Your query 'kriss' produced 19 results New search

 
Korea, Republic of, KRISS (Korea Research Institute of Standards and Science)
Complete CMCs in Photometry and Radiometry for Korea, Republic of (.PDF file)
Distribution temperature. Tungsten lamp, 2500 K to 3200 K
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 21 to 42
Reference spectroradiometer
Approved on 09 September 2011
 
Correlated color temperature. Tungsten lamp, 2500 K to 3200 K
Absolute expanded uncertainty (k = 2, level of confidence 95 %) in K: 15 to 20
Reference spectroradiometer
Approved on 09 September 2011
 
Color, emitted. General source, color space x, y,
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.001 to 0.005 (depending on type of source)
Reference spectroradiometer
Bandwidth: 3 nm
Type of source: tungsten-based source, LEDs (R, G, B, W)
Approved on 09 September 2011
 
Color, emitted. General source, color space u, v,
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.0004 to 0.006 (depending on type of source)
Reference spectroradiometer
Bandwidth: 3 nm
Type of source: tungsten-based source, LEDs (R, G, B, W)
Approved on 09 September 2011
 
Color, emitted. General source, color space u-prime, v-prime,
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.0004 to 0.009 (depending on type of source)
Reference spectroradiometer
Bandwidth: 3 nm
Type of source: tungsten-based source, LEDs (R, G, B, W)
Approved on 09 September 2011
 
Chromaticity response. Colorimeter, (CIE 1931 x, y)
Absolute expanded uncertainty (k = 2, level of confidence 95 %): 0.003 to 0.008 (depending on type of source)
Comparison at LCD display
Type of source: LCD display with colour patterns (R, G, B, W)
Approved on 09 September 2011
 
Responsivity, laser, power. General detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.4
Power-stabilized continuous-wave lasers
Wavelengths: 633 nm, 650 nm, 780 nm
Power level: 0.01 mW to 10 mW
Type of detector: laser power meter
Approved on 04 June 2007
 
Responsivity, blackbody total irradiance. Broadband detector. Measurand unit, A/(W m-2), V/(W m-2), reading/(W m-2)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.3
Variable high-temperature blackbody
Irradiance level: 1 kW/m2 to 10 kW/m2
Distribution temperature: 3000 K to 3100 K
Type of detector: Schmidt-Boelter heat-flux meter
Approved on 03 September 2009
 
Responsivity, UV BB irradiance. UV radiometer. Measurand unit, rA/(W m-2), V/(W m-2), reading/(W m-2)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.6
Comparison at filtered Hg lamp
Wavelength range: 365 nm ± 5 nm
Irradiance level: 0.1 mW/cm2 to 250 mW/cm2
Type of source: Hg lamp with neutral density and bandpass filters
Approved on 03 September 2009
 
Responsivity, UV BB irradiance. UV radiometer. Measurand unit, A/(W m-2), V/(W m-2), reading/(W m-2)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 3.6
Comparison at filtered Hg lamp
Wavelength range: 405 nm ± 5 nm
Irradiance level: 0.1 mW/cm2 to 250 mW/cm2
Type of source: Hg lamp with neutral density and bandpass filters
Approved on 03 September 2009
 
Power, spectral total radiant. Laser, 1.0E-05 W to 1.0E-02 W
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.4
Reference detector
Wavelength range: 325 nm to 1064 nm
Type of laser: continuous-wave
Approved on 04 June 2007
 
Responsivity, spectral, power. Broad band detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 7.5 to 1.1
Comparison at lamp-based monochromatic source
Wavelength range: 300 nm to 450 nm
Bandwidth: 1 nm
Power level: 4E-10 W to 5E-07 W
Approved on 09 September 2011
 
Responsivity, spectral, power. Broad band detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.1
Comparison at lamp-based monochromatic source
Wavelength range: 450 nm to 950 nm
Bandwidth: 1 nm
Power level: 4E-10 W to 5E-07 W
Approved on 09 September 2011
 
Responsivity, spectral, power. Broad band detector. Measurand unit, A/W, V/W, reading/W, W, dB, dBm
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.1 to 3.6
Comparison at lamp-based monochromatic source
Wavelength range: 950 nm to 1000 nm
Bandwidth: 1 nm
Power level: 4E-10 W to 5E-07 W
Approved on 09 September 2011
 
Responsivity, spectral, irradiance. Spectroradiometer. Measurand unit, A/(W m-2 nm), V/(W m-2 nm), reading/(W m-2)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 4.6 to 2.6
Irradiance standard lamp
Wavelength range: 250 nm to 850 nm
Bandwidth: 5 nm
Power level: 1E-04 W/(m2nm) to 3E-01 W/(m2nm)
Approved on 09 September 2011
 
Responsivity, spectral, irradiance. Spectroradiometer. Measurand unit, A/(W m-2 nm), V/(W m-2 nm), reading/(W m-2)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 2.6 to 4.2
Irradiance standard lamp
Wavelength range: 850 nm to 1700 nm
Bandwidth: 5 nm
Power level: 1E-01 W/(m2nm) to 3E-01 W/(m2nm)
Approved on 09 September 2011
 
Irradiance, spectral. Tungsten lamp, 1.0E-04 W/(m2 nm) to 1 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.7 to 0.6
Spectral comparison with high-temperature blackbody
Wavelength range: 250 nm to 700 nm
Bandwidth: 4.8 nm
Approved on 09 September 2011
 
Irradiance, spectral. Tungsten lamp, 1.0E-04 W/(m2 nm) to 1 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 0.6 to 1.5
Spectral comparison with high-temperature blackbody
Wavelength range: 700 nm to 1100 nm
Bandwidth: 4.8 nm
Approved on 09 September 2011
 
Irradiance, spectral. Tungsten lamp, 1.0E-04 W/(m2 nm) to 1 W/(m2 nm)
Relative expanded uncertainty (k = 2, level of confidence 95 %) in %: 1.5 to 2.5
Spectral comparison with high-temperature blackbody
Wavelength range: 1100 nm to 2500 nm
Bandwidth: 4.8 nm
Approved on 09 September 2011