIMT (National Institute of Metrology (Thailand))
Complete CMCs in Photometry and Radiometry for Thailand (.PDF file)
Irradiance, spectral. Tungsten lamp, 3E-05 W/(m2 nm) to 4E-02 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 6.6
Monochromator
Wavelength range: 290 nm to 300 nm
Bandwidth: 5 nm
Approved on 16 January 2011
Internal NMI service identifier: NIMT/13030-10101
 
Irradiance, spectral. Tungsten lamp, 3E-05 W/(m2 nm) to 4E-02 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 5.6
Monochromator
Wavelength range: 300 nm to 500 nm (excludes 300 nm)
Bandwidth: 5 nm
Approved on 16 January 2011
Internal NMI service identifier: NIMT/13030-10101
 
Irradiance, spectral. Tungsten lamp, 6E-03 W/(m2 nm) to 0.13 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.3
Monochromator
Wavelength range: 500 nm to 600 nm (excludes 500 nm and 600 nm)
Bandwidth: 5 nm
Approved on 16 January 2011
Internal NMI service identifier: NIMT/13030-10101
 
Irradiance, spectral. Tungsten lamp, 2E-02 W/(m2 nm) to 0.2 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.9
Monochromator
Wavelength range: 600 nm to 830 nm (excludes 830 nm)
Bandwidth: 5 nm
Approved on 16 January 2011
Internal NMI service identifier: NIMT/13030-10101
 
Irradiance, spectral. Tungsten lamp, 5E-03 W/(m2 nm) to 0.2 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 6.5
Monochromator
Wavelength range: 830 nm to 900 nm
Bandwidth: 10 nm
Approved on 16 January 2011
Internal NMI service identifier: NIMT/13030-10101