Japan, NMIJ AIST (National Metrology Institute of Japan)
Complete CMCs in Photometry and Radiometry for Japan (.PDF file)
Distribution temperature. Tungsten lamp, 2000 K to 3400 K
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 15
Spectroradiometer
Approved on 16 January 2011
 
Responsivity, laser power. General detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.7
Laser comparator
Wavelength: 488 nm, 515 nm
Power level: > 200 mW to 1 W
Type of detector: thermopile, pyroelectric, solid state
Approved on 16 January 2011
 
Responsivity, laser power. General detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.22
Laser comparator
Wavelength: 404 nm
Power level: 0.05 mW to < 0.1 mW
Type of detector: thermopile, pyroelectric, solid state
Approved on 16 January 2011
 
Responsivity, laser power. General detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.17
Laser comparator
Wavelength: 404 nm
Power level: 0.1 mW to < 1 mW
Type of detector: thermopile, pyroelectric, solid state
Approved on 16 January 2011
 
Responsivity, laser power. General detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.13
Laser comparator
Wavelength: 404 nm
Power level: 1 mW to 10 mW
Type of detector: thermopile, pyroelectric, solid state
Approved on 16 January 2011
 
Responsivity, laser power. General detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.1
Laser comparator
Wavelength: 1.06 µm
Power level: 1 W to 10 W
Type of detector: thermopile, pyroelectric, solid state
Approved on 16 January 2011
 
Responsivity, laser power. General detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.3
Laser comparator
Wavelength: 10.6 µm
Power level: 1 W to 10 W
Type of detector: thermopile, pyroelectric, solid state
Approved on 16 January 2011
 
Irradiance, spectral. Tungsten lamp, 3E-05 W/(m2 nm) to 9E-03 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.8
Monochromator
Wavelength range: 250 nm to 350 nm
Bandwidth: 5 nm
Approved on 16 January 2011
 
Irradiance, spectral. Tungsten lamp, 1E-03 W/(m2 nm) to 4E-02 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.2
Monochromator
Wavelength range: 350 nm to 450 nm (excludes 350 nm)
Bandwidth: 5 nm
Approved on 16 January 2011
 
Irradiance, spectral. Tungsten lamp, 6E-03 W/(m2 nm) to 0.15 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.8
Monochromator
Wavelength range: 450 nm to 600 nm (excludes 450 nm)
Bandwidth: 5 nm
Approved on 16 January 2011
 
Irradiance, spectral. Tungsten lamp, 2E-02 W/(m2 nm) to 0.25 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.0
Monochromator
Wavelength range: 600 nm to 830 nm (excludes 600 nm)
Bandwidth: 5 nm
Approved on 16 January 2011
 
Irradiance, spectral. Tungsten lamp, 1.5E-02 W/(m2 nm) to 0.25 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.4
Monochromator
Wavelength range: 830 nm to 2000 nm (excludes 830 nm)
Bandwidth: 5 nm to 3 nm, varies with wavelength
Approved on 16 January 2011
 
Irradiance, spectral. Tungsten lamp, 1.5E-02 W/(m2 nm) to 0.25 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.4
Monochromator
Wavelength range: 2000 nm to 2300 nm (excludes 2000 nm)
Bandwidth: 10 nm to 8 nm, varies with wavelength
Approved on 16 January 2011
 
Irradiance, spectral. Tungsten lamp, 5E-03 W/(m2 nm) to 5.5E-02 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 6.0
Monochromator
Wavelength range: 2300 nm to 2500 nm (excludes 2300 nm)
Bandwidth: 9 nm to 7 nm, varies with wavelength
Approved on 16 January 2011
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 4.8 to 1.6, varies with wavelength
Double grating monochromator
Wavelength: 200 nm to 250 nm (excludes 250 nm)
Bandwidth: 5 nm
Power level: < 10 µW
Approved on 16 January 2011
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.6
Double grating monochromator
Wavelength: 250 nm to 380 nm (excludes 380 nm)
Bandwidth: 5 nm
Power level: < 10 µW
Approved on 16 January 2011
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.0 to 0.45, varies with wavelength
Double grating monochromator
Wavelength: 380 nm to 650 nm (excludes 650 nm)
Bandwidth: 5 nm
Power level: < 10 µW
Approved on 16 January 2011
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.45 to 0.56, varies with wavelength
Double grating monochromator
Wavelength: 650 nm to 930 nm (excludes 930 nm)
Bandwidth: 5 nm
Power level: < 10 µW
Approved on 16 January 2011
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.56 to 2.9, varies with wavelength
Double grating monochromator
Wavelength: 930 nm to 1150 nm
Bandwidth: 5 nm
Power level: < 5 µW
Approved on 16 January 2011
 
Responsivity, spectral, power. Broadband detector. Measurand unit, A/W
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 7.9 to 5.7, varies with wavelength
Seya-Namioka monochromator
Wavelength: 140 nm to 160 nm (excludes 160 nm)
Bandwidth: 3.5 nm
Power level: < 2 µW
Approved on 16 January 2011